A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload.

H C de Boer1, B J Heijmen.   

Abstract

PURPOSE: To evaluate a new off-line patient setup correction protocol that minimizes the required number of portal images and perform a comparison with currently applied protocols. METHODS AND MATERIALS: We compared two types of off-line protocols: (a) the widely applied shrinking action level (SAL) protocol, in which the setup error, averaged over the measured treatment fractions, is compared with a threshold that decreases with the number of measurements, to decide if a correction is necessary; and (b) a new "no-action-level" (NAL) protocol, which simply calculates the mean setup error over a fixed number of fractions, and always corrects for it. The performance of the protocols was evaluated by applying them to (a) a database of measured setup errors from 600 prostate patients (with, on average, 10 imaged fractions/patient) and (b) Monte Carlo-generated setup error distributions for various values of the population systematic and random errors.
RESULTS: The NAL protocol achieved a significantly higher accuracy than the SAL protocol for a similar workload in terms of image acquisition and analysis, as well as in setup corrections. The SAL protocol required approximately three times more images than the NAL protocol to obtain the same reduction of systematic errors. Application of the NAL protocol to measured setup errors confirmed its efficacy in systematic error reduction in a real patient population.
CONCLUSION: The NAL protocol performed much more efficiently than the SAL protocol for both actually measured and simulated setup data. The resulting decrease in required portal images not only reduces workload, but also dose to healthy tissue, if dedicated large fields are required for portal imaging (double exposure).

Entities:  

Mesh:

Year:  2001        PMID: 11483348     DOI: 10.1016/s0360-3016(01)01624-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  48 in total

1.  Uncertainties and CTV to PTV margins quantitative assessment using cone-beam CT technique in clinical application for prostate, and head and neck irradiation tumours.

Authors:  X J Juan-Senabre; J López-Tarjuelo; A Conde-Moreno; A Santos-Serra; A L Sánchez-Iglesias; J D Quirós-Higueras; N de Marco Blancas; S Calzada-Feliu; C Ferrer-Albiach
Journal:  Clin Transl Oncol       Date:  2011-11       Impact factor: 3.405

2.  Quality assurance for kilo- and megavoltage in-room imaging and localization for off- and online setup error correction.

Authors:  James M Balter; Larry E Antonuk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

3.  CTV to PTV margins for prostate irradiation. Three-dimensional quantitative assessment of interfraction uncertainties using portal imaging and serial CT scans.

Authors:  Luis A Pérez-Romasanta; Eva Lozano-Martín; Joaquín Velasco-Jiménez; Fermín Mendicote-León; Miguel Sanz-Martín; Javier Torres-Donaire; Carmen Carrascosa-Fernández; Juan Carlos Zapata-Jimínez; Jacinto Arjona-Gutiérrez; Antonio Gil-Agudo
Journal:  Clin Transl Oncol       Date:  2009-09       Impact factor: 3.405

4.  Set-up errors and planning target volume margins in head and neck cancer radiotherapy: a clinical study of image guidance with on-line cone-beam computed tomography.

Authors:  Francesco Dionisi; Mauro Filippo Palazzi; Francesco Bracco; Maria Grazia Brambilla; Claudia Carbonini; Diego Dario Asnaghi; Angelo Filippo Monti; Alberto Torresin
Journal:  Int J Clin Oncol       Date:  2012-03-03       Impact factor: 3.402

5.  Verification of the dose attenuation of a newly developed vacuum cushion for intensity-modulated radiation therapy of prostate cancer.

Authors:  Toru Takakura; Yoshiyuki Ito; Akinori Higashikawa; Tomohiro Nishiyama; Takashi Sakamoto
Journal:  Radiol Phys Technol       Date:  2016-06-03

6.  Comparison of set-up errors by breast size on wing board by portal imaging.

Authors:  Waseem Raza; Sushma Agarwal; Koilpillai Joseph Maria Das; Sellepolyam Kaliyaperumal Senthil Kumar; Punita Lal
Journal:  Rep Pract Oncol Radiother       Date:  2016-05-06

7.  A hybrid strategy of offline adaptive planning and online image guidance for prostate cancer radiotherapy.

Authors:  Yu Lei; Qiuwen Wu
Journal:  Phys Med Biol       Date:  2010-03-30       Impact factor: 3.609

8.  Comparative analysis of image guidance in two institutions for prostate cancer patients.

Authors:  Tomasz Piotrowski; Slav Yartsev; George Rodrigues; Tomasz Bajon
Journal:  Rep Pract Oncol Radiother       Date:  2014-01-02

9.  Technical note: improved positioning protocol for patient setup accuracy in conventional radiotherapy for lung cancer.

Authors:  Hongbo Chai; Yuichiro Narita; Masafumi Takagi; Mikiko Kudo; Tomomi Kimura; Keiichi Kattou
Journal:  Radiol Phys Technol       Date:  2019-09-23

10.  Reducing radiation-associated toxicity using online image guidance (IGRT) in prostate cancer patients undergoing dose-escalated radiation therapy.

Authors:  Martina Becker-Schiebe; Ali Abaci; Tahera Ahmad; Wolfgang Hoffmann
Journal:  Rep Pract Oncol Radiother       Date:  2016-02-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.