Radiation dose reduction in hepatic multidetector computed tomography with a novel adaptive noise reduction filter.

Yoshinori Funama1, Kazuo Awai, Osamu Miyazaki, Taiga Goto, Yoshiharu Nakayama, Masamitchi Shimamura, Kumiko Hiraishi, Shinichi Hori, Yasuyuki Yamashita.   

Abstract

PURPOSE: The aim of this study was to optimize a novel adaptive noise reduction filter based on patient body weight and to investigate its utility for improving the image quality of low-dose hepatic computed tomography (CT) scans.
MATERIALS AND METHODS: The tube current-time product was changed from 140 to 180 and from 60 to 100 mAs at standard- and low-dose CT, respectively, based on the body weights of 45 patients. Unenhanced and two-phase contrast-enhanced helical scans were obtained at the standard dose during the hepatic arterial and equilibrium phases. During the equilibrium phase, we obtained low-dose scans of the liver immediately after standard-dose scans. The low-dose CT images were postprocessed with the filter. Two radiologists visually evaluated artifacts in the liver parenchyma and its graininess, the sharpness of the liver contour, tumor conspicuity, homogeneity of the enhancement of the portal vein, and overall image quality.
RESULTS: There was no statistically significant difference between standard and filtered low-dose images with respect to artifacts in the liver, the graininess of the liver parenchyma, tumor conspicuity, homogeneity of enhancement of the portal vein, or overall image quality.
CONCLUSION: The adaptive noise reduction filter effectively reduced image noise. We confirmed the effectiveness of the filter by examining clinical hepatic images obtained at low-dose CT.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18683574     DOI: 10.1007/s11604-007-0202-y

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  23 in total

1.  The ALARA concept in pediatric CT: myth or reality?

Authors:  Thomas L Slovis
Journal:  Radiology       Date:  2002-04       Impact factor: 11.105

2.  Low-dose CT of the abdomen: evaluation of image improvement with use of noise reduction filters pilot study.

Authors:  Mannudeep K Kalra; Michael M Maher; Dushyant V Sahani; Michael A Blake; Peter F Hahn; Gopal B Avinash; Thomas L Toth; Elkan Halpern; Sanjay Saini
Journal:  Radiology       Date:  2003-05-08       Impact factor: 11.105

3.  A computer-simulated liver phantom (virtual liver phantom) for multidetector computed tomography evaluation.

Authors:  Yoshinori Funama; Kazuo Awai; Osamu Miyazaki; Yoshiharu Nakayama; Da Liu; Taiga Goto; Yasuyuki Yamashita; Shinichi Hori
Journal:  Eur Radiol       Date:  2005-10-20       Impact factor: 5.315

4.  Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients.

Authors:  Tejas Dalal; Mannudeep K Kalra; Stefania M R Rizzo; Bernhard Schmidt; Christoph Suess; Thomas Flohr; Michael A Blake; Sanjay Saini
Journal:  Radiology       Date:  2005-08       Impact factor: 11.105

5.  Small hepatocellular carcinoma in patients with chronic liver damage: prospective comparison of detection with dynamic MR imaging and helical CT of the whole liver.

Authors:  Y Yamashita; K Mitsuzaki; T Yi; I Ogata; T Nishiharu; J Urata; M Takahashi
Journal:  Radiology       Date:  1996-07       Impact factor: 11.105

6.  Radiation effective doses to patients undergoing abdominal CT examinations.

Authors:  D E Ware; W Huda; P J Mergo; A L Litwiller
Journal:  Radiology       Date:  1999-03       Impact factor: 11.105

7.  A method for describing the doses delivered by transmission x-ray computed tomography.

Authors:  T B Shope; R M Gagne; G C Johnson
Journal:  Med Phys       Date:  1981 Jul-Aug       Impact factor: 4.071

8.  Application of European Commission reference dose levels in CT examinations in Crete, Greece.

Authors:  V Tsapaki; S Kottou; D Papadimitriou
Journal:  Br J Radiol       Date:  2001-09       Impact factor: 3.039

9.  Improvement of low-contrast detectability in low-dose hepatic multidetector computed tomography using a novel adaptive filter: evaluation with a computer-simulated liver including tumors.

Authors:  Yoshinori Funama; Kazuo Awai; Osamu Miyazaki; Yoshiharu Nakayama; Taiga Goto; Yasuo Omi; Toshiaki Shimonobo; Duo Liu; Yasuyuki Yamashita; Shinichi Hori
Journal:  Invest Radiol       Date:  2006-01       Impact factor: 6.016

10.  Small hypervascular hepatocellular carcinoma revealed by double arterial phase CT performed with single breath-hold scanning and automatic bolus tracking.

Authors:  Tonsok Kim; Takamichi Murakami; Masatoshi Hori; Manabu Takamura; Satoru Takahashi; Atsuya Okada; Shuji Kawata; Modesto Cruz; Michael P Federle; Hironobu Nakamura
Journal:  AJR Am J Roentgenol       Date:  2002-04       Impact factor: 3.959

View more
  2 in total

1.  Noise-reducing algorithms do not necessarily provide superior dose optimisation for hepatic lesion detection with multidetector CT.

Authors:  K L Dobeli; S J Lewis; S R Meikle; D L Thiele; P C Brennan
Journal:  Br J Radiol       Date:  2013-02-07       Impact factor: 3.039

2.  Radiation dose reduction with application of non-linear adaptive filters for abdominal CT.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Mi Kim Sung; Anni Back; Michael A Blake
Journal:  World J Radiol       Date:  2012-01-28
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.