Abstract
PURPOSE: To investigate set-up errors, suggest the adequate planning target volume (PTV) margin and image-guided radiotherapy frequency in head and neck (H&N) cancer treated with intensity-modulated radiotherapy (IMRT) assessed by kV cone-beam computed tomography (CBCT).Entities:
Keywords: Chemotherapy; Cone-beam computed tomography; Head and neck cancer; IMRT technique; Oncology; Radiotherapy
Year: 2018 PMID: 29621873 PMCID: PMC5903362 DOI: 10.3857/roj.2017.00493
Source DB: PubMed Journal: Radiat Oncol J ISSN: 2234-1900
Patient and tumor characteristics
| Characteristic | Value |
|---|---|
| Age (yr) | 68 (44–88) |
| Gender | |
| Male | 45 (75) |
| Female | 15 (25) |
| Smoking status | |
| Smokers | 48 (80) |
| Non-smokers | 12 (20) |
| Alcohol | |
| Potus | 33 (55) |
| Non-potus | 27 (45) |
| Site | |
| Nasopharynx | 3 (5) |
| Oropharynx | 14 (23.5) |
| Oral cavity | 17 (28) |
| Hypopharynx | 4 (7) |
| Larynx | 17 (28) |
| Parotid | 5 (8.5) |
| Histology | |
| Squamous cell carcinoma | 52 (86.6) |
| Undifferentiated carcinoma | 8 (13.4) |
| Radiotherapy treatment | |
| Definitive RT+/-CT | 42 (70) |
| Adjuvant RT+/-CT | 18 (30) |
Values are presented as median (range) or number (%).
RT, radiotherapy; CT, computed tomography.
Fig. 1.Summary diagram of our protocol study. A pretreatment kV cone-beam computed tomography (CBCT) was performed to all patients at first, second, and third day of treatment. Isocenter correction was performed before the treatment if the translational shifts was greater than 3 mm.
Frequencies of CBCT translational shifts greater than 3 and 5 mm in ML, SI, AP directions with and without the application of the systematic correction protocol
| ML | SI | AP | All | |
|---|---|---|---|---|
| Translational shifts >3 mm | ||||
| Before corrections (180 CBCT) | 10 (5.6) | 19 (10.6) | 33 (18.3) | 62 (34.5) |
| After corrections (180 CBCT) | 4 (2.2) | 11 (6.1) | 24 (13.3) | 39 (21.6) |
| Globally (360 CBCT) | 14 (3.8) | 30 (8.3) | 57 (15.9) | 101 (28) |
| Translational shifts >5 mm | ||||
| Before corrections (180 CBCT) | 2 (1.1) | 2 (1.1) | 10 (5.6) | 14 (7.7) |
| After corrections (180 CBCT) | 0 (0) | 1 (0.5) | 4 (2.2) | 5 (2.7) |
| Globally (360 CBCT) | 2 (0.5) | 3 (0.8) | 14 (3.9) | 19 (5.2) |
Values are presented as number (%).
CBCT, cone-beam computed tomography; ML, medial-lateral; SI, superior-Inferior; AP, anterior-posterior.
Fig. 2.Scatter-plot of translational displacements in the three axes without and with the application of the systematic correction protocol: (A) ML direction, (B) SI direction, and (C) AP direction. CBCT, cone-beam computed tomography; ML, medial-lateral; SI, superior-inferior; AP, anterior-posterior.
Overall set-up accuracy with and without the application of the systematic correction protocol
| Entire cohort | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| No systemic correction | Systemic correction | All 360 CBCT | |||||||
| ML | SI | AP | ML | SI | AP | ML | SI | AP | |
| Mean (mm) | 1.28 | 1.56 | 1.99 | 1.23 | 1.28 | 1.95 | 1.26 | 1.28 | 1.97 |
| Range (mm) | 0–9 | 0–6 | 0–10 | 0–4 | 0–6 | 0–6 | 0–9 | 0–5 | 0–10 |
| Systematic error (∑) | 1 | 1 | 1.3 | 0.8 | 0.7 | 0.8 | 0.7 | 0.8 | 0.9 |
| random error (σ) | 1.8 | 1.4 | 1.8 | 1.1 | 1.2 | 1.4 | 1.5 | 1.3 | 1.6 |
| CTV–PTV margin (mm) | 3.76 | 3.48 | 4.51 | 2.84 | 2.59 | 2.98 | 2.8 | 2.91 | 3.37 |
CBCT, cone-beam computed tomography; ML, medial-lateral; SI, superior-Inferior; AP, anterior-posterior; CTV, clinical target volume; PTV, planning target volume.