Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source.

Konrad Buchauer1, Guido Henke, Hans Schiefer, Ludwig Plasswilm.   

Abstract

BACKGROUND: Conical surface applicators with an Ir-192 high-dose-rate brachytherapy source are a common modality for the treatment of non-melanomatous skin cancer with high tumour control rates. Surface dose characterisation of the Varian Varisource GammaMed+ IX afterloader vertical type surface applicators is performed two dimensionally using high-resolution film dosimetry. AIM: The focus of this study was to determine if Varian surface applicators with a vertical source suffer from the dose distribution irregularities reported for comparable applicators. Our goal was to evaluate if the irregularities found affected treatment and dose output verification procedures.
METHODS: Ionisation chamber-based verification of applicator output was established according to guidelines provided by the manufacturer. For additional measurement of surface dose Gafchromic EBT3 film dosimetry was used. The term "therapeutic dose" was defined as 85% of the prescribed dose level.
RESULTS: For the 10 different applicator inserts evaluated, cold spots were observed. Mean cold spot size was 2.0 mm × 3.6 mm (± 0.6 mm). The cold spots were dosimetrically well below 85% of the prescribed dose. The cold spot was situated 2.2 mm (1.4-2.7 mm) unilaterally from the central axis and caused general asymmetry in the dose profiles intersecting the cold spot area. A source tilt of approximately 8° (± 1°) was determined for the source used for irradiation.
CONCLUSIONS: A central underdosed area exceeding 15 % of the prescribed dose has not been previously reported. Source tilt was observed and found to affect clinical use and possibly treatment outcome in applicators using a vertically arranged source. Surface applicators with a vertically orientated source were subject to dose irregularities that could impact on chamber-based applicator output verification procedures. We recommend film dosimetry-backed applicator commissioning to avoid systematic errors.

Entities:  

Mesh:

Year:  2014        PMID: 25081332     DOI: 10.1007/s00066-014-0713-2

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  20 in total

1.  Optimal flattening filter shape of a surface brachytherapy applicator.

Authors:  Robert Jeraj; Attila Sarvary; Tomas Kron
Journal:  Phys Med Biol       Date:  2002-03-07       Impact factor: 3.609

2.  Technique for routine output verification of Leipzig applicators with a well chamber.

Authors:  J Pérez-Calatayud; D Granero; F Ballester; V Crispín; R Van der Laarse
Journal:  Med Phys       Date:  2006-01       Impact factor: 4.071

3.  Calibration of 192Ir high-dose-rate afterloading systems.

Authors:  S J Goetsch; F H Attix; D W Pearson; B R Thomadsen
Journal:  Med Phys       Date:  1991 May-Jun       Impact factor: 4.071

4.  GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

Authors:  Sigrun Saur; Jomar Frengen
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

5.  Dose calculation for photon-emitting brachytherapy sources with average energy higher than 50 keV: report of the AAPM and ESTRO.

Authors:  Jose Perez-Calatayud; Facundo Ballester; Rupak K Das; Larry A Dewerd; Geoffrey S Ibbott; Ali S Meigooni; Zoubir Ouhib; Mark J Rivard; Ron S Sloboda; Jeffrey F Williamson
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

6.  GAFCHROMIC EBT photospectral dose response dependence on temperature and implications for flat bed scanning.

Authors:  Konrad Buchauer; Elmar Hillbrand; Alexander de Vries
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

7.  Radiochromic film dosimetry of HDR (192)Ir source radiation fields.

Authors:  Saad Aldelaijan; Huriyyah Mohammed; Nada Tomic; Li-Heng Liang; Francois Deblois; Arman Sarfehnia; Wamied Abdel-Rahman; Jan Seuntjens; Slobodan Devic
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

8.  [The indications for and results of HDR afterloading therapy in diseases of the skin and mucosa with standardized surface applicators (the Leipzig applicator)].

Authors:  A Köhler-Brock; W Prager; S Pohlmann; S Kunze
Journal:  Strahlenther Onkol       Date:  1999-04       Impact factor: 3.621

9.  Treatment of skin carcinomas of the face by high-dose-rate brachytherapy and custom-made surface molds.

Authors:  B Guix; F Finestres; J Tello; C Palma; A Martinez; J Guix; R Guix
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-04-01       Impact factor: 7.038

10.  Locally advanced verrucous carcinoma of the oral cavity: treatment using customized mold HDR brachytherapy instead of hemi-maxillectomy.

Authors:  K M J van Gestel; D J M Buurman; R Pijls; P A W H Kessler; P L A van den Ende; A L Hoffmann; E G C Troost
Journal:  Strahlenther Onkol       Date:  2013-08-22       Impact factor: 3.621

View more
  1 in total

1.  Impact of source position on high-dose-rate skin surface applicator dosimetry.

Authors:  Jeho Jeong; Christopher A Barker; Marco Zaider; Gil'ad N Cohen
Journal:  Brachytherapy       Date:  2016-06-01       Impact factor: 2.362

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.