| Literature DB >> 35548030 |
Seby George1, Y Retna Ponmalar1, Henry Finlay Godson1, A Sathish Kumar1, B Paul Ravindran1.
Abstract
Background: The experimental determination of relative output factors presents the greatest challenge, especially for small fields with different detectors. The aim of this study is to evaluate the influence of jaw positions on small-field output factors for the fields defined by micro-multileaf collimator and circular cones with different detectors. Materials andEntities:
Keywords: Detectors; jaw effect; output factor; small-field dosimetry
Year: 2022 PMID: 35548030 PMCID: PMC9084590 DOI: 10.4103/jmp.jmp_111_21
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
The geometric configuration of the fields defined by micro-multileaf collimator and the secondary jaws
| mMLC field (mm×mm) | Jaw field (mm×mm) | mMLC field (mm×mm) | Jaw field (mm×mm) |
|---|---|---|---|
| 6×6 | 8×8 | 12×12 | 14×14 |
| 10×10 | 20×20 | ||
| 20×20 | 44×44 | ||
| 44×44 | 80×80 | ||
| 80×80 | 98×98 | ||
| 98×98 | |||
| 18×18 | 20×20 | 24×24 | 44×44 |
| 44×44 | 60×60 | ||
| 80×80 | 80×80 | ||
| 98×98 | 98×98 | ||
| 30×30 | 44×44 | 36×36 | 44×44 |
| 60×60 | 60×60 | ||
| 80×80 | 80×80 | ||
| 98×98 | 98×98 | ||
| 42×42 | 44×44 | 60×60 | 60×60 |
| 60×60 | 80×80 | ||
| 80×80 | 98×98 | ||
| 98×98 | |||
| 80×80 | 80×80 | 98×98 | 98×98 |
| 98×98 |
mMLC: micro-multileaf collimator
Figure 1Experimental setup for the output factor measurements
Figure 2Comparison of output factors obtained with various detectors for the fields defined by BrainLab micro-multileaf collimator when X and Y jaws were positioned closer to the micro-multileaf collimator defined field and at the maximum jaw position (98 mm × 98 mm)
Output factors obtained with different detectors for the small fields defined by BrainLab micro-multileaf collimator and the percentage deviation from microDiamond reference detector
| mMLC field (mm×mm) | Jaw field (mm×mm) | Output factor | Percentage deviation | |||
|---|---|---|---|---|---|---|
|
|
| |||||
| microDiamond | SRS diode | PinPoint | Diode | PinPoint | ||
| 6×6 | 8×8 | 0.367 | 0.374 | 0.279 | 1.9 | −23.9 |
| 10×10 | 0.453 | 0.462 | 0.368 | 2.0 | −18.8 | |
| 20×20 | 0.615 | 0.619 | 0.511 | 0.5 | −17.0 | |
| 44×44 | 0.631 | 0.634 | 0.527 | 0.5 | −16.5 | |
| 80×80 | 0.635 | 0.638 | 0.532 | 0.5 | −16.2 | |
| 98×98 | 0.636 | 0.640 | 0.536 | 0.6 | −15.8 | |
| 12×12 | 14×14 | 0.643 | 0.643 | 0.603 | −0.1 | −6.3 |
| 20×20 | 0.726 | 0.723 | 0.699 | −0.3 | −3.7 | |
| 44×44 | 0.759 | 0.756 | 0.735 | −0.3 | −3.1 | |
| 80×80 | 0.767 | 0.766 | 0.746 | −0.1 | −2.8 | |
| 98×98 | 0.771 | 0.770 | 0.750 | −0.2 | −2.7 | |
| 18×18 | 20×20 | 0.756 | 0.753 | 0.740 | −0.4 | −2.1 |
| 44×44 | 0.801 | 0.798 | 0.798 | −0.3 | −0.4 | |
| 80×80 | 0.822 | 0.819 | 0.820 | −0.4 | −0.2 | |
| 98×98 | 0.828 | 0.826 | 0.826 | −0.3 | −0.3 | |
| 24×24 | 44×44 | 0.822 | 0.820 | 0.825 | −0.2 | 0.4 |
| 60×60 | 0.834 | 0.832 | 0.838 | −0.3 | 0.4 | |
| 80×80 | 0.848 | 0.845 | 0.851 | −0.4 | 0.4 | |
| 98×98 | 0.859 | 0.856 | 0.862 | −0.3 | 0.3 | |
| 30×30 | 44×44 | 0.837 | 0.835 | 0.840 | −0.3 | 0.3 |
| 60×60 | 0.851 | 0.848 | 0.855 | −0.4 | 0.4 | |
| 80×80 | 0.868 | 0.866 | 0.871 | −0.3 | 0.3 | |
| 98×98 | 0.878 | 0.875 | 0.880 | −0.4 | 0.2 | |
| 36×36 | 44×44 | 0.854 | 0.850 | 0.857 | −0.4 | 0.3 |
| 60×60 | 0.866 | 0.864 | 0.869 | −0.2 | 0.4 | |
| 80×80 | 0.884 | 0.882 | 0.886 | −0.3 | 0.2 | |
| 98×98 | 0.895 | 0.892 | 0.897 | −0.3 | 0.2 | |
| 42×42 | 44×44 | 0.864 | 0.859 | 0.866 | −0.5 | 0.3 |
| 60×60 | 0.881 | 0.877 | 0.883 | −0.5 | 0.2 | |
| 80×80 | 0.896 | 0.892 | 0.899 | −0.4 | 0.3 | |
| 98×98 | 0.908 | 0.904 | 0.911 | −0.4 | 0.3 | |
| 60×60 | 60×60 | 0.911 | 0.908 | 0.914 | −0.3 | 0.3 |
| 80×80 | 0.932 | 0.929 | 0.935 | −0.4 | 0.3 | |
| 98×98 | 0.946 | 0.943 | 0.948 | −0.3 | 0.2 | |
| 80×80 | 80×80 | 0.960 | 0.958 | 0.962 | −0.2 | 0.2 |
| 98×98 | 0.977 | 0.976 | 0.979 | −0.1 | 0.2 | |
| 98×98 | 98×98 | 1 | 1 | 1 | - | - |
SRS: Stereotactic radiosurgery
The increase in output factor for fields defined by micro-multileaf collimator when the jaws were moved from the minimum jaw field to maximum jaw field setting
| mMLC field (mm×mm) | Jaw field (mm×mm) | Increase in output factor | |||
|---|---|---|---|---|---|
|
|
| ||||
| Minimum | Maximum | microDiamond | SRS diode | PinPoint | |
| 6×6 | 8×8 | 98×98 | 1.73 | 1.71 | 1.92 |
| 12×12 | 14×14 | 1.20 | 1.20 | 1.24 | |
| 18×18 | 20×20 | 1.10 | 1.10 | 1.12 | |
| 24×24 | 44×44 | 1.05 | 1.04 | 1.04 | |
| 30×30 | 44×44 | 1.05 | 1.05 | 1.05 | |
| 36×36 | 44×44 | 1.05 | 1.05 | 1.05 | |
| 42×42 | 44×44 | 1.05 | 1.05 | 1.05 | |
| 60×60 | 60×60 | 1.04 | 1.04 | 1.04 | |
| 80×80 | 80×80 | 1.02 | 1.02 | 1.02 | |
| 98×98 | 98×98 | 1 | 1 | 1 | |
SRS: Stereotactic radiosurgery, mMLC: micro-multileaf collimator
Figure 3(a-d) Influence of jaw position on output factor for four different fields defined by micro-multileaf collimator
Output factors obtained with different detectors for the small fields defined by BrainLab stereotactic cones and the percentage deviation from PTW microDiamond detector reference value
| Cone size (cm) | Jaw size (cm) | Output factor | Percentage difference | |||
|---|---|---|---|---|---|---|
|
|
| |||||
| microDiamond | SRS diode | PinPoint | Diode | PinPoint | ||
| 1 | 1.5 | 0.646±0.03 | 0.657±0.04 | 0.615±0.06 | 1.7 | −4.8 |
| 1.5 | 2 | 0.720±0.03 | 0.728±0.04 | 0.699±0.05 | 1.1 | −2.9 |
| 2 | 2.5 | 0.783±0.03 | 0.781±0.03 | 0.768±0.04 | −0.2 | −1.9 |
| 2.5 | 3 | 0.803±0.02 | 0.801±0.02 | 0.802±0.02 | −0.2 | −0.1 |
| 3 | 3.5 | 0.823±0.01 | 0.820±0.01 | 0.821±0.01 | −0.3 | −0.2 |
| 3.5 | 4 | 0.840±0.01 | 0.837±0.01 | 0.839±0.01 | −0.4 | −0.1 |
| 4 | 4.5 | 0.854±0.01 | 0.851±0.01 | 0.853±0.01 | −0.4 | −0.1 |
Figure 4The output factors obtained before and after applying the detector-specific output correction factor of various detectors for the fields defined by stereotactic cones