Intrafraction verification of gated RapidArc by using beam-level kilovoltage X-ray images.

Ruijiang Li1, Edward Mok, Daniel T Chang, Megan Daly, Billy W Loo, Maximilian Diehn, Quynh-Thu Le, Albert Koong, Lei Xing.   

Abstract

PURPOSE: To verify the geometric accuracy of gated RapidArc treatment using kV images acquired during dose delivery. METHODS AND MATERIALS: Twenty patients were treated using the gated RapidArc technique with a Varian TrueBeam STx linear accelerator. One to 7 metallic fiducial markers were implanted inside or near the tumor target before treatment simulation. For patient setup and treatment verification purposes, the internal target volume (ITV) was created, corresponding to each implanted marker. The gating signal was generated from the Real-time Position Management (RPM) system. At the beginning of each fraction, individualized respiratory gating amplitude thresholds were set based on fluoroscopic image guidance. During the treatment, we acquired kV images immediately before MV beam-on at every breathing cycle, using the on-board imaging system. After the treatment, all implanted markers were detected, and their 3-dimensional (3D) positions in the patient were estimated using software developed in-house. The distance from the marker to the corresponding ITV was calculated for each patient by averaging over all markers and all fractions.
RESULTS: The average 3D distance between the markers and their ITVs was 0.8 ± 0.5 mm (range, 0-1.7 mm) and was 2.1 ± 1.2 mm at the 95th percentile (range, 0-3.8 mm). On average, a left-right margin of 0.6 mm, an anterior-posterior margin of 0.8 mm, and a superior-inferior margin of 1.5 mm is required to account for 95% of the intrafraction uncertainty in RPM-based RapidArc gating.
CONCLUSION: To our knowledge, this is the first clinical report of intrafraction verification of respiration-gated RapidArc treatment in stereotactic ablative radiation therapy. For some patients, the markers deviated significantly from the ITV by more than 2 mm at the beginning of the MV beam-on. This emphasizes the need for gating techniques with beam-on/-off controlled directly by the actual position of the tumor target instead of external surrogates such as RPM.
Copyright © 2012 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22554582      PMCID: PMC4476315          DOI: 10.1016/j.ijrobp.2012.03.006

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  20 in total

1.  High retention and safety of percutaneously implanted endovascular embolization coils as fiducial markers for image-guided stereotactic ablative radiotherapy of pulmonary tumors.

Authors:  Julian C Hong; Yao Yu; Aarti K Rao; Sonja Dieterich; Peter G Maxim; Quynh-Thu Le; Maximilian Diehn; Daniel Y Sze; Nishita Kothary; Billy W Loo
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-08-01       Impact factor: 7.038

2.  Stereotactic body radiation therapy for early-stage non-small-cell lung carcinoma: four-year results of a prospective phase II study.

Authors:  Achilles J Fakiris; Ronald C McGarry; Constantin T Yiannoutsos; Lech Papiez; Mark Williams; Mark A Henderson; Robert Timmerman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-27       Impact factor: 7.038

3.  Four-dimensional inverse treatment planning with inclusion of implanted fiducials in IMRT segmented fields.

Authors:  Yunzhi Ma; Louis Lee; O Keshet; Paul Keall; Lei Xing
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

4.  Gated treatment delivery verification with on-line megavoltage fluoroscopy.

Authors:  An Tai; James D Christensen; Elizabeth Gore; Ali Khamene; Thomas Boettger; X Allen Li
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-02-03       Impact factor: 7.038

5.  A Bayesian approach to real-time 3D tumor localization via monoscopic x-ray imaging during treatment delivery.

Authors:  Ruijiang Li; Benjamin P Fahimian; Lei Xing
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

6.  Prostate intrafraction motion assessed by simultaneous kilovoltage fluoroscopy at megavoltage delivery I: clinical observations and pattern analysis.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-06-25       Impact factor: 7.038

7.  Physical aspects of a real-time tumor-tracking system for gated radiotherapy.

Authors:  H Shirato; S Shimizu; T Kunieda; K Kitamura; M van Herk; K Kagei; T Nishioka; S Hashimoto; K Fujita; H Aoyama; K Tsuchiya; K Kudo; K Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-11-01       Impact factor: 7.038

8.  A feasibility study of markerless fluoroscopic gating for lung cancer radiotherapy using 4DCT templates.

Authors:  Ruijiang Li; John H Lewis; Laura I Cerviño; Steve B Jiang
Journal:  Phys Med Biol       Date:  2009-09-24       Impact factor: 3.609

9.  Stereotactic body radiation therapy for liver tumors: impact of daily setup corrections and day-to-day anatomic variations on dose in target and organs at risk.

Authors:  Alejandra Méndez Romero; Roel Th Zinkstok; Wouter Wunderink; Rob M van Os; Hans Joosten; Yvette Seppenwoolde; Peter J C M Nowak; Rene P Brandwijk; Cornelis Verhoef; Jan N M IJzermans; Peter C Levendag; Ben J M Heijmen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-04-20       Impact factor: 7.038

10.  Interfractional uncertainty in the treatment of pancreatic cancer with radiation.

Authors:  Priya Jayachandran; A Yuriko Minn; Jacques Van Dam; Jeffrey A Norton; Albert C Koong; Daniel T Chang
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-10-30       Impact factor: 7.038

View more
  10 in total

1.  First study of on-treatment volumetric imaging during respiratory gated VMAT.

Authors:  Kihwan Choi; Lei Xing; Albert Koong; Ruijiang Li
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

2.  Clinical implementation of intrafraction cone beam computed tomography imaging during lung tumor stereotactic ablative radiation therapy.

Authors:  Ruijiang Li; Bin Han; Bowen Meng; Peter G Maxim; Lei Xing; Albert C Koong; Maximilian Diehn; Billy W Loo
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-10-08       Impact factor: 7.038

3.  Nonisocentric treatment strategy for breast radiation therapy: a proof of concept study.

Authors:  Ruijiang Li; Lei Xing; Kathleen C Horst; Karl Bush
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-03-15       Impact factor: 7.038

4.  Imaged-guided liver stereotactic body radiotherapy using VMAT and real-time adaptive tumor gating. Concerns about technique and preliminary clinical results.

Authors:  Carmen Llacer-Moscardo; Olivier Riou; David Azria; Ludovic Bedos; Norbert Ailleres; Francois Quenet; Philippe Rouanet; Marc Ychou; Pascal Fenoglietto
Journal:  Rep Pract Oncol Radiother       Date:  2016-09-05

5.  Simultaneous modulated accelerated radiation therapy for esophageal cancer: a feasibility study.

Authors:  Wu-Zhe Zhang; Jian-Zhou Chen; De-Rui Li; Zhi-Jian Chen; Hong Guo; Ting-Ting Zhuang; Dong-Sheng Li; Ming-Zhen Zhou; Chuang-Zhen Chen
Journal:  World J Gastroenterol       Date:  2014-10-14       Impact factor: 5.742

6.  Intrafractional 3D localization using kilovoltage digital tomosynthesis for sliding-window intensity modulated radiation therapy.

Authors:  Pengpeng Zhang; Margie Hunt; Hai Pham; Grace Tang; Gig Mageras
Journal:  Phys Med Biol       Date:  2015-08-25       Impact factor: 3.609

7.  Kilovoltage Imaging of Implanted Fiducials to Monitor Intrafraction Motion With Abdominal Compression During Stereotactic Body Radiation Therapy for Gastrointestinal Tumors.

Authors:  Ellen Yorke; Ying Xiong; Qian Han; Pengpeng Zhang; Gikas Mageras; Michael Lovelock; Hai Pham; Jian-Ping Xiong; Karyn A Goodman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-11-18       Impact factor: 7.038

8.  Local control rates in stereotactic body radiotherapy (SBRT) of lung metastases associated with the biologically effective dose.

Authors:  Daniel Zucca Aparicio; Ovidio Hernando Requejo; Miguel Ángel de la Casa de Julián; Carmen Rubio Rodríguez; Pedro Fernández Letón
Journal:  Rep Pract Oncol Radiother       Date:  2019-01-22

9.  Simultaneous MV-kV imaging for intrafractional motion management during volumetric-modulated arc therapy delivery.

Authors:  Margie A Hunt; Mark Sonnick; Hai Pham; Rajesh Regmi; Jian-ping Xiong; Daniel Morf; Gig S Mageras; Michael Zelefsky; Pengpeng Zhang
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

10.  Stereotactic body radiation therapy for adrenal gland metastases: Outcomes and toxicity.

Authors:  Diego A S Toesca; Amanda J Koong; Rie von Eyben; Albert C Koong; Daniel T Chang
Journal:  Adv Radiat Oncol       Date:  2018-05-24
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.