Literature DB >> 19386435

Stereotactic body radiation therapy for liver tumors: impact of daily setup corrections and day-to-day anatomic variations on dose in target and organs at risk.

Alejandra Méndez Romero1, Roel Th Zinkstok, Wouter Wunderink, Rob M van Os, Hans Joosten, Yvette Seppenwoolde, Peter J C M Nowak, Rene P Brandwijk, Cornelis Verhoef, Jan N M IJzermans, Peter C Levendag, Ben J M Heijmen.   

Abstract

PURPOSE: To assess day-to-day differences between planned and delivered target volume (TV) and organ-at-risk (OAR) dose distributions in liver stereotactic body radiation therapy (SBRT), and to investigate the dosimetric impact of setup corrections. METHODS AND MATERIALS: For 14 patients previously treated with SBRT, the planning CT scan and three treatment scans (one for each fraction) were included in this study. For each treatment scan, two dose distributions were calculated: one using the planned setup for the body frame (no correction), and one using the clinically applied (corrected) setup derived from measured tumor displacements. Per scan, the two dose distributions were mutually compared, and the clinically delivered distribution was compared with planning. Doses were recalculated in equivalent 2-Gy fraction doses. Statistical analysis was performed with the linear mixed model.
RESULTS: With setup corrections, the mean loss in TV coverage relative to planning was 1.7%, compared with 6.8% without corrections. For calculated equivalent uniform doses, these figures were 2.3% and 15.5%, respectively. As for the TV, mean deviations of delivered OAR doses from planning were small (between -0.4 and +0.3 Gy), but the spread was much larger for the OARs. In contrast to the TV, the mean impact of setup corrections on realized OAR doses was close to zero, with large positive and negative exceptions.
CONCLUSIONS: Daily correction of the treatment setup is required to obtain adequate TV coverage. Because of day-to-day patient anatomy changes, large deviations in OAR doses from planning did occur. On average, setup corrections had no impact on these doses. Development of new procedures for image guidance and adaptive protocols is warranted.

Entities:  

Mesh:

Year:  2009        PMID: 19386435     DOI: 10.1016/j.ijrobp.2008.12.020

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  11 in total

Review 1.  [Radiotherapy of hepatic metastases].

Authors:  S E Combs; K K Herfarth; D Habermehl; J Debus
Journal:  Chirurg       Date:  2010-06       Impact factor: 0.955

2.  Accumulated dose in liver stereotactic body radiotherapy: positioning, breathing, and deformation effects.

Authors:  Michael Velec; Joanne L Moseley; Tim Craig; Laura A Dawson; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-12-28       Impact factor: 7.038

Review 3.  Adaptive management of liver cancer radiotherapy.

Authors:  Kristy K Brock; Laura A Dawson
Journal:  Semin Radiat Oncol       Date:  2010-04       Impact factor: 5.934

4.  Imaged-guided liver stereotactic body radiotherapy using VMAT and real-time adaptive tumor gating. Concerns about technique and preliminary clinical results.

Authors:  Carmen Llacer-Moscardo; Olivier Riou; David Azria; Ludovic Bedos; Norbert Ailleres; Francois Quenet; Philippe Rouanet; Marc Ychou; Pascal Fenoglietto
Journal:  Rep Pract Oncol Radiother       Date:  2016-09-05

5.  Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images.

Authors:  Ruijiang Li; Edward Mok; Bin Han; Albert Koong; Lei Xing
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

6.  Intrafraction verification of gated RapidArc by using beam-level kilovoltage X-ray images.

Authors:  Ruijiang Li; Edward Mok; Daniel T Chang; Megan Daly; Billy W Loo; Maximilian Diehn; Quynh-Thu Le; Albert Koong; Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-05-02       Impact factor: 7.038

7.  Interfraction liver shape variability and impact on GTV position during liver stereotactic radiotherapy using abdominal compression.

Authors:  Cynthia L Eccles; Laura A Dawson; Joanne L Moseley; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-13       Impact factor: 7.038

8.  Dose escalated liver stereotactic body radiation therapy at the mean respiratory position.

Authors:  Michael Velec; Joanne L Moseley; Laura A Dawson; Kristy K Brock
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-07-08       Impact factor: 7.038

9.  Intrafractional fiducial marker position variations in stereotactic liver radiotherapy during voluntary deep inspiration breath-hold.

Authors:  Line Bjerregaard Stick; Ivan Richter Vogelius; Signe Risum; Mirjana Josipovic
Journal:  Br J Radiol       Date:  2020-09-11       Impact factor: 3.039

10.  Dosimetric differences among volumetric modulated arc radiotherapy (RapidArc) plans based on different target volumes in radiotherapy of hepatocellular carcinoma.

Authors:  GuanZhong Gong; Yong Yin; YuJie Guo; TongHai Liu; JinHu Chen; Jie Lu; ChangSheng Ma; Tao Sun; Tong Bai; GuiFang Zhang; DengWang Li; RuoZheng Wang
Journal:  J Radiat Res       Date:  2012-08-21       Impact factor: 2.724

View more