Comparative dose accuracy of durable and patch insulin infusion pumps.

Luis G Jahn1, Jorge J Capurro, Brian L Levy.   

Abstract

BACKGROUND: As all major insulin pump manufacturers comply with the international infusion pump standard EN 60601-2-24:1998, there may be a general assumption that all pumps are equal in insulin-delivery accuracy. This research investigates single-dose and averaged-dose accuracy of incremental basal deliveries for one patch model and three durable models of insulin pumps.
METHOD: For each pump model, discrete single doses delivered during 0.5 U/h basal rate infusion over a 20 h period were measured using a time-stamped microgravimetric system. Dose accuracy was analyzed by comparing single doses and time-averaged doses to specific accuracy thresholds (±5% to ±30%).
RESULTS: The percentage of single doses delivered outside accuracy thresholds of ±5%, ±10%, and ±20% were as follows: Animas OneTouch® Ping® (43.2%, 14.3%, and 1.8%, respectively), Roche Accu-Chek® Combo (50.6%, 24.4%, and 5.5%), Medtronic Paradigm® Revel™/Veo™ (54.2%, 26.7%, and 6.6%), and Insulet OmniPod® (79.1%, 60.5%, and 34.9%). For 30 min, 1 h, and 2 h averaging windows, the percentage of doses delivered outside a ±15% accuracy were as follows: OneTouch Ping (1.0%, 0.4%, and 0%, respectively), Accu-Chek Combo (4.2%, 3.5%, and 3.1%), Paradigm Revel/Veo (3.9%, 3.1%, and 2.2%), and OmniPod (33.9%, 19.9%, and 10.3%).
CONCLUSIONS: This technical evaluation demonstrates significant differences in single-dose and averaged-dose accuracy among the insulin pumps tested. Differences in dose accuracy were most evident between the patch pump model and the group of durable pump models. Of the pumps studied, the Animas OneTouch Ping demonstrated the best single-dose and averaged-dose accuracy. Further research on the clinical relevance of these findings is warranted.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23911184      PMCID: PMC3879767          DOI: 10.1177/193229681300700425

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  5 in total

1.  Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes.

Authors:  David M Nathan; Patricia A Cleary; Jye-Yu C Backlund; Saul M Genuth; John M Lachin; Trevor J Orchard; Philip Raskin; Bernard Zinman
Journal:  N Engl J Med       Date:  2005-12-22       Impact factor: 91.245

2.  Novel methodology to determine the accuracy of the OmniPod insulin pump: a key component of the artificial pancreas system.

Authors:  Howard Zisser; Marc Breton; Eyal Dassau; Kateryna Markova; Wendy Bevier; Dale Seborg; Boris Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

3.  The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus.

Authors:  D M Nathan; S Genuth; J Lachin; P Cleary; O Crofford; M Davis; L Rand; C Siebert
Journal:  N Engl J Med       Date:  1993-09-30       Impact factor: 91.245

4.  Accuracy and precision of low-dose insulin administration using syringes, pen injectors, and a pump.

Authors:  Katherine Keith; David Nicholson; Douglas Rogers
Journal:  Clin Pediatr (Phila)       Date:  2004 Jan-Feb       Impact factor: 1.168

5.  Pharmacokinetics and pharmacodynamics of insulin analogs in special populations with type 2 diabetes mellitus.

Authors:  Candis M Morello
Journal:  Int J Gen Med       Date:  2011-12-12
  5 in total
  11 in total

Review 1.  Boluses in Insulin Therapy.

Authors:  Ralph Ziegler; Guido Freckmann; Lutz Heinemann
Journal:  J Diabetes Sci Technol       Date:  2016-07-10

2.  Establishing Methods to Determine Clinically Relevant Bolus and Basal Rate Delivery Accuracy of Insulin Pumps.

Authors:  Ulrike Kamecke; Delia Waldenmaier; Cornelia Haug; Ralph Ziegler; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-07-13

3.  Accuracy of a new patch pump based on a microelectromechanical system (MEMS) compared to other commercially available insulin pumps: results of the first in vitro and in vivo studies.

Authors:  Sophie Borot; Sylvia Franc; Justine Cristante; Alfred Penfornis; Pierre-Yves Benhamou; Bruno Guerci; Hélène Hanaire; Eric Renard; Yves Reznik; Chantal Simon; Guillaume Charpentier
Journal:  J Diabetes Sci Technol       Date:  2014-07-30

4.  Patch Pumps: Are They All the Same?

Authors:  Lutz Heinemann; Delia Waldenmaier; Bernd Kulzer; Ralph Ziegler; Barry Ginsberg; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-08-22

5.  Analysis of "Laboratory and Benchtop Performance of a Mealtime Insulin Delivery System".

Authors:  Eric Zijlstra
Journal:  J Diabetes Sci Technol       Date:  2018-04-05

6.  Reporting Insulin Pump Accuracy: Trumpet Curves According to IEC 60601-2-24 and Beyond.

Authors:  Stefan Pleus; Ulrike Kamecke; Delia Waldenmaier; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2018-10-12

Review 7.  Insulin pump (dose-to-dose) accuracy: what does it mean and when is it important?

Authors:  Howard Zisser
Journal:  J Diabetes Sci Technol       Date:  2014-08-26

8.  10th Annual Symposium on Self-Monitoring of Blood Glucose, April 27-29, 2017, Warsaw, Poland.

Authors:  Christopher G Parkin; Anita Homberg; Rolf Hinzmann
Journal:  Diabetes Technol Ther       Date:  2017-11-14       Impact factor: 6.118

9.  Accuracy of Bolus and Basal Rate Delivery of Different Insulin Pump Systems.

Authors:  Guido Freckmann; Ulrike Kamecke; Delia Waldenmaier; Cornelia Haug; Ralph Ziegler
Journal:  Diabetes Technol Ther       Date:  2019-03-22       Impact factor: 6.118

10.  A Comparative Pulse Accuracy Study of Two Commercially Available Patch Insulin Infusion Pumps.

Authors:  Jenna L Bowen; Chris J Allender
Journal:  Eur Endocrinol       Date:  2016-08-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.