Pantak Therapax SXT 150: performance assessment and dose determination using IAEA TRS-398 protocol.

D Jurado1, T Eudaldo, P Carrasco, N Jornet, A Ruiz, M Ribas.   

Abstract

The performance assessment and beam characteristics of the Therapax SXT 150 unit, which encompass both low and medium-energy beams, were evaluated. Dose determination was carried out by implementing the International Atomic Energy Agency (IAEA) TRS-398 protocol and measuring all the dosimetric parameters in order to have a solid, consistent and reliable data set for the unit. Mechanical movements, interlocks and applicator characteristics agreed with specifications. The timer exhibited good accuracy and linearity. The output was very stable, with good repeatability, long-term reproducibility and no dependence on tube head orientation. The measured dosimetric parameters included beam first and second half-value layers (HVLs), absorbed dose rate to water under reference conditions, central axis depth dose distributions, output factors and beam profiles. Measured first HVLs agreed with comparable published data, but the homogeneity coefficients were low in comparison with typical values found in the literature. The timer error was significant for all filters and should be taken into consideration for the absorbed dose rate determination under reference conditions as well as for the calculation of treatment times. Percentage depth-dose (PDD) measurements are strongly recommended for each filter-applicator combination. The output factor definition of the IAEA TRS-398 protocol for medium-energy X-ray qualities involves the use of data that is difficult to measure. Beam profiles had small penumbras and good symmetry and flatness except for the lowest energy beam, for which a heel effect was observed.

Entities:  

Mesh:

Year:  2005        PMID: 16046424     DOI: 10.1259/bjr/15782649

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  5 in total

1.  Acceptance, commissioning and clinical use of the WOmed T-200 kilovoltage X-ray therapy unit.

Authors:  Maria M Aspradakis; Paolo Zucchetti
Journal:  Br J Radiol       Date:  2015-07-30       Impact factor: 3.039

2.  Validating plastic scintillation detectors for photon dosimetry in the radiologic energy range.

Authors:  Francois Lessard; Louis Archambault; Mathieu Plamondon; Philippe Despres; Franccois Therriault-Proulx; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

3.  Use of calculations to validate beam quality and relative dose measurements for a kilovoltage X-ray therapy unit.

Authors:  B J Healy; R F Hill
Journal:  Phys Eng Sci Med       Date:  2022-04-05

4.  Dosimetric characteristics of a new unit for electronic skin brachytherapy.

Authors:  Teresa Garcia-Martinez; Jan-Pieter Chan; Jose Perez-Calatayud; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2014-02-19

5.  Modeling a superficial radiotherapy X-ray source for relative dose calculations.

Authors:  Christopher D Johnstone; Richard LaFontaine; Yannick Poirier; Mauro Tambasco
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.