| Literature DB >> 22163627 |
Kuo-Chih Liao1, Shih-Chieh Chang, Cheng-Yang Chiu, Yu-Hsiang Chou.
Abstract
The objective of this study was to evaluate the acute response of Sencil(™), a fiber-optic sensor, in point accuracy for glucose monitoring in vivo on healthy dogs under anesthesia. A total of four dogs with clinically normal glycemia were implanted with one sensor each in the chest region to measure the interstitial glucose concentration during the ovariohysterectomy procedure. The data was acquired every 10 seconds after initiation, and was compared to the concentration of venous plasma glucose sampled during the surgery procedures for accuracy of agreement analysis. In the four trials with a range of 71-297 mg/dL plasma glucose, the collected 21 pairs of ISF readings from the Sencil™ and the plasma reference showed superior dispersion of residue values than the conventional system, and a linear correlation (the Pearson correlation coefficient is 0.9288 and the y-intercept is 14.22 mg/dL). The MAD (17.6 mg/dL) and RMAD (16.16%) of Sencil™ measurements were in the comparable range of the conventional system. The Clarke error grid analysis indicated that 100% of the paired points were in the clinically acceptable zone A (61.9%) and B (38.1%).Entities:
Keywords: Clarke error grid analysis; calibration; fiber-optic sensor; in vivo glucose monitoring; point accuracy
Mesh:
Substances:
Year: 2010 PMID: 22163627 PMCID: PMC3231153 DOI: 10.3390/s100807789
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.Illustration of the Sencil™ system for glucose sensing. (a) Similarity of shape and size between Sencil™ and human hair with attached follicle (white bar = 100 μm). (b) Sensor components and relationships to tissue in vivo. (c) Changes in FRET between fluorophores covalently immobilized on the flexible PEG matrix depend on changes in the distance between them, which in turn depends on the competitive natural affinity between Con A and various saccharides such as betacyclodextrin and glucose. The figure is modified with permission from [10], published by Biosensors Bioelectronics, 2008.
Figure 2.Illustration of the relative spacial positions for implantation, blood sampling and IV infusion on each subject (a). Arrangement of devices around the operation table (b).
Figure 3.Sensor calibration from the linear fitting result of in vitro measurements. (a) Absolute emissions at 525 nm and 570 nm (arrows) were applied to calculate the ratio for indicating glucose concentration. (b) Linear correlation between the emission ratios and the corresponding glucose concentrations. The standard deviations (from the six sensors manufactured in the same batch) of the calibration values at corresponding glucose concentrations are shown as green error bars.
Figure 4.Fluorescence emission ratio change trend before and after implantation.
Figure 5.Profiles of the ISF reading from Sencil™ and plasma reference from Hitachi 7050. (a) case study #1 (b) case study #2 (c) case study #3 (d) case study #4.
Figure 6.Accuracy of agreement between Sencil reading and plasma reference: (a) Bland-Altman method analysis; (b) least square linear fitting result; (c) Clarke error grid analysis.