Correlation of fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes.

S K Garg1, R O Potts, N R Ackerman, S J Fermi, J A Tamada, H P Chase.   

Abstract

OBJECTIVE: The purpose of this study was to compare measurements of glucose obtained via iontophoretic extraction with the GlucoWatch automatic glucose biographer (Cygnus, Inc., Redwood City, CA) with capillary blood glucose values that were determined 1) in a controlled outpatient clinic setting and 2) in a home setting. RESEARCH DESIGN AND METHODS: There were 76 GlucoWatch biographers used on 28 different young adults (21 women and 7 men) with type 1 diabetes (age 30.9 +/- 6.9 years and duration of diabetes 18.4 +/- 8.1 years [mean +/- SD]) in a controlled outpatient clinic setting. Some subjects participated on multiple days. Subjects wore two GlucoWatch biographers, each on the forearm (ventral aspect). Comparisons were made to HemoCue blood glucose analyzer (Aktiebolgat Leo, Helsingborg, Sweden) capillary blood glucose measurements. In addition, GlucoWatch biographers (one each day for 3 consecutive days) were used by 12 subjects (8 women, 4 men) in a home setting. Comparisons were made to capillary blood glucose values determined using the One Touch Profile meter (Johnson & Johnson, New Brunswick, NJ).
RESULTS: GlucoWatch biographer glucose values correlated well with capillary blood glucose values determined using the HemoCue analyzer in the clinic setting (r = 0.90, 1,554 paired data points) and using the One Touch Profile meter in the home setting (r = 0.85, 204 paired data points). When 36 subjects wore two biographers simultaneously, the correlation between the two biographers was r = 0.94. The error grid analysis demonstrated that > 96% of biographer glucose values determined in the clinic or home setting were in the clinically acceptable A and B regions.
CONCLUSIONS: This study confirms the accuracy and precision of glucose values as determined using the GlucoWatch biographer in clinic and home settings.

Entities:  

Mesh:

Year:  1999        PMID: 10526740     DOI: 10.2337/diacare.22.10.1708

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  34 in total

Review 1.  Severe hypoglycemia in adults.

Authors:  Mary F Carroll; Mark R Burge; David S Schade
Journal:  Rev Endocr Metab Disord       Date:  2003-05       Impact factor: 6.514

2.  The need for clinical accuracy guidelines for blood glucose monitors.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2012-01-01

3.  The Future of Glucose Monitoring.

Authors:  Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2016-02       Impact factor: 6.118

Review 4.  In vitro, in vivo and post explantation testing of glucose-detecting biosensors: current methods and recommendations.

Authors:  Heidi E Koschwanez; William M Reichert
Journal:  Biomaterials       Date:  2007-04-19       Impact factor: 12.479

5.  DTT: 19 Years and Counting ….

Authors:  Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2017-02       Impact factor: 6.118

6.  The Future of Continuous Glucose Monitoring.

Authors:  Satish K Garg; Halis K Akturk
Journal:  Diabetes Technol Ther       Date:  2017-06       Impact factor: 6.118

7.  Flash Glucose Monitoring: The Future Is Here.

Authors:  Satish K Garg; Halis Kaan Akturk
Journal:  Diabetes Technol Ther       Date:  2017-05       Impact factor: 6.118

8.  Development of a highly responsive needle-type glucose sensor using polyimide for a wearable artificial endocrine pancreas.

Authors:  Shinji Ichimori; Kenro Nishida; Seiya Shimoda; Taiji Sekigami; Yasuto Matsuo; Kenshi Ichinose; Motoaki Shichiri; Michiharu Sakakida; Eiichi Araki
Journal:  J Artif Organs       Date:  2006       Impact factor: 1.731

9.  Recruitment and Collection of Dermal Interstitial Fluid Using a Microneedle Patch.

Authors:  Chandana Kolluru; Mikayla Williams; Jeremy Chae; Mark R Prausnitz
Journal:  Adv Healthc Mater       Date:  2019-01-04       Impact factor: 9.933

10.  Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.

Authors:  Tai-Ping Sun; Hsiu-Li Shieh; Congo Tak-Shing Ching; Yan-Dong Yao; Su-Hua Huang; Chia-Ming Liu; Wei-Hao Liu; Chung-Yuan Chen
Journal:  Int J Nanomedicine       Date:  2010-05-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.