Feasibility of a high-speed gamma-camera design using the high-yield-pileup-event-recovery method.

W H Wong1, H Li, J Uribe, H Baghaei, Y Wang, S Yokoyama.   

Abstract

UNLABELLED: Higher count-rate gamma cameras than are currently used are needed if the technology is to fulfill its promise in positron coincidence imaging, radionuclide therapy dosimetry imaging, and cardiac first-pass imaging. The present single-crystal design coupled with conventional detector electronics and the traditional Anger-positioning algorithm hinder higher count-rate imaging because of the pileup of gamma-ray signals in the detector and electronics. At an interaction rate of 2 million events per second, the fraction of nonpileup events is < 20% of the total incident events. Hence, the recovery of pileup events can significantly increase the count-rate capability, increase the yield of imaging photons, and minimize image artifacts associated with pileups. A new technology to significantly enhance the performance of gamma cameras in this area is introduced.
METHODS: We introduce a new electronic design called high-yield-pileup-event-recovery (HYPER) electronics for processing the detector signal in gamma cameras so that the individual gamma energies and positions of pileup events, including multiple pileups, can be resolved and recovered despite the mixing of signals. To illustrate the feasibility of the design concept, we have developed a small gamma-camera prototype with the HYPER-Anger electronics. The camera has a 10 x 10 x 1 cm NaI(Tl) crystal with four photomultipliers. Hot-spot and line sources with very high 99mTc activities were imaged. The phantoms were imaged continuously from 60,000 to 3,500,000 counts per second to illustrate the efficacy of the method as a function of counting rates.
RESULTS: At 2-3 million events per second, all phantoms were imaged with little distortion, pileup, and dead-time loss. At these counting rates, multiple pileup events (> or = 3 events piling together) were the predominate occurrences, and the HYPER circuit functioned well to resolve and recover these events. The full width at half maximum of the line-spread function at 3,000,000 counts per second was 1.6 times that at 60,000 counts per second.
CONCLUSION: This feasibility study showed that the HYPER electronic concept works; it can significantly increase the count-rate capability and dose efficiency of gamma cameras. In a larger clinical camera, multiple HYPER-Anger circuits may be implemented to further improve the imaging counting rates that we have shown by multiple times. This technology would facilitate the use of gamma cameras for radionuclide therapy dosimetry imaging, cardiac first-pass imaging, and positron coincidence imaging and the simultaneous acquisition of transmission and emission data using different isotopes with less cross-contamination between transmission and emission data.

Entities:  

Mesh:

Year:  2001        PMID: 11337552

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  6 in total

1.  A Real Time Coincidence System for High Count-Rate TOF or Non-TOF PET Cameras Using Hybrid Method Combining AND-Logic and Time-Mark Technology.

Authors:  Chao Wang; Hongdi Li; Rocio A Ramirez; Yuxuan Zhang; Hossain Baghaei; Shitao Liu; Shaohui An; Wai-Hoi Wong
Journal:  IEEE Trans Nucl Sci       Date:  2010-04-01       Impact factor: 1.679

2.  C-SPECT - a Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential.

Authors:  Wei Chang; Caesar E Ordonez; Haoning Liang; Yusheng Li; Jingai Liu
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-06       Impact factor: 1.679

3.  A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

4.  A New Statistics-Based Online Baseline Restorer for a High Count-Rate Fully Digital System.

Authors:  Hongdi Li; Chao Wang; Hossain Baghaei; Yuxuan Zhang; Rocio Ramirez; Shitao Liu; Shaohui An; Wai-Hoi Wong
Journal:  IEEE Trans Nucl Sci       Date:  2010-04       Impact factor: 1.679

5.  HIGH-RESOLUTION L(Y)SO DETECTORS USING PMT-QUADRANT-SHARING FOR HUMAN & ANIMAL PET CAMERAS.

Authors:  Rocio A Ramirez; Shitao Liu; Jiguo Liu; Yuxuan Zhang; Soonseok Kim; Hossain Baghaei; Hongdi Li; Yu Wang; Wai-Hoi Wong
Journal:  IEEE Trans Nucl Sci       Date:  2008-06-01       Impact factor: 1.679

6.  Is absolute quantification of dopaminergic neurotransmission studies with 123I SPECT ready for clinical use?

Authors:  Habib Zaidi; Georges El Fakhri
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-07       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.