Literature DB >> 9685531

Active electrode monitoring. How to prevent unintentional thermal injury associated with monopolar electrosurgery at laparoscopy.

T G Vancaillie1.   

Abstract

BACKGROUND: In recent years, the use of minimally invasive surgery (MIS) has expanded to a wide variety of surgical specialties. The increased popularity of the procedure, however, has been accompanied by its share of complications, including trocar lacerations and inadvertent thermal injuries to nontargeted tissues during monopolar electrosurgery.
METHODS: A survey on electrosurgical thermal injuries and three case studies are presented. The new technology of active electrode monitoring (AEM) is described.
RESULTS: AEM eliminates stray currents generated by insulation failure and capacitive coupling.
CONCLUSIONS: To reduce the incidence of injury by monopolar electrosurgery at laparoscopy, there is a need for advanced technology, such as AEM. In addition, laparoscopic surgeons should be encouraged to study the basic concepts of the biophysics of electrosurgery.

Entities:  

Mesh:

Year:  1998        PMID: 9685531     DOI: 10.1007/s004649900769

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  11 in total

1.  Intraoperative electrode burns.

Authors:  Michael J Russell; Michael Gaetz
Journal:  J Clin Monit Comput       Date:  2004-02       Impact factor: 2.502

2.  Laparoscopy for acute small bowel obstruction: indication or contraindication?

Authors:  Ioannis Tierris; Constantinos Mavrantonis; Constantinos Stratoulias; George Panousis; Afrodite Mpetsou; Nicolaos Kalochristianakis
Journal:  Surg Endosc       Date:  2010-07-07       Impact factor: 4.584

3.  Insulation failure in electrosurgery instrumentation: a prospective evaluation.

Authors:  Floriane Tixier; Mélanie Garçon; Françoise Rochefort; Stéphane Corvaisier
Journal:  Surg Endosc       Date:  2016-03-16       Impact factor: 4.584

4.  Insulation failure in laparoscopic instruments.

Authors:  Paul N Montero; Thomas N Robinson; John S Weaver; Greg V Stiegmann
Journal:  Surg Endosc       Date:  2009-07-02       Impact factor: 4.584

Review 5.  Common uses and cited complications of energy in surgery.

Authors:  Ganesh Sankaranarayanan; Rajeswara R Resapu; Daniel B Jones; Steven Schwaitzberg; Suvranu De
Journal:  Surg Endosc       Date:  2013-04-23       Impact factor: 4.584

6.  Iliac vein injury due to a damaged Hot Shears™ tip cover during robot assisted radical prostatectomy.

Authors:  Enrique Ian Lorenzo; Wooju Jeong; Sangun Park; Won Tae Kim; Sung Joon Hong; Koon Ho Rha
Journal:  Yonsei Med J       Date:  2011-03       Impact factor: 2.759

Review 7.  Recurrences in laparoscopic incisional hernia repairs: a personal series and review of the literature.

Authors:  R H Koehler; G Voeller
Journal:  JSLS       Date:  1999 Oct-Dec       Impact factor: 2.172

Review 8.  Energy sources in neonatal surgery: principles and practice.

Authors:  Shandip K Sinha; Anjan Dhua
Journal:  J Neonatal Surg       Date:  2014-04-01

Review 9.  Electrosurgical unit: Iatrogenic injuries and medico-legal aspect. Italian legal rules, experience and article review.

Authors:  Patrizia Gualniera; Serena Scurria; Daniela Sapienza; Alessio Asmundo
Journal:  Ann Med Surg (Lond)       Date:  2021-01-04

10.  Principles and safety measures of electrosurgery in laparoscopy.

Authors:  Ibrahim Alkatout; Thoralf Schollmeyer; Nusrat A Hawaldar; Nidhi Sharma; Liselotte Mettler
Journal:  JSLS       Date:  2012 Jan-Mar       Impact factor: 2.172

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.