Spatial resolution and sensitivity of the Inveon small-animal PET scanner.

Eric P Visser1, Jonathan A Disselhorst, Maarten Brom, Peter Laverman, Martin Gotthardt, Wim J G Oyen, Otto C Boerman.   

Abstract

UNLABELLED: The Inveon small-animal PET scanner is characterized by a large, 127-mm axial length and a 161-mm crystal ring diameter. The associated high sensitivity is obtained by using all lines of response (LORs) up to the maximum ring difference (MRD) of 79, for which the most oblique LORs form acceptance angles of 38.3 degrees with transaxial planes. The result is 2 phenomena that are normally not encountered in PET scanners: a parallax or depth-of-interaction effect in the axial direction and the breakdown of Fourier rebinning (FORE). Both effects cause a deterioration of axial spatial resolution. Limiting the MRD to smaller values reduces this axial blurring at the cost of sensitivity. Alternatively, 3-dimensional (3D) reconstruction techniques can be used in which the rebinning step is absent. The aim of this study was to experimentally determine the spatial resolution and sensitivity of the Inveon for its whole field of view (FOV).
METHODS: Spatial resolution and sensitivity were measured using filtered backprojection (FBP) with FORE, FBP with LOR angle-weighted adapted FORE (AFORE), and 3D ordered-subset expectation maximization followed by maximum a posteriori reconstruction (OSEM3D/MAP).
RESULTS: Tangential and radial full width at half maximum (FWHM) showed almost no dependence on the MRD using FORE and FBP. Tangential FWHMs were 1.5 mm in the center of the FOV (CFOV) and 1.8 mm at the edge of the FOV (EFOV). Radial FWHMs were 1.5 and 3.0 mm in the CFOV and EFOV, respectively. In contrast, axial FWHMs increased with the MRD and ranged between 1.1 and 2.0 mm in the CFOV and between 1.5 and 2.7 mm in the EFOV for a MRD between 1 and 79. AFORE improved the axial resolution for a large part of the FOV, but image noise increased. OSEM3D/MAP yielded uniform spatial resolution in all directions, with an average FWHM of 1.65+/-0.06 mm. Sensitivity in the CFOV for the default energy and coincidence time window was 0.068; peak sensitivity was 0.111.
CONCLUSION: The Inveon showed high spatial resolution and high sensitivity, both of which can be maintained using OSEM3D/MAP reconstruction instead of rebinning and 2D algorithms.

Entities:  

Mesh:

Year:  2009        PMID: 19139188     DOI: 10.2967/jnumed.108.055152

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  64 in total

1.  Molecular Imaging of Proteases in Cancer.

Authors:  Yunan Yang; Hao Hong; Yin Zhang; Weibo Cai
Journal:  Cancer Growth Metastasis       Date:  2009-08-17

2.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Authors:  Keiichi Magota; Naoki Kubo; Yuji Kuge; Ken-Ichi Nishijima; Songji Zhao; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-10       Impact factor: 9.236

3.  On the assessment of spatial resolution of PET systems with iterative image reconstruction.

Authors:  Kuang Gong; Simon R Cherry; Jinyi Qi
Journal:  Phys Med Biol       Date:  2016-02-11       Impact factor: 3.609

4.  Approach to assessing myocardial perfusion in rats using static [13N]-ammonia images and a small-animal PET.

Authors:  Juan José Vaquero; Dong-Wei Gao; Carmen García-Villaba; Stephen Bacharach; Henry Vanbrocklin; Qizhi Fang; Manuel Desco; Randall Lee; Michael Dae
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

5.  Small-animal imaging using clinical positron emission tomography/computed tomography and super-resolution.

Authors:  Frank P DiFilippo; Sagar Patel; Kewal Asosingh; Serpil C Erzurum
Journal:  Mol Imaging       Date:  2012-06       Impact factor: 4.488

6.  A novel facile method of labeling octreotide with (18)F-fluorine.

Authors:  Peter Laverman; William J McBride; Robert M Sharkey; Annemarie Eek; Lieke Joosten; Wim J G Oyen; David M Goldenberg; Otto C Boerman
Journal:  J Nucl Med       Date:  2010-02-11       Impact factor: 10.057

7.  Investigation of Image Reconstruction Parameters of the Mediso nanoScan PC Small-Animal PET/CT Scanner for Two Different Positron Emitters Under NEMA NU 4-2008 Standards.

Authors:  Anastasios Gaitanis; George A Kastis; Elena Vlastou; Penelope Bouziotis; Panayotis Verginis; Constantinos D Anagnostopoulos
Journal:  Mol Imaging Biol       Date:  2017-08       Impact factor: 3.488

8.  Insights into bone metabolism of avian embryos in ovo via 3D and 4D 18F-fluoride positron emission tomography.

Authors:  Lydia Würbach; Alexander Heidrich; Thomas Opfermann; Peter Gebhardt; Hans Peter Saluz
Journal:  Mol Imaging Biol       Date:  2012-12       Impact factor: 3.488

9.  Pharmacoimaging of Blood-Brain Barrier Permeable (FDG) and Impermeable (FLT) Substrates After Intranasal (IN) Administration.

Authors:  Laura L Boles Ponto; Susan Walsh; Jiangeng Huang; Christine Mundt; Katherine Thede-Reynolds; G Leonard Watkins; John Sunderland; Michael Acevedo; Maureen Donovan
Journal:  AAPS J       Date:  2017-12-07       Impact factor: 4.009

Review 10.  Small-animal SPECT and SPECT/CT: application in cardiovascular research.

Authors:  Reza Golestani; Chao Wu; René A Tio; Clark J Zeebregts; Artiom D Petrov; Freek J Beekman; Rudi A J O Dierckx; Hendrikus H Boersma; Riemer H J A Slart
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-01-13       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.