| Literature DB >> 22330760 |
Nadège Anizan1, Thomas Carlier, Cecilia Hindorf, Jacques Barbet, Manuel Bardiès.
Abstract
BACKGROUND: Noninvasive multimodality imaging is essential for preclinical evaluation of the biodistribution and pharmacokinetics of radionuclide therapy and for monitoring tumor response. Imaging with nonstandard positron-emission tomography [PET] isotopes such as 124I is promising in that context but requires accurate activity quantification. The decay scheme of 124I implies an optimization of both acquisition settings and correction processing. The PET scanner investigated in this study was the Inveon PET/CT system dedicated to small animal imaging.Entities:
Year: 2012 PMID: 22330760 PMCID: PMC3311594 DOI: 10.1186/2191-219X-2-7
Source DB: PubMed Journal: EJNMMI Res Impact factor: 3.138
Figure 1Schema of the NEMA NU 4-2008 image quality phantom.
NECR value for mouse phantoms filled with 3 MBq of 18F
| Phantom | Energy window (keV) | Coincidence time window (ns) | NECR (kcps) |
|---|---|---|---|
| Mouse | 250 to 750 | 2.8 | 140 |
| 4.7 | 150 | ||
| 350 to 650 | 2.8 | 120 | |
| 4.7 | 130 |
NECR, noise equivalent count rate.
Figure 2NECR variation with LLD for the rat (ULD = 550 keV) and mouse (ULD = 590 keV) phantoms.
Figure 3Profiles drawn on summed sinograms for .
Figure 4NECR curve for mouse-sized (A) and rat-sized (B) phantoms filled with .
Figure 5Coincidence rates for (A) mouse-sized (250 to 590 keV) and (B) rat-sized (250 to 550 keV) phantoms.
SF and GF for the mouse and rat phantoms
| Phantom | Energy window (keV) | SF (%) | GF (%) |
|---|---|---|---|
| Rat | 250 to 550 | 25.4 | 29.2 |
| 390 to 550 | 19.3 | 10.3 | |
| Mouse | 250 to 590 | 16.6 | 19.1 |
| 400 to 590 | 13.7 | 10.2 |
SF, scatter coincidence fraction; GF, gamma-prompt coincidence fraction.
%SDu and the relative error between average activity measured and injected activity
| Isotope | Correction | %SDu | Error (%) | ||||
|---|---|---|---|---|---|---|---|
| 2DFBP | 2DOSEM | 3DOSEM/MAP | 2DFBP | 2DOSEM | 3DOSEM/MAP | ||
| 18F | No correction | 6.3 | 6.7 | 6.2 | -28.3 | -28.4 | -25.2 |
| AC | 5.0 | 5.5 | 5.8 | 0.6 | 0.5 | 6.2 | |
| AC-SC | 5.1 | 5.7 | 4.8 | -1.7 | -2.1 | 3.7 | |
| 124I | No correction | 14.3 | 16.6 | 15.6 | -36.2 | -34.7 | -34.3 |
| AC | 13.8 | 16.3 | 14.5 | -8.0 | -6.9 | -3.6 | |
| AC-SC | 14.4 | 17.8 | 15.2 | -13.7 | -13.4 | -9.9 | |
| AC-SC-GC | 16.9 | 21.3 | 14.2 | -27.2 | -26.6 | -23.0 | |
%SDU, percentage standard deviation; 2DFBP, 2D filtered back-projection; 2DOSEM, 2D ordered-subset expectation maximization; 3DOSEM/MAP, 3DOSEM in combination with maximum a posteriori; AC, attenuation correction; SC, scatter correction; GC, gamma-coincidence correction.
SORa and SORw for the image quality phantom
| Isotope | Correction | SORa (%) | SORw (%) | ||||
|---|---|---|---|---|---|---|---|
| 2DFBP | 2DOSEM | 3DOSEM/MAP | 2DFBP | 2DOSEM | 3DOSEM/MAP | ||
| 18F | No correction | 15.9 | 15.8 | 19.9 | -0.8 | 4.4 | 0.1 |
| AC | 2.6 | 5.8 | 1.2 | 2.4 | 5.6 | 3.3 | |
| AC-SC | 1.4 | 4.9 | 3.0 | 0.2 | 4.2 | 2.8 | |
| 124I | No correction | 17.3 | 18.1 | 17.9 | 4.8 | 9.7 | 3.6 |
| AC | 3.2 | 7.0 | < 0.01 | 8.5 | 12.5 | 8.9 | |
| AC-SC | -0.9 | 4.5 | < 0.01 | 3.7 | 7.4 | 3.6 | |
| AC-SC-GC | -6.6 | 2.2 | < 0.01 | -4.4 | 2.9 | 0.9 | |
SORa, spillover ratio for the air region; SORw, spillover ratio for the water region; 2DFBP, 2D filtered back-projection; 2DOSEM, 2D ordered-subset expectation maximization; 3DOSEM/MAP, 3DOSEM in combination with maximum a posteriori; AC, attenuation correction; SC, scatter correction; GC, gamma-coincidence correction.