Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams.

S Reinhardt1, M Hillbrand, J J Wilkens, W Assmann.   

Abstract

PURPOSE: Dose verification in highly conformal radiation therapy such as IMRT or proton therapy can benefit from the high spatial resolution offered by radio-chromic films such as Gafchromic EBT or EBT2. Recently, a new generation of these films, EBT3, has become available. The composition and thickness of the sensitive layer are the same as for the previous EBT2 films. The most important change is the symmetric layer configuration to eliminate side orientation dependence, which is reported for EBT2 films.
METHODS: The general film characteristics such as sensitivity to read-out orientation and postexposure darkening evolution of the new EBT3 film are evaluated. Film response has been investigated in clinical photon and proton beams and compared to former EBT2 films. Quenching effects in the proton Bragg peak region have been studied for both, EBT2 and EBT3 films.
RESULTS: The general performance of EBT3 is comparable to EBT2, and the orientation dependence with respect to film side is completely eliminated in EBT3 films. Response differences of EBT2 and EBT3 films are of the same order of magnitude as batch-to-batch variations observed for EBT2 films. No significant difference has been found for both generations of EBT films between photon and proton exposure. Depth dose measurements of EBT2 and EBT3 show an excellent agreement, though underestimating dose by up to 20% in the Bragg peak region.
CONCLUSIONS: The symmetric configuration of EBT3 presents a major improvement for film handling. EBT3 has similar dosimetric performance as its precursor EBT2 and can, thus, be applied to dose verification in IMRT in the same way. For dose verification in proton therapy the underresponse in the Bragg peak region has to be taken into account.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22894450     DOI: 10.1118/1.4737890

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  46 in total

1.  The influence of the channel size on the reduction of side effects in microchannel proton therapy.

Authors:  Stefanie Girst; Christoph Greubel; Judith Reindl; Christian Siebenwirth; Olga Zlobinskaya; Günther Dollinger; Thomas E Schmid
Journal:  Radiat Environ Biophys       Date:  2015-05-09       Impact factor: 1.925

2.  Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

Authors:  S Reinhardt; M Würl; C Greubel; N Humble; J J Wilkens; M Hillbrand; A Mairani; W Assmann; K Parodi
Journal:  Radiat Environ Biophys       Date:  2015-01-09       Impact factor: 1.925

3.  Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source.

Authors:  Konrad Buchauer; Guido Henke; Hans Schiefer; Ludwig Plasswilm
Journal:  Strahlenther Onkol       Date:  2014-08-01       Impact factor: 3.621

4.  Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film.

Authors:  Maria F Chan; Chin-Cheng Chen; Chengyu Shi; Jingdong Li; Xiaoli Tang; Xiang Li; Dennis Mah
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2017-05-16

5.  Comprehensive quality assurance phantom for the small animal radiation research platform (SARRP).

Authors:  M Jermoumi; H Korideck; M Bhagwat; P Zygmanski; G M Makrigiogos; R I Berbeco; R C Cormack; W Ngwa
Journal:  Phys Med       Date:  2015-05-08       Impact factor: 2.685

6.  Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Authors:  Khaled Soliman; Marouf Adili; Abdullah Alrushoud
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-07

7.  An Anthropomorphic Head and Neck Quality Assurance Phantom for Credentialing of Intensity-Modulated Proton Therapy.

Authors:  Daniela Branco; Paige Taylor; Xiaodong Zhang; Heng Li; Michele Guindani; David Followill
Journal:  Int J Part Ther       Date:  2018-03-21

8.  A study on rectal dose measurement in phantom and in vivo using Gafchromic EBT3 film in IMRT and CyberKnife treatments of carcinoma of prostate.

Authors:  K Ganapathy; P G G Kurup; V Murali; M Muthukumaran; S Balaji Subramanian; J Velmurugan
Journal:  J Med Phys       Date:  2013-07

9.  The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT.

Authors:  Jung-In Kim; So-Yeon Park; Hak Jae Kim; Jin Ho Kim; Sung-Joon Ye; Jong Min Park
Journal:  Radiat Oncol       Date:  2014-07-28       Impact factor: 3.481

10.  Clinical utility of Gafchromic film in an MRI-guided linear accelerator.

Authors:  Ilma Xhaferllari; Joshua P Kim; Ruchira Liyanage; Chang Liu; Dongsu Du; Anthony Doemer; Indrin J Chetty; Ning Wen
Journal:  Radiat Oncol       Date:  2021-06-26       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.