| Literature DB >> 22231212 |
Gocha Khelashvili1, James Chu, Aidnag Diaz, Julius Turian.
Abstract
The purpose was to study the dosimetric characteristics of the small diameter (≤ 10.0 mm) BrainLAB cones used for stereotactic radiosurgery (SRS) treatments in conjunction with a Varian Trilogy accelerator. Required accuracy and precision in dose delivery during SRS can be achieved only when the geometric and dosimetric characteristics of the small radiation fields is completely understood. Although a number of investigators have published the dosimetric characteristics of SRS cones, to our knowledge, there is no generally accepted value for the relative output factor (ROF) for the 5.0 mm diameter cone. Therefore, we have investigated the dosimetric properties of the small (≤ 10.0 mm) diameter BrainLAB SRS cones used in conjunction with the iPlan TPS and a Trilogy linear accelerator with a SRS beam mode. Percentage depth dose (PDD), off-axis ratios (OAR), and ROF were measured using a SRS diode and verified with Monte Carlo (MC) simulations. The dependence of ROF on detector material response was studied. The dependence of PDD, OAR, and ROF on the alignment of the beam CAX with the detector motion line was also investigated using MC simulations. An agreement of 1% and 1 mm was observed between measurements and MC for PDD and OAR. The calculated ROF for the 5.0 mm diameter cone was 0.692 ± 0.008--in good agreement with the measured value of 0.683 ± 0.007 after the diode response was corrected. Simulations of the misalignment between the beam axis and detector motion axis for angles between 0.5°-1.0° have shown a deviation > 2% in PDD beyond a certain depth. We have also provided a full set of dosimetric data for BrainLAB SRS cones. Monte Carlo calculated ROF values for cones with diameters less than 10.0 mm agrees with measured values to within 1.8%. Care should be exercised when measuring PDD and OAR for small cones. We recommend the use of MC to confirm the measurement under these conditions.Entities:
Mesh:
Year: 2012 PMID: 22231212 PMCID: PMC5716132 DOI: 10.1120/jacmp.v13i1.3610
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Lateral electron equilibrium (LEE) diameter vs. SRS cone diameter.
| Cone (mm) | 5 | 7.5 | 10 | 12.5 | 15 | 17.5 | 20 | 25 | 30 |
|---|---|---|---|---|---|---|---|---|---|
|
| 16.5 | 16.9 | 17.6 | 18.4 | 18.5 | 18.6 | 18.7 | 19.2 | 19.7 |
Figure 1Results of Monte Carlo simulations vs. measurements for open field for Trilogy in SRS mode.
Figure 2Monte Carlo simulation geometry for calculation of and factors.
Figure 3Monte Carlo simulations vs. measurements for 5 mm diameter cone for Trilogy in SRS mode.
Figure 4Dependence of PPD for 5 mm SDS cone on beam CAX alignment with detector line of motion.
Figure 5A complete set of PDDs and OARs for all cones.
ROF measured and calculated values ± 1 σ.
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| 5.0 |
|
|
|
| 1.32% |
| 7.5 |
|
|
|
| 1.79% |
| 10.0 |
|
|
|
| 1.67% |
| 12.5 |
|
|
|
| 1.48% |
| 15.0 |
|
|
|
| 1.00% |
| 17.5 |
|
|
|
| 1.32% |
| 20.0 | 1.000 |
|
|
| 0.75% |
Figure 6Calculated and factors.
Figure 7Relative output factors vs. cone diameter for Trilogy in SRS mode.