Cardiac computed tomography angiography with automatic tube potential selection: effects on radiation dose and image quality.

Brian B Ghoshhajra1, Leif-Christopher Engel, Mihály Károlyi, Manavjot Singh Sidhu, Bryan Wai, Mitya Barreto, Uthamalingam Shanmugam, Udo Hoffmann, Thomas J Brady, Manudeep Kalra, Suhny Abbara.   

Abstract

PURPOSE: Automatic exposure control (AEC) algorithms are widely available in coronary computed tomography angiography (CTA) and have been shown to reduce radiation doses by adjusting tube current to patient size. However, the effects of anthropometry-based automatic potential selection (APS) on image quality and radiation dose are unknown. We sought to investigate the effect of an APS algorithm on coronary CTA radiation dose and image quality.
MATERIALS AND METHODS: For this retrospective case-control study we selected 38 patients who had undergone coronary CTA for coronary artery assessment in whom tube potential and tube current were selected automatically by a combined automatic tube potential and tube current selection algorithm (APS-AEC) and compared them with 38 controls for whom tube voltage was selected according to standard body mass index (BMI) cutoffs and tube current was selected using automatic exposure control (BMI-AEC). Controls were matched for BMI, heart rate, heart rhythm, sex, acquisition mode, and indication for cardiac CTA. Image quality was assessed as contrast-to-noise ratio and signal-to-noise ratio in the proximal coronary arteries. Subjective reader assessment was also made. Total radiation dose (volume-weighted computed tomography dose index) was measured and compared between the 2 groups. In the study group, comparison was made with conventional BMI-guided prior protocols (site protocols and Society of Cardiovascular Computed Tomography recommendations) through disagreement analysis.
RESULTS: The APS-AEC cases received 29.8% lower overall radiation dose compared with controls (P=not significant). APS-AEC resulted in a significantly higher signal-to-noise ratio of the proximal coronary arteries (P<0.01) and contrast-to-noise ratio of the left main (P=0.01). In the study cases, the APS resulted in a change in tube potential versus site protocols and Society of Cardiovascular Computed Tomography recommendations in 45% (n=17) and 50% (n=19) of patients, respectively.
CONCLUSION: Automated tube potential selection software resulted in significantly improved objective image quality versus standard BMI-based methods of tube potential selection, without increased radiation doses.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 22847638     DOI: 10.1097/RTI.0b013e3182631e8a

Source DB:  PubMed          Journal:  J Thorac Imaging        ISSN: 0883-5993            Impact factor:   3.000


  23 in total

1.  Application of intelligent optimal kV scanning technology (CARE kV) in dual-source computed tomography (DSCT) coronary angiography.

Authors:  Jun Zhang; Shaolei Kang; Dan Han; Xiaojie Xie; Yaming Deng
Journal:  Int J Clin Exp Med       Date:  2015-10-15

2.  Radiation dose reduction in pediatric cardiac computed tomography: experience from a tertiary medical center.

Authors:  Brian B Ghoshhajra; Ashley M Lee; Leif-Christopher Engel; Csilla Celeng; Mannudeep K Kalra; Thomas J Brady; Udo Hoffmann; Sjirk J Westra; Suhny Abbara
Journal:  Pediatr Cardiol       Date:  2013-07-20       Impact factor: 1.655

3.  Coronary CT Angiography Using Low Iodine Delivery Rate and Tube Voltage Determined by Body Mass Index: Superiority in Clinical Practice.

Authors:  Wang Yuan; Ting-Ting Qu; Hui-Juan Wang; Mei-Yu Wang; Yuan Qu; Gang Niu; Jian Yang
Journal:  Curr Med Sci       Date:  2019-10-14

4.  Image quality and radiation reduction of 320-row area detector CT coronary angiography with optimal tube voltage selection and an automatic exposure control system: comparison with body mass index-adapted protocol.

Authors:  Jiyeon Lim; Eun-Ah Park; Whal Lee; Hackjoon Shim; Jin Wook Chung
Journal:  Int J Cardiovasc Imaging       Date:  2015-01-22       Impact factor: 2.357

5.  Image quality and diagnostic value of ultra low-voltage, ultra low-contrast coronary CT angiography.

Authors:  Chong-Fu Jia; Jie Zhong; Xin-Yi Meng; Xi-Xia Sun; Zhi-Qiang Yang; Yu-Jie Zou; Xiang-Yue Wang; Shuang Pan; Da Yin; Zhao-Qian Wang
Journal:  Eur Radiol       Date:  2019-03-19       Impact factor: 5.315

6.  Coronary CT angiography in the emergency department utilizing second and third generation dual source CT.

Authors:  Nandini M Meyersohn; Balint Szilveszter; Pedro V Staziaki; Jan-Erik Scholtz; Richard A P Takx; Udo Hoffmann; Brian B Ghoshhajra
Journal:  J Cardiovasc Comput Tomogr       Date:  2017-03-22

7.  Assessment of image quality and radiation dose of prospectively ECG-triggered adaptive dual-source coronary computed tomography angiography (cCTA) with arrhythmia rejection algorithm in systole versus diastole: a retrospective cohort study.

Authors:  Ashley M Lee; Jonathan Beaudoin; Leif-Christopher Engel; Manavjot S Sidhu; Suhny Abbara; Thomas J Brady; Udo Hoffmann; Brian B Ghoshhajra
Journal:  Int J Cardiovasc Imaging       Date:  2013-03-24       Impact factor: 2.357

8.  Feasibility of a radiation dose conserving CT protocol for myocardial function assessment.

Authors:  A Pursnani; A Lee; T Mayrhofer; M Panagia; U Sharma; S Abbara; U Hoffmann; B B Ghoshhajra
Journal:  Br J Radiol       Date:  2014-06-02       Impact factor: 3.039

9.  Topogram-based automated selection of the tube potential and current in thoraco-abdominal trauma CT - a comparison to fixed kV with mAs modulation alone.

Authors:  Claudia Frellesen; Wenzel Stock; J Matthias Kerl; Thomas Lehnert; Julian L Wichmann; Christoph Nau; Emanuel Geiger; Sebastian Wutzler; Martin Beeres; Boris Schulz; Boris Bodelle; Hanns Ackermann; Thomas J Vogl; Ralf W Bauer
Journal:  Eur Radiol       Date:  2014-05-10       Impact factor: 5.315

Review 10.  Role of Cardiac PET in Clinical Practice.

Authors:  Brian M Salata; Parmanand Singh
Journal:  Curr Treat Options Cardiovasc Med       Date:  2017-11-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.