Set-up errors and planning target volume margins in head and neck cancer radiotherapy: a clinical study of image guidance with on-line cone-beam computed tomography.

Francesco Dionisi1, Mauro Filippo Palazzi, Francesco Bracco, Maria Grazia Brambilla, Claudia Carbonini, Diego Dario Asnaghi, Angelo Filippo Monti, Alberto Torresin.   

Abstract

BACKGROUND: Set-up errors represent a source of uncertainty in head and neck (H&N) cancer radiotherapy. The present study evaluated set-up accuracy with the use of cone-beam computed tomography (CBCT) in order to establish the proper clinical target volume (CTV) to planning target volume (PTV) margins to be adopted.
METHODS: Local set-up accuracy was analysed for 44 H&N cancer patients since the implementation of CBCT. An on-line correction protocol was adopted, with the first 3 scans used to correct systematic errors with a 3-mm action level. The overall mean displacement (M), the population systematic (Σ) and random (σ) errors and the 3D vector length were calculated. PTV margins were calculated according to the van Herk formula (2.5Σ + 0.7σ).
RESULTS: A total of 420 CBCT scans were analysed. A systematic correction was needed in 43% of patients. The value of M was <1 mm in all directions; the values of Σ and σ ranged over 1-1.2 and 1.4-1.9 mm, respectively. Pre-correction PTV margins were 3.48, 4.08 and 4.33 mm along the 3 axes. The PTV margins calculated after online correction were <2.5 mm in all directions.
CONCLUSIONS: Kilovoltage CBCT is effective in evaluating set-up accuracy in H&amp;N patients. CTV-PTV margins of 5 mm are safe and are currently adopted at our centre; however, some special situations, such as re-irradiation or the close proximity of organs at risk and high-dose regions, could benefit from daily image registration and lower (i.e., 3 mm) margins.

Entities:  

Mesh:

Year:  2012        PMID: 22389140     DOI: 10.1007/s10147-012-0395-7

Source DB:  PubMed          Journal:  Int J Clin Oncol        ISSN: 1341-9625            Impact factor:   3.402


  25 in total

1.  Setup error in radiotherapy: on-line correction using electronic kilovoltage and megavoltage radiographs.

Authors:  L Pisani; D Lockman; D Jaffray; D Yan; A Martinez; J Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-06-01       Impact factor: 7.038

2.  When should systematic patient positioning errors in radiotherapy be corrected?

Authors:  Thomas Bortfeld; Marcel van Herk; Steve B Jiang
Journal:  Phys Med Biol       Date:  2002-12-07       Impact factor: 3.609

3.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

4.  Quality assurance for the geometric accuracy of cone-beam CT guidance in radiation therapy.

Authors:  Jean-Pierre Bissonnette; Doug Moseley; Elizabeth White; Michael Sharpe; Tom Purdie; David A Jaffray
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

Review 5.  The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide.

Authors:  Stine Korreman; Coen Rasch; Helen McNair; Dirk Verellen; Uwe Oelfke; Philippe Maingon; Ben Mijnheer; Vincent Khoo
Journal:  Radiother Oncol       Date:  2010-02-12       Impact factor: 6.280

6.  Analysis of interfractional set-up errors and intrafractional organ motions during IMRT for head and neck tumors to define an appropriate planning target volume (PTV)- and planning organs at risk volume (PRV)-margins.

Authors:  Minoru Suzuki; Yasumasa Nishimura; Kiyoshi Nakamatsu; Masahiko Okumura; Hisayuki Hashiba; Ryuta Koike; Shuichi Kanamori; Toru Shibata
Journal:  Radiother Oncol       Date:  2006-03-29       Impact factor: 6.280

7.  A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload.

Authors:  H C de Boer; B J Heijmen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-08-01       Impact factor: 7.038

8.  Nonrigid patient setup errors in the head-and-neck region.

Authors:  Buelent Polat; Juergen Wilbert; Kurt Baier; Michael Flentje; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2007-09       Impact factor: 3.621

9.  A comprehensive assessment by tumor site of patient setup using daily MVCT imaging from more than 3,800 helical tomotherapy treatments.

Authors:  Leah K Schubert; David C Westerly; Wolfgang A Tomé; Minesh P Mehta; Emilie T Soisson; Thomas R Mackie; Mark A Ritter; Deepak Khuntia; Paul M Harari; Bhudatt R Paliwal
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-03-15       Impact factor: 7.038

10.  Setup uncertainties of anatomical sub-regions in head-and-neck cancer patients after offline CBCT guidance.

Authors:  Simon van Kranen; Suzanne van Beek; Coen Rasch; Marcel van Herk; Jan-Jakob Sonke
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-04-01       Impact factor: 7.038

View more
  13 in total

1.  IMPACT OF THREE DIFFERENT MATCHING METHODS ON PATIENT SET-UP ERROR IN X-RAY VOLUMETRIC IMAGING FOR HEAD AND NECK CANCER.

Authors:  P Mohandass; D Khanna; B Nishaanth; C Saravanan; Narendra Bhalla; Abhishek Puri; Blessy Mohandass
Journal:  Rep Pract Oncol Radiother       Date:  2020-09-21

2.  A pilot study of highly accelerated 3D MRI in the head and neck position verification for MR-guided radiotherapy.

Authors:  Yihang Zhou; Oi Lei Wong; Kin Yin Cheung; Siu Ki Yu; Jing Yuan
Journal:  Quant Imaging Med Surg       Date:  2019-07

3.  Dose distribution comparison in volumetric-modulated arc therapy plans for head and neck cancers with and without an external body contour extended technique.

Authors:  Yoshihiro Tanaka; Hajime Monzen; Kenji Matsumoto; Shinichiro Inomata; Toshiaki Fuse
Journal:  Rep Pract Oncol Radiother       Date:  2019-11-02

4.  Evaluation of inter-fraction and intra-fraction errors during volumetric modulated arc therapy in nasopharyngeal carcinoma patients.

Authors:  Wen-Jing Yin; Ying Sun; Feng Chi; Jian-Lan Fang; Rui Guo; Xiao-Li Yu; Yan-Ping Mao; Zhen-Yu Qi; Ying Guo; Meng-Zhong Liu; Jun Ma
Journal:  Radiat Oncol       Date:  2013-04-02       Impact factor: 3.481

5.  Different setup errors assessed by weekly cone-beam computed tomography on different registration in nasopharyngeal carcinoma treated with intensity-modulated radiation therapy.

Authors:  Jiqing Su; Wen Chen; Huiyun Yang; Jidong Hong; Zijian Zhang; Guangzheng Yang; Li Li; Rui Wei
Journal:  Onco Targets Ther       Date:  2015-09-14       Impact factor: 4.147

6.  Estimation of daily interfractional larynx residual setup error after isocentric alignment for head and neck radiotherapy: quality assurance implications for target volume and organs-at-risk margination using daily CT on- rails imaging.

Authors:  Charles A Baron; Musaddiq J Awan; Abdallah S R Mohamed; Imad Akel; David I Rosenthal; G Brandon Gunn; Adam S Garden; Brandon A Dyer; Laurence Court; Parag R Sevak; Esengul Kocak-Uzel; Clifton D Fuller
Journal:  J Appl Clin Med Phys       Date:  2014-01-08       Impact factor: 2.102

7.  Evaluation of accuracy and reproducibility of a relocatable maxillary fixation system for fractionated intracranial stereotactic radiation therapy.

Authors:  Majella Russo; Rebecca Owen; Anne Bernard; Vaughan Moutrie; Matthew Foote
Journal:  J Med Radiat Sci       Date:  2015-09-22

8.  Optimization of the margin expanded from the clinical to the planned target volume during intensity-modulated radiotherapy for nasopharyngeal carcinoma.

Authors:  Wang Fangzheng; Jiang Chuner; Sun Quanquan; Ye Zhimin; Yang Shuangyan; Yu Huanhuan; Shi Jianfang; Masoto Sakamoto; Qin Weifeng; Fu Zhenfu; Jiang Yangming; Wang Yuezhen
Journal:  Oncotarget       Date:  2017-11-20

9.  Set-up errors in head and neck cancer treated with IMRT technique assessed by cone-beam computed tomography: a feasible protocol.

Authors:  Durim Delishaj; Stefano Ursino; Francesco Pasqualetti; Fabrizio Matteucci; Agostino Cristaudo; Carlo Pietro Soatti; Amelia Barcellini; Fabiola Paiar
Journal:  Radiat Oncol J       Date:  2018-03-30

10.  Robust Intensity Modulated Proton Therapy (IMPT) Increases Estimated Clinical Benefit in Head and Neck Cancer Patients.

Authors:  Lisanne V van Dijk; Roel J H M Steenbakkers; Bennie ten Haken; Hans Paul van der Laan; Aart A van 't Veld; Johannes A Langendijk; Erik W Korevaar
Journal:  PLoS One       Date:  2016-03-31       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.