E Di Castro, R Pani, R Pellegrini, C Bacci.
Abstract
A method is introduced for the evaluation of x-ray spectra from x-ray machines operating in the range 50-100 kVp using a cadmium telluride (CdTe) detector with low detection efficiency. The pulse height distribution obtained with this kind of detector does not represent the true photon spectra owing to the presence of K-escape, Compton scattering, etc.; these effects were evaluated using a Monte Carlo method. A stripping procedure is described for implementation on a Univac 1100/82 computer. The validity of our method was finally tested by comparison with experimental results obtained with a Ge detector and with data from the literature; the results are in good agreement with published data.Entities:
Mesh:
Substances:
Year: 1984 PMID: 6483976 DOI: 10.1088/0031-9155/29/9/008
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609