Performance evaluation of a pediatric viscous impeller pump for Fontan cavopulmonary assist.

Guruprasad A Giridharan1, Steven C Koenig, Jeffrey Kennington, Michael A Sobieski, Jun Chen, Steven H Frankel, Mark D Rodefeld.   

Abstract

OBJECTIVE: The anatomic and physiologic constraints for pediatric cavopulmonary assist differ markedly from adult Fontan circulations owing to smaller vessel sizes and risk of elevated pulmonary resistance. In this study, hemodynamic and hemolysis performance of a catheter-based viscous impeller pump (VIP) to power the Fontan circulation is assessed at a pediatric scale (∼15 kg) and performance range (0-30 mm Hg).
METHODS: Computer simulation and mock circulation studies were conducted to assess the hydraulic performance, acute hemodynamic response to different levels VIP support, and the potential for vena caval collapse. Computational fluid dynamics simulations were used to estimate VIP hydraulic performance, shear rates, and potential for hemolysis. Hemolysis was quantified in a mock loop with fresh bovine blood.
RESULTS: A VIP augmented 4-way total cavopulmonary connection flow at pediatric scales and restored systemic pressures and flows to biventricular values, without causing flow obstruction or suction. VIP generated flows up to 4.1 L/min and pressure heads of up to 38 mm Hg at 11,000 rpm. Maximal shear rate was 160 Pa, predicting low hemolysis risk. Observed hemolysis was low with plasma free hemoglobin of 11.4 mg · dL(-1) · h(-1).
CONCLUSIONS: A VIP will augment Fontan cavopulmonary flow in the proper pressure and flow ranges, with low hemolysis risk under more stringent pediatric scale and physiology compared with adult scale. This technology may be developed to simultaneously reduce systemic venous pressure and improve cardiac output after stage 2 or 3 Fontan repair. It may serve to compress surgical staging, lessening the pathophysiologic burden of repair.
Copyright © 2013 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2012        PMID: 22421403      PMCID: PMC3525746          DOI: 10.1016/j.jtcvs.2012.01.082

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  26 in total

1.  Changes in venous return parameters associated with univentricular Fontan circulations.

Authors:  L Macé; P Dervanian; A Bourriez; G M Mazmanian; V Lambert; J Losay; J Y Neveux
Journal:  Am J Physiol Heart Circ Physiol       Date:  2000-11       Impact factor: 4.733

2.  Cavopulmonary assist: circulatory support for the univentricular Fontan circulation.

Authors:  Mark D Rodefeld; Jack H Boyd; Cynthia D Myers; Brian J LaLone; Alex J Bezruczko; Andrew W Potter; John W Brown
Journal:  Ann Thorac Surg       Date:  2003-12       Impact factor: 4.330

3.  Characterization of an adult mock circulation for testing cardiac support devices.

Authors:  George M Pantalos; Steven C Koenig; Kevin J Gillars; Guruprasad A Giridharan; Dan L Ewert
Journal:  ASAIO J       Date:  2004 Jan-Feb       Impact factor: 2.872

4.  Basal pulmonary vascular resistance and nitric oxide responsiveness late after Fontan-type operation.

Authors:  S Khambadkone; J Li; M R de Leval; S Cullen; J E Deanfield; A N Redington
Journal:  Circulation       Date:  2003-06-23       Impact factor: 29.690

5.  HEART: an automated beat-to-beat cardiovascular analysis package using Matlab.

Authors:  M J Mark J Schroeder; Bill Perreault; D L Daniel L Ewert; S C Steven C Koenig
Journal:  Comput Biol Med       Date:  2004-07       Impact factor: 4.589

6.  Cavopulmonary assist in the neonate: an alternative strategy for single-ventricle palliation.

Authors:  Mark D Rodefeld; Jack H Boyd; Cynthia D Myers; Robert G Presson; Wiltz W Wagner; John W Brown
Journal:  J Thorac Cardiovasc Surg       Date:  2004-03       Impact factor: 5.209

7.  Blood trauma testing for mechanical circulatory support devices.

Authors:  Guruprasad A Giridharan; Michael A Sobieski; Mickey Ising; Mark S Slaughter; Steven C Koenig
Journal:  Biomed Instrum Technol       Date:  2011 Jul-Aug

8.  Outcomes after the Norwood operation in neonates with critical aortic stenosis or aortic valve atresia.

Authors:  David A Ashburn; Brian W McCrindle; Christo I Tchervenkov; Marshall L Jacobs; Gary K Lofland; Edward L Bove; Thomas L Spray; William G Williams; Eugene H Blackstone
Journal:  J Thorac Cardiovasc Surg       Date:  2003-05       Impact factor: 5.209

9.  Ventricular afterload and ventricular work in fontan circulation: comparison with normal two-ventricle circulation and single-ventricle circulation with blalock-taussig shunts.

Authors:  Hideaki Senzaki; Satoshi Masutani; Jun Kobayashi; Toshiki Kobayashi; Nozomu Sasaki; Haruhiko Asano; Shunei Kyo; Yuji Yokote; Akira Ishizawa
Journal:  Circulation       Date:  2002-06-18       Impact factor: 29.690

10.  Integrated data acquisition system for medical device testing and physiology research in compliance with good laboratory practices.

Authors:  Steven C Koenig; Cary Woolard; Guy Drew; Lauren Unger; Kevin Gillars; Dan Ewert; Laman Gray; George Pantalos
Journal:  Biomed Instrum Technol       Date:  2004 May-Jun
View more
  16 in total

1.  Cavopulmonary assist for the failing Fontan circulation: impact of ventricular function on mechanical support strategy.

Authors:  Guruprasad A Giridharan; Mickey Ising; Michael A Sobieski; Steven C Koenig; Jun Chen; Steven Frankel; Mark D Rodefeld
Journal:  ASAIO J       Date:  2014 Nov-Dec       Impact factor: 2.872

2.  Experimental Investigation of the Effect of Non-Newtonian Behavior of Blood Flow in the Fontan Circulation.

Authors:  Andrew L Cheng; Niema M Pahlevan; Derek G Rinderknecht; John C Wood; Morteza Gharib
Journal:  Eur J Mech B Fluids       Date:  2017-12-27       Impact factor: 2.183

3.  In Vitro Ultrasound Measurements of Powered and Unpowered Total Cavopulmonary Connection.

Authors:  Bp Iliff; Aem Kerlo; J Chen; Md Rodefeld; Cj Goergen
Journal:  Austin J Biomed Eng       Date:  2014

Review 4.  Recent advances in computational methodology for simulation of mechanical circulatory assist devices.

Authors:  Alison L Marsden; Yuri Bazilevs; Christopher C Long; Marek Behr
Journal:  Wiley Interdiscip Rev Syst Biol Med       Date:  2014-01-21

Review 5.  Current approaches to device implantation in pediatric and congenital heart disease patients.

Authors:  Jacob R Miller; Timothy S Lancaster; Pirooz Eghtesady
Journal:  Expert Rev Cardiovasc Ther       Date:  2015-03-03

Review 6.  Ventricular assist device use in congenital heart disease with a comparison to heart transplant.

Authors:  Jacob R Miller; Pirooz Eghtesady
Journal:  J Comp Eff Res       Date:  2014-09       Impact factor: 1.744

7.  A paired membrane umbrella double-lumen cannula ensures consistent cavopulmonary assistance in a Fontan sheep model.

Authors:  Dongfang Wang; Guodong Gao; Mark Plunkett; Guangfeng Zhao; Stephen Topaz; Cherry Ballard-Croft; Joseph B Zwischenberger
Journal:  J Thorac Cardiovasc Surg       Date:  2014-05-16       Impact factor: 5.209

8.  Using a Novel In Vitro Fontan Model and Condition-Specific Real-Time MRI Data to Examine Hemodynamic Effects of Respiration and Exercise.

Authors:  Michael Tree; Zhenglun Alan Wei; Phillip M Trusty; Vrishank Raghav; Mark Fogel; Kevin Maher; Ajit Yoganathan
Journal:  Ann Biomed Eng       Date:  2017-10-24       Impact factor: 3.934

Review 9.  An overview of mechanical circulatory support in single-ventricle patients.

Authors:  Jacob R Miller; Timothy S Lancaster; Connor Callahan; Aaron M Abarbanell; Pirooz Eghtesady
Journal:  Transl Pediatr       Date:  2018-04

Review 10.  Computational modeling and engineering in pediatric and congenital heart disease.

Authors:  Alison L Marsden; Jeffrey A Feinstein
Journal:  Curr Opin Pediatr       Date:  2015-10       Impact factor: 2.856

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.