Impact of the arterial input function on microvascularization parameter measurements using dynamic contrast-enhanced ultrasonography.

Marianne Gauthier1, Stéphanie Pitre-Champagnat, Farid Tabarout, Ingrid Leguerney, Mélanie Polrot, Nathalie Lassau.   

Abstract

AIM: To evaluate the sources of variation influencing the microvascularization parameters measured by dynamic contrast-enhanced ultrasonography (DCE-US).
METHODS: Firstly, we evaluated, in vitro, the impact of the manual repositioning of the ultrasound probe and the variations in flow rates. Experiments were conducted using a custom-made phantom setup simulating a tumor and its associated arterial input. Secondly, we evaluated, in vivo, the impact of multiple contrast agent injections and of examination day, as well as the influence of the size of region of interest (ROI) associated with the arterial input function (AIF). Experiments were conducted on xenografted B16F10 female nude mice. For all of the experiments, an ultrasound scanner along with a linear transducer was used to perform pulse inversion imaging based on linear raw data throughout the experiments. Semi-quantitative and quantitative analyses were performed using two signal-processing methods.
RESULTS: In vitro, no microvascularization parameters, whether semi-quantitative or quantitative, were significantly correlated (P values from 0.059 to 0.860) with the repositioning of the probe. In addition, all semi-quantitative microvascularization parameters were correlated with the flow variation while only one quantitative parameter, the tumor blood flow, exhibited P value lower than 0.05 (P = 0.004). In vivo, multiple contrast agent injections had no significant impact (P values from 0.060 to 0.885) on microvascularization parameters. In addition, it was demonstrated that semi-quantitative microvascularization parameters were correlated with the tumor growth while among the quantitative parameters, only the tissue blood flow exhibited P value lower than 0.05 (P = 0.015). Based on these results, it was demonstrated that the ROI size of the AIF had significant influence on microvascularization parameters: in the context of larger arterial ROI (from 1.17 ± 0.6 mm(3) to 3.65 ± 0.3 mm(3)), tumor blood flow and tumor blood volume were correlated with the tumor growth, exhibiting P values lower than 0.001.
CONCLUSION: AIF selection is an essential aspect of the deconvolution process to validate the quantitative DCE-US method.

Entities:  

Keywords:  Angiogenesis; Arterial input function; Dynamic contrast-enhanced ultrasonography; Functional imaging; Linear raw data

Year:  2012        PMID: 22900130      PMCID: PMC3419865          DOI: 10.4329/wjr.v4.i7.291

Source DB:  PubMed          Journal:  World J Radiol        ISSN: 1949-8470


  48 in total

1.  High-resolution functional imaging with ultrasound contrast agents based on RF processing in an in vivo kidney experiment.

Authors:  X A Verbeek; J M Willigers; F W Prinzen; M Peschar; L A Ledoux; A P Hoeks
Journal:  Ultrasound Med Biol       Date:  2001-02       Impact factor: 2.998

2.  Delay and dispersion effects in dynamic susceptibility contrast MRI: simulations using singular value decomposition.

Authors:  F Calamante; D G Gadian; A Connelly
Journal:  Magn Reson Med       Date:  2000-09       Impact factor: 4.668

3.  Characterization of focal liver lesions with contrast-specific US modes and a sulfur hexafluoride-filled microbubble contrast agent: diagnostic performance and confidence.

Authors:  Emilio Quaia; Fabrizio Calliada; Michele Bertolotto; Sandro Rossi; Lorena Garioni; Laura Rosa; Roberto Pozzi-Mucelli
Journal:  Radiology       Date:  2004-08       Impact factor: 11.105

4.  Imaging chemically modified adenovirus for targeting tumors expressing integrin alphavbeta3 in living mice with mutant herpes simplex virus type 1 thymidine kinase PET reporter gene.

Authors:  Zhengming Xiong; Zhen Cheng; Xianzhong Zhang; Manish Patel; Joseph C Wu; Sanjiv S Gambhir; Xiaoyuan Chen
Journal:  J Nucl Med       Date:  2006-01       Impact factor: 10.057

5.  Modeling ultrasound contrast measurement of blood flow and perfusion in biological tissue.

Authors:  Johan M Thijssen; Chris L de Korte
Journal:  Ultrasound Med Biol       Date:  2005-02       Impact factor: 2.998

6.  Dynamic CT measurement of cerebral blood flow: a validation study.

Authors:  A Cenic; D G Nabavi; R A Craen; A W Gelb; T Y Lee
Journal:  AJNR Am J Neuroradiol       Date:  1999-01       Impact factor: 3.825

7.  Characteristics of SonoVuetrade mark.

Authors:  Michel Schneider
Journal:  Echocardiography       Date:  1999-10       Impact factor: 1.724

8.  Reproducibility of perfusion parameters in dynamic contrast-enhanced MRI of lung and liver tumors: effect on estimates of patient sample size in clinical trials and on individual patient responses.

Authors:  Chaan S Ng; David L Raunig; Edward F Jackson; Edward A Ashton; Frederick Kelcz; Kevin B Kim; Razelle Kurzrock; Teresa M McShane
Journal:  AJR Am J Roentgenol       Date:  2010-02       Impact factor: 3.959

9.  Semiquantitative human cerebral perfusion assessment with ultrasound in brain space-occupying lesions: preliminary data.

Authors:  Edoardo Vicenzini; Roberto Delfini; Francesca Magri; Francesco Puccinelli; Marta Altieri; Antonio Santoro; Maria Fabrizia Giannoni; Luigi Bozzao; Vittorio Di Piero; Gian Luigi Lenzi
Journal:  J Ultrasound Med       Date:  2008-05       Impact factor: 2.153

10.  Differential sensitivity of breast cancer and melanoma cells to proteasome inhibitor Velcade.

Authors:  Azmi Yerlikaya; Nuray Erin
Journal:  Int J Mol Med       Date:  2008-12       Impact factor: 4.101

View more
  3 in total

1.  Maximum Entropy Technique and Regularization Functional for Determining the Pharmacokinetic Parameters in DCE-MRI.

Authors:  Zahra Amini Farsani; Volker J Schmid
Journal:  J Digit Imaging       Date:  2022-05-26       Impact factor: 4.903

2.  Reproducibility of Contrast-Enhanced Ultrasound in Mice with Controlled Injection.

Authors:  Alexandre Dizeux; Thomas Payen; Guillaume Barrois; Delphine Le Guillou Buffello; S Lori Bridal
Journal:  Mol Imaging Biol       Date:  2016-10       Impact factor: 3.488

3.  Modified Maximum Entropy Method and Estimating the AIF via DCE-MRI Data Analysis.

Authors:  Zahra Amini Farsani; Volker J Schmid
Journal:  Entropy (Basel)       Date:  2022-01-20       Impact factor: 2.524

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.