Literature DB >> 22328968

Radiation dose reduction with application of non-linear adaptive filters for abdominal CT.

Sarabjeet Singh1, Mannudeep K Kalra, Mi Kim Sung, Anni Back, Michael A Blake.   

Abstract

AIM: To evaluate the effect of non-linear adaptive filters (NLAF) on abdominal computed tomography (CT) images acquired at different radiation dose levels.
METHODS: Nineteen patients (mean age 61.6 ± 7.9 years, M:F = 8:11) gave informed consent for an Institutional Review Board approved prospective study involving acquisition of 4 additional image series (200, 150, 100, 50 mAs and 120 kVp) on a 64 slice multidetector row CT scanner over an identical 10 cm length in the abdomen. The CT images acquired at 150, 100 and 50 mAs were processed with the NLAF. Two radiologists reviewed unprocessed and processed images for image quality in a blinded randomized manner. CT dose index volume, dose length product, patient weight, transverse diameters, objective noise and CT numbers were recorded. Data were analyzed using Analysis of Variance and Wilcoxon signed rank test.
RESULTS: Of the 31 lesions detected in abdominal CT images, 28 lesions were less than 1 cm in size. Subjective image noise was graded as unacceptable in unprocessed images at 50 and 100 mAs, and in NLAF processed images at 50 mAs only. In NLAF processed images, objective image noise was decreased by 21% (14.4 ± 4/18.2 ± 4.9) at 150 mAs, 28.3% (15.7 ± 5.6/21.9 ± 4) at 100 mAs and by 39.4% (18.8 ± 9/30.4 ± 9.2) at 50 mAs compared to unprocessed images acquired at respective radiation dose levels. At 100 mAs the visibility of smaller structures improved from suboptimal in unprocessed images to excellent in NLAF processed images, whereas diagnostic confidence was respectively improved from probably confident to fully confident.
CONCLUSION: NLAF lowers image noise, improves the visibility of small structures and maintains lesion conspicuity at down to 100 mAs for abdominal CT.

Entities:  

Keywords:  Abdominal computed tomography image quality; Computed tomography radiation dose reduction; Noise reduction filters; Non-linear adaptive filters

Year:  2012        PMID: 22328968      PMCID: PMC3272617          DOI: 10.4329/wjr.v4.i1.21

Source DB:  PubMed          Journal:  World J Radiol        ISSN: 1949-8470


  19 in total

1.  CT images of abdomen and pelvis: effect of nonlinear three-dimensional optimized reconstruction algorithm on image quality and lesion characteristics.

Authors:  Stefania M R Rizzo; Mannudeep K Kalra; Bernhard Schmidt; Rainer Raupach; Michael M Maher; Michael A Blake; Sanjay Saini
Journal:  Radiology       Date:  2005-10       Impact factor: 11.105

2.  New method of evaluating edge-preserving adaptive filters for computed tomography (CT): digital phantom method.

Authors:  Miwa Okumura; Takamasa Ota; Shinsuke Tsukagoshi; Kazuhiro Katada
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2006-07-20

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

4.  [Evaluation of non-linear adaptive smoothing filter by digital phantom].

Authors:  Kazuhiro Sato; Mitsunori Goto; Hiroki Ishiya; Ryosuke Oshita; Issei Mori; Isao Yanagawa
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2008-04-20

5.  [Dose reduction and image quality in MDCT of the upper abdomen: potential of an adaptive post-processing filter].

Authors:  P Kröpil; R S Lanzman; C Walther; S Röhlen; E Godehardt; U Mödder; M Cohnen
Journal:  Rofo       Date:  2009-11-06

6.  Combination of a low-tube-voltage technique with hybrid iterative reconstruction (iDose) algorithm at coronary computed tomographic angiography.

Authors:  Yoshinori Funama; Katsuyuki Taguchi; Daisuke Utsunomiya; Seitaro Oda; Yumi Yanaga; Yasuyuki Yamashita; Kazuo Awai
Journal:  J Comput Assist Tomogr       Date:  2011 Jul-Aug       Impact factor: 1.826

7.  Adaptive statistical iterative reconstruction technique for radiation dose reduction in chest CT: a pilot study.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Matthew D Gilman; Jiang Hsieh; Homer H Pien; Subba R Digumarthy; Jo-Anne O Shepard
Journal:  Radiology       Date:  2011-03-08       Impact factor: 11.105

8.  Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Jiang Hsieh; Paul E Licato; Synho Do; Homer H Pien; Michael A Blake
Journal:  Radiology       Date:  2010-09-09       Impact factor: 11.105

9.  Radiation dose reduction in hepatic multidetector computed tomography with a novel adaptive noise reduction filter.

Authors:  Yoshinori Funama; Kazuo Awai; Osamu Miyazaki; Taiga Goto; Yoshiharu Nakayama; Masamitchi Shimamura; Kumiko Hiraishi; Shinichi Hori; Yasuyuki Yamashita
Journal:  Radiat Med       Date:  2008-04

10.  Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Michael A Moore; Randheer Shailam; Bob Liu; Thomas L Toth; Ellen Grant; Sjirk J Westra
Journal:  Radiology       Date:  2009-05-12       Impact factor: 11.105

View more
  1 in total

1.  Dose reduction with adaptive statistical iterative reconstruction for paediatric CT: phantom study and clinical experience on chest and abdomen CT.

Authors:  F Gay; Y Pavia; N Pierrat; S Lasalle; S Neuenschwander; H J Brisse
Journal:  Eur Radiol       Date:  2013-08-31       Impact factor: 5.315

  1 in total