| Literature DB >> 29299141 |
Wang Fangzheng1,2, Jiang Chuner3, Sun Quanquan1,2, Ye Zhimin1,2, Yang Shuangyan4, Yu Huanhuan4, Shi Jianfang4, Masoto Sakamoto5, Qin Weifeng1,2, Fu Zhenfu1,2, Jiang Yangming6, Wang Yuezhen1,2.
Abstract
During the radiotherapy process, the emergence of set-up errors is nearly inevitable. Because set-up errors were not detected and corrected daily, planned target volumes were formed by expanding the clinical target volume according to each unit's experience. We optimized the margins of clinical and planned target volumes during administration of intensity-modulated radiotherapy for nasopharyngeal carcinoma. A total of 72 patients newly diagnosed with non-metastatic nasopharyngeal carcinoma and treated with Tomotherapy were prospectively enrolled in the study. For each patient, one megavoltage computed tomography scan was obtained after conventional positioning, online correction, and daily tomotherapy delivery. The interfraction set-up errors were determined using a planning CT based on the registered scan. The mean interfraction errors were -2.437±2.0529 mm, 0.0652±2.3844 mm, 0.318±1.8314 mm, and 0.197±1.8721° for the medial-lateral, superior-inferior, and anterior-posterior directions, and the direction of rotation, respectively. The total MPTV in the three directions was 7.53 mm, 1.83 mm, and 2.08 mm, respectively. The 3-mm margins in the superior-inferior and anterior-posterior directions uniformly expanded from the clinical target volume should be sufficient, and the marging in the medial-lateral direction was up to 7.5 mm. These results suggest that personalized MPTV may be adopted for intensity-modulated radiotherapy planning.Entities:
Keywords: clinical target volume; intensity-modulated radiotherapy; margins; megavoltage computed tomography; nasopharyngeal carcinoma
Year: 2017 PMID: 29299141 PMCID: PMC5746376 DOI: 10.18632/oncotarget.22518
Source DB: PubMed Journal: Oncotarget ISSN: 1949-2553
Figure 1The distributions of the set-up errors in ML, SI, AP directions and Roll
Figure 2The histogram of the distribution frequency of the absolute values for the set-up errors in all directions
Figure 3The box diagram of the absolute setup errors in the three-dimensional orientation
The estimation of the extended boundary in the direction for the patients with NPC (mm)
| Direction | System error | Random error | MPTV |
|---|---|---|---|
| ML(mm) | 2.437 | 2.0529 | 7.53 |
| SI(mm) | 0.0652 | 2.3844 | 1.83 |
| AP(mm) | 0.318 | 1.8314 | 2.08 |