Novel methodology to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system.

Howard Zisser1, Marc Breton, Eyal Dassau, Kateryna Markova, Wendy Bevier, Dale Seborg, Boris Kovatchev.   

Abstract

BACKGROUND: This article describes two novel and easy approaches for assessing the accuracy of insulin pumps as implemented within the artificial pancreas system. The approaches are illustrated by data testing the OmniPod Insulin Management System at its lowest delivery volume (0.05 U) and at doses of 0.1, 0.2, 1, and 6U.
METHOD: In method 1, a pipette, digital microscope, and imaging software were used to measure average bolus delivery on a linear scale for multiple volumes. In method 2, a digital microscope and imaging software were used to measure the volume of a spherical bolus of 0.05 U of insulin.
RESULTS: Bench testing results using the two novel methods demonstrated that the OmniPod is extremely accurate, with a relative error ranging from -0.90% to +0.96% for all measured doses (0.05, 0.1, 0.2, 1, and 6 U). In method 1, at target bolus dose of 0.05 U, the mean delivered dose (± standard deviation) was 0.0497 ± 0.003 U, 0.099 ± 0.005 U at 0.1 U, 0.2 ± <1e-5 U at 0.2 U, 1.001 ± 0.018 U at 1 U, and 6.03 ± 0.04 U at 6 U. In method 2, at target bolus dose of 0.5 ml, the mean delivered dose for both OmniPods was 0.505 ± 0.014.
CONCLUSIONS: Both methods confirmed a high degree of accuracy for the OmniPod insulin pump. These techniques can be used to estimate delivery volume in other infusion pumps as well.
© 2011 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22226273      PMCID: PMC3262722          DOI: 10.1177/193229681100500627

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  16 in total

1.  Retrospective analysis and patient satisfaction assessment of insulin pump therapy in patients with type 2 diabetes.

Authors:  Chad K Gentry; L Brian Cross; Benjamin N Gross; M Shawn McFarland; William H Bestermann
Journal:  South Med J       Date:  2011-01       Impact factor: 0.954

2.  Using insulin pump therapy in poorly controlled type 2 diabetes.

Authors:  Phyllis Wolff-McDonagh; Judith Kaufmann; Stephen Foreman; Steven Wisotsky; Jo Ann Wisotsky; Craig Wexler
Journal:  Diabetes Educ       Date:  2010-06-22       Impact factor: 2.140

3.  Insulin pump therapy in patients with type 2 diabetes safely improved glycemic control using a simple insulin dosing regimen.

Authors:  Steven V Edelman; Bruce W Bode; Timothy S Bailey; Mark S Kipnes; Rocco Brunelle; Xiaojing Chen; Juan P Frias
Journal:  Diabetes Technol Ther       Date:  2010-08       Impact factor: 6.118

4.  Comparison of continuous subcutaneous insulin infusion and multiple daily injection regimens using insulin lispro in type 1 diabetic patients on intensified treatment: a randomized study. The Study Group for the Development of Pump Therapy in Diabetes.

Authors:  H Hanaire-Broutin; V Melki; S Bessières-Lacombe; J P Tauber
Journal:  Diabetes Care       Date:  2000-09       Impact factor: 19.112

5.  Insulin pump therapy: a meta-analysis.

Authors:  Jill Weissberg-Benchell; Jeanne Antisdel-Lomaglio; Roopa Seshadri
Journal:  Diabetes Care       Date:  2003-04       Impact factor: 19.112

6.  Insulin pump therapy in youth with type 1 diabetes: a retrospective paired study.

Authors:  Revital Nimri; Naomi Weintrob; Hadassa Benzaquen; Regina Ofan; Gila Fayman; Moshe Phillip
Journal:  Pediatrics       Date:  2006-06       Impact factor: 7.124

7.  Effects of pulsatile subcutaneous injections of insulin lispro on plasma insulin concentration levels.

Authors:  Alice Chan; Marc D Breton; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2008-09

8.  Modular artificial beta-cell system: a prototype for clinical research.

Authors:  Eyal Dassau; Howard Zisser; Cesar C Palerm; Bruce A Buckingham; Lois Jovanovic; Francis J Doyle
Journal:  J Diabetes Sci Technol       Date:  2008-09

9.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

10.  Accuracy and reproducibility of low dose insulin administration using pen-injectors and syringes.

Authors:  M G Gnanalingham; P Newland; C P Smith
Journal:  Arch Dis Child       Date:  1998-07       Impact factor: 3.791

View more
  13 in total

1.  Analysis of novel methods to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system.

Authors:  Manuel Ochoa; Babak Ziaie
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

Review 2.  Boluses in Insulin Therapy.

Authors:  Ralph Ziegler; Guido Freckmann; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-07-10

3.  Establishing Methods to Determine Clinically Relevant Bolus and Basal Rate Delivery Accuracy of Insulin Pumps.

Authors:  Ulrike Kamecke; Delia Waldenmaier; Cornelia Haug; Ralph Ziegler; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-07-13

4.  Accuracy of a new patch pump based on a microelectromechanical system (MEMS) compared to other commercially available insulin pumps: results of the first in vitro and in vivo studies.

Authors:  Sophie Borot; Sylvia Franc; Justine Cristante; Alfred Penfornis; Pierre-Yves Benhamou; Bruno Guerci; Hélène Hanaire; Eric Renard; Yves Reznik; Chantal Simon; Guillaume Charpentier
Journal:  J Diabetes Sci Technol       Date:  2014-07-30

5.  Patch Pumps: Are They All the Same?

Authors:  Lutz Heinemann; Delia Waldenmaier; Bernd Kulzer; Ralph Ziegler; Barry Ginsberg; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-08-22

6.  Reporting Insulin Pump Accuracy: Trumpet Curves According to IEC 60601-2-24 and Beyond.

Authors:  Stefan Pleus; Ulrike Kamecke; Delia Waldenmaier; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-12

7.  Artificial pancreas (AP) clinical trial participants' acceptance of future AP technology.

Authors:  Wendy C Bevier; Serena M Fuller; Ryan P Fuller; Richard R Rubin; Eyal Dassau; Francis J Doyle; Lois Jovanovič; Howard C Zisser
Journal:  Diabetes Technol Ther       Date:  2014-05-08       Impact factor: 6.118

8.  Comparative dose accuracy of durable and patch insulin infusion pumps.

Authors:  Luis G Jahn; Jorge J Capurro; Brian L Levy
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

Review 9.  Insulin pump (dose-to-dose) accuracy: what does it mean and when is it important?

Authors:  Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2014-08-26

10.  10th Annual Symposium on Self-Monitoring of Blood Glucose, April 27-29, 2017, Warsaw, Poland.

Authors:  Christopher G Parkin; Anita Homberg; Rolf Hinzmann
Journal:  Diabetes Technol Ther       Date:  2017-11-14       Impact factor: 6.118

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.