Establishing Methods to Determine Clinically Relevant Bolus and Basal Rate Delivery Accuracy of Insulin Pumps.

Ulrike Kamecke1, Delia Waldenmaier1, Cornelia Haug1, Ralph Ziegler2, Guido Freckmann1.   

Abstract

BACKGROUND: Adequate testing of delivery accuracy of insulin pumps is under discussion. Especially for patch pumps, test settings are challenging. In addition, evaluation and presentation of accuracy results in a way that is reasonable and useful for clinicians, not only for technicians, is important.
METHODS: Test setups based on IEC 60601-2-24 were used and, in addition, different setups for patch pumps were compared to identify an adequate alternative for pumps without external infusion set. These setups are applicable for both bolus and basal rate accuracy testing. In addition, evaluation procedures considering clinical relevance were compiled.
RESULTS: A setup for patch pumps that provides reliable results could be realized. Evaluation of basal rate accuracy data should also consider the actual clinical use of insulin pumps and thus, deviating from IEC 60601-2-24, compose the whole measurement period without excluding the first 24 hours. In addition to the presentation using trumpet curves, accuracy of 1-hour windows should be evaluated and displayed.
CONCLUSIONS: This article proposes an approach on how to test, evaluate, and present bolus and basal rate accuracy of insulin pumps from a clinical perspective.

Entities:  

Keywords:  CSII; accuracy; basal rate; bolus; insulin pump

Mesh:

Substances:

Year:  2018        PMID: 30003823      PMCID: PMC6313283          DOI: 10.1177/1932296818788349

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  16 in total

1.  Guidelines for insulin dosing in continuous subcutaneous insulin infusion using new formulas from a retrospective study of individuals with optimal glucose levels.

Authors:  John Walsh; Ruth Roberts; Timothy Bailey
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Siphon effects on continuous subcutaneous insulin infusion pump delivery performance.

Authors:  Howard C Zisser; Wendy Bevier; Eyal Dassau; Lois Jovanovic
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

3.  ISPAD Clinical Practice Consensus Guidelines 2014. Insulin treatment in children and adolescents with diabetes.

Authors:  Thomas Danne; Hans-Jacob Bangstad; Larry Deeb; Przemyslawa Jarosz-Chobot; Lucy Mungaie; Banshi Saboo; Tatsuhiko Urakami; Tadej Battelino; Ragnar Hanas
Journal:  Pediatr Diabetes       Date:  2014-09       Impact factor: 4.866

4.  Subcutaneous insulin infusion: change in basal infusion rate has no immediate effect on insulin absorption rate.

Authors:  P Hildebrandt; K Birch; B M Jensen; C Kühl
Journal:  Diabetes Care       Date:  1986 Nov-Dec       Impact factor: 19.112

5.  Use of insulin pump therapy in the pediatric age-group: consensus statement from the European Society for Paediatric Endocrinology, the Lawson Wilkins Pediatric Endocrine Society, and the International Society for Pediatric and Adolescent Diabetes, endorsed by the American Diabetes Association and the European Association for the Study of Diabetes.

Authors:  Moshe Phillip; Tadej Battelino; Henry Rodriguez; Thomas Danne; Francine Kaufman
Journal:  Diabetes Care       Date:  2007-03-19       Impact factor: 19.112

6.  Basal rates and circadian profiles in continuous subcutaneous insulin infusion (CSII) differ for preschool children, prepubertal children, adolescents and young adults.

Authors:  Rainer Bachran; Peter Beyer; Christof Klinkert; Bettina Heidtmann; Joachim Rosenbauer; Reinhard W Holl
Journal:  Pediatr Diabetes       Date:  2011-05-05       Impact factor: 4.866

7.  Changes in basal insulin infusion rates with subcutaneous insulin infusion: time until a change in metabolic effect is induced in patients with type 1 diabetes.

Authors:  Lutz Heinemann; Leszek Nosek; Christoph Kapitza; Matthias-Axel Schweitzer; Lars Krinelke
Journal:  Diabetes Care       Date:  2009-06-01       Impact factor: 19.112

8.  Comparative dose accuracy of durable and patch insulin infusion pumps.

Authors:  Luis G Jahn; Jorge J Capurro; Brian L Levy
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

Review 9.  Insulin pump (dose-to-dose) accuracy: what does it mean and when is it important?

Authors:  Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2014-08-26

10.  A Comparative Pulse Accuracy Study of Two Commercially Available Patch Insulin Infusion Pumps.

Authors:  Jenna L Bowen; Chris J Allender
Journal:  Eur Endocrinol       Date:  2016-08-28
View more
  6 in total

1.  Reporting Insulin Pump Accuracy: Trumpet Curves According to IEC 60601-2-24 and Beyond.

Authors:  Stefan Pleus; Ulrike Kamecke; Delia Waldenmaier; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-12

Review 2.  Evaluation of hypoglycemic therapeutics and nutritional supplementation for type 2 diabetes mellitus management: An insight on molecular approaches.

Authors:  Murugan Prasathkumar; Robert Becky; Salim Anisha; Chenthamara Dhrisya; Subramaniam Sadhasivam
Journal:  Biotechnol Lett       Date:  2022-02-04       Impact factor: 2.461

3.  Description of a Novel Patch Pump for Insulin Delivery and Comparative Accuracy Evaluation.

Authors:  Annette Baumstark; Jochen Mende; Joji Uchiyama; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2021-03-19

4.  Evaluation of the Accuracy and Reliability of a Tubeless Insulin Infusion System Under Laboratory Conditions.

Authors:  Jochen Mende; Manuel Eichenlaub; Delia Waldenmaier; Cornelia Haug; Guido Freckmann; Ralph Ziegler
Journal:  J Diabetes Sci Technol       Date:  2022-01-19

5.  Accuracy of Bolus and Basal Rate Delivery of Different Insulin Pump Systems.

Authors:  Guido Freckmann; Ulrike Kamecke; Delia Waldenmaier; Cornelia Haug; Ralph Ziegler
Journal:  Diabetes Technol Ther       Date:  2019-03-22       Impact factor: 6.118

6.  Performance evaluation study of a commercially available smart patient-controlled analgesia pump with the microbalance method and an infusion analyzer.

Authors:  Jinsoo Park; Bongsu Jung
Journal:  J Dent Anesth Pain Med       Date:  2022-03-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.