Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries.

K L Harold1, H Pollinger, B D Matthews, K W Kercher, R F Sing, B T Heniford.   

Abstract

BACKGROUND: Advanced laparoscopic procedures have necessitated the development of new technology for vascular control. Suture ligation can be time-consuming and cumbersome during laparoscopic dissection. Titanium clips have been used for hemostasis, and recently plastic clips and energy sources such as ultrasonic coagulating shears and bipolar thermal energy devices have become popular. The purpose of this study was to compare the bursting pressure of arteries sealed with ultrasonic coagulating shears (UCS), electrothermal bipolar vessel sealer (EBVS), titanium laparoscopic clips (LCs), and plastic laparoscopic clips (PCs). In addition, the spread of thermal injury from the UCS and the EBVS was compared.
METHODS: Arteries in three size groups (2-3, 4-5 and 6-7 mm) were harvested from freshly euthanized pigs. Each of the four devices was used to seal 16 specimens from each size group for burst testing. A 5-Fr catheter was placed into the open end of the specimen and secured with a purse-string suture. The catheter was connected to a pressure monitor and saline was infused until there was leakage from the sealed end. This defined the bursting pressure in mmHg. The ultrasonic shears and bipolar thermal device were used to seal an additional 8 vessels in each size group, which were sent for histologic examination. These were examined with hematoxylin and eosin stains, and the extent of thermal injury, defined by coagulation necrosis, was measured in millimeters. Analysis of variance was performed and, where appropriate, a Tukey's test was also performed.
RESULTS: The EBVS's mean burst pressure was statistically higher than that of the UCS at 4 or 5 mm (601 vs 205 mmHg) and 6 or 7 mm (442 vs 175 mmHg). EBVS had higher burst pressures for the 4 or 5-mm group (601 mmHg) and 6 or 7-mm group (442 mmHg) compared with its pressure at 2 or 3 mm (128 mmHg) ( p = 0.0001). The burst pressures of the UCS and EBVS at 2 or 3 mm were not significantly different. Both clips were statistically stronger than the thermal devices except at 4 or 5 mm, in which case the EBVS was as strong as the LC (601 vs 593 mmHg). The PC and LC were similar except at 4 or 5 mm, where the PC was superior (854 vs 593 mmHg). The PC burst pressure for 4 or 5 mm (854 mmHg) was statistically higher than that for vessels 2 or 3 mm (737 mmHg) but not different from the 6 or 7 mm pressure (767 mmHg). Thermal spread was not statistically different when comparing EBVS and UCS at any size (EBVS mean = 2.57 mm vs UCS mean = 2.18 mm).
CONCLUSIONS: Both the PC and LC secured all vessel sizes to well above physiologic levels. The EBVS can be used confidently in vessels up to 7 mm. There is no difference in the thermal spread of the LigaSure vessel sealer and the UCS.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12799888     DOI: 10.1007/s00464-002-8833-7

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  5 in total

1.  Use of fibrin sealant for prosthetic mesh fixation in laparoscopic extraperitoneal inguinal hernia repair.

Authors:  N Katkhouda; E Mavor; M H Friedlander; R J Mason; M Kiyabu; S W Grant; K Achanta; E L Kirkman; K Narayanan; R Essani
Journal:  Ann Surg       Date:  2001-01       Impact factor: 12.969

2.  Initial results with an electrothermal bipolar vessel sealer.

Authors:  B T Heniford; B D Matthews; R F Sing; C Backus; B Pratt; F L Greene
Journal:  Surg Endosc       Date:  2001-05-14       Impact factor: 4.584

3.  Comparison of thermal spread after ureteral ligation with the Laparo-Sonic ultrasonic shears and the Ligasure system.

Authors:  Sharon L Goldstein; Kristi L Harold; Alan Lentzner; Brent D Matthews; Kent W Kercher; Ronald F Sing; Broc Pratt; Edward H Lipford; B Todd Heniford
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2002-02       Impact factor: 1.878

4.  The use of bipolar cautery, laparosonic coagulating shears, and vascular clips for hemostasis of small and medium-sized vessels.

Authors:  H Spivak; W S Richardson; J G Hunter
Journal:  Surg Endosc       Date:  1998-02       Impact factor: 4.584

5.  High-burst-strength, feedback-controlled bipolar vessel sealing.

Authors:  J S Kennedy; P L Stranahan; K D Taylor; J G Chandler
Journal:  Surg Endosc       Date:  1998-06       Impact factor: 4.584

  5 in total
  74 in total

1.  Efficacy and safety of the novel electrosurgical vessel sealing and cutting instrument BiCision®.

Authors:  Ralf Rothmund; Bernhard Kraemer; Felix Neis; Sara Brucker; Markus Wallwiener; Ali Reda; Andrea Hausch; Marcus Scharpf; Mara Natascha Szyrach
Journal:  Surg Endosc       Date:  2012-06-08       Impact factor: 4.584

2.  Thermal fusion: effects and interactions of temperature, compression, and duration variables.

Authors:  Deogracias A G Reyes; Stuart I Brown; Lynda Cochrane; Luisa S Motta; Alfred Cuschieri
Journal:  Surg Endosc       Date:  2012-07-07       Impact factor: 4.584

3.  BiClamp forceps significantly shorten the operation time for breast surgery.

Authors:  Hideki Takeuchi; Toshiaki Saeki; Takashi Shigekawa; Hiroshi Sano; Noriko Nakamiya; Kazuo Matsuura; Misono Misumi; Takao Takahashi; Nobuko Fujiuchi; Katsuhiko Okubo; Akihiko Osaki; Takaki Sakurai; Isamu Koyama
Journal:  Surg Today       Date:  2010-07-30       Impact factor: 2.549

4.  An evaluation of electrosurgical vessel-sealing devices in biliary tract surgery in a porcine model.

Authors:  William W Hope; Srikanth Padma; William L Newcomb; Thomas M Schmelzer; Jessica J Heath; Amy E Lincourt; B Todd Heniford; H James Norton; John B Martinie; David A Iannitti
Journal:  HPB (Oxford)       Date:  2010-12       Impact factor: 3.647

5.  Electrothermal bipolar vessel ligation improves operative time during laparoscopic total proctocolectomy: a large single-center experience.

Authors:  Richard Garfinkle; Marylise Boutros; Neha Hippalgaonkar; Geva Maimon; Giovanna da Silva; Fabio Potenti; Steven D Wexner
Journal:  Surg Endosc       Date:  2015-10-28       Impact factor: 4.584

6.  Energy vessel sealing systems versus mechanical ligature of the inferior mesenteric artery in laparoscopic sigmoidectomy.

Authors:  Bertrand Trilling; Romain Riboud; Julio Abba; Edouard Girard; Jean-Luc Faucheron
Journal:  Int J Colorectal Dis       Date:  2016-02-03       Impact factor: 2.571

7.  Efficacy and quality of vessel sealing: comparison of a reusable with a disposable device and effects of clamp surface geometry and structure.

Authors:  S Richter; O Kollmar; M K Schilling; G A Pistorius; M D Menger
Journal:  Surg Endosc       Date:  2006-05-12       Impact factor: 4.584

8.  Comparative study of electrothermal bipolar vessel sealer and ultrasonic coagulating shears in laparoscopic colectomy.

Authors:  M Takada; T Ichihara; Y Kuroda
Journal:  Surg Endosc       Date:  2004-12-02       Impact factor: 4.584

9.  In vivo comparison of the coagulation capability of SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries.

Authors:  Ronald H Clements; Rajendra Palepu
Journal:  Surg Endosc       Date:  2007-05-04       Impact factor: 4.584

10.  Use of the electrothermal bipolar vessel system (EBVS) in laparoscopic adrenalectomy: a prospective study.

Authors:  Mario Guerrieri; Francesca Crosta; Angelo De Sanctis; Maddalena Baldarelli; Giovanni Lezoche; Roberto Campagnacci
Journal:  Surg Endosc       Date:  2007-05-04       Impact factor: 4.584

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.