Ibrahim Alkatout1, Thoralf Schollmeyer, Nusrat A Hawaldar, Nidhi Sharma, Liselotte Mettler.
Abstract
BACKGROUND: Electrosurgical units are the most common type of electrical equipment in the operating room. A basic understanding of electricity is needed to safely apply electrosurgical technology for patient care.Entities:
Mesh:
Year: 2012 PMID: 22906341 PMCID: PMC3407433 DOI: 10.4293/108680812X13291597716348
Source DB: PubMed Journal: JSLS ISSN: 1086-8089 Impact factor: 2.172
Comparison of Tissue Effects of 4 Energy Modalities[4]
| Monopolar | Traditional Bipolar | Advanced Bipolar | Ultrasonic | |
|---|---|---|---|---|
| Tissue Effect | Cutting, Coagulation | Coagulation | Cutting, coagulation | Cutting, coagulation |
| Power Setting | 50–80 W | 30–50W | DEFAULT | 55,000 Hz frequency |
| Thermal Spread | Not well assessed | 2–6mm | 1–4mm | 1–4mm |
| Maximum Temperature | >100°C | >100°C | Not well assessed | <80°C |
| Vessel Sealing Ability | Not applicable | Not applicable | Seals vessels ≤7mm | Seals vessels ≤5mm |
| Technique | Not applicable | Not applicable | Tension free application | Tension free application |