Literature DB >> 22116296

Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section.

A Butwick1, G Hilton, B Carvalho.   

Abstract

BACKGROUND: The ability to measure haemoglobin (Hb) real-time and non-invasively offers important clinical value in the assessment of acute changes in maternal Hb during the peripartum period. This study evaluates the Masimo Rainbow SET(®) Radical-7 Pulse CO-Oximeter in a pregnant population undergoing Caesarean section (CS).
METHODS: Fifty patients undergoing elective CS were enrolled in this prospective, controlled study and followed for 48 h after surgery. Non-invasive Masimo Hb (SpHb) values were compared with laboratory Hb values from venous blood samples drawn at baseline, immediately post-CS, and 24 h post-CS using the Bland-Altman plots. Longitudinal analysis of SpHb changes over time was performed using mixed-effects regression modelling.
RESULTS: For the comparison between SpHb and laboratory Hb, SpHb displayed a significant positive bias at baseline {1.22 g dl(-1) [95% confidence interval (CI): 0.89-1.54]} and at 24 h post-CS [1.36 g dl(-1) (95% CI: 1.04-1.68)]. The bias immediately post-CS was 0.14 g dl(-1) (95% CI: -0.18 to 0.46). The limits of agreement at baseline, immediately post-CS, and at 24 h post-CS were: -0.9 and 3.33, -2.35 and 2.56, and -0.55 and 3.27 g dl(-1), respectively. The mean decrease in SpHb from baseline to 48 h post-CS was ∼1 g dl(-1).
CONCLUSIONS: The variability in bias and limits of agreements of the Rainbow SET(®) Radical-7 Pulse CO-Oximeter SpHb may limit its clinical utility for assessing Hb concentration in patients undergoing elective CS. Modifications are needed in the calibration of the device to improve accuracy and precision in an obstetric setting. The study was registered at clinicaltrials.gov (NCT01108471) before participant enrolment: URL=http://clinicaltrials.gov/ct2/show/NCT01108471?term=butwick&rank=1.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22116296     DOI: 10.1093/bja/aer373

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  17 in total

1.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

2.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

3.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

4.  Effect of a High-Rate Versus a Low-Rate Oxytocin Infusion for Maintaining Uterine Contractility During Elective Cesarean Delivery: A Prospective Randomized Clinical Trial.

Authors:  Adrienne Duffield; Christine McKenzie; Brendan Carvalho; Bharathi Ramachandran; Victoria Yin; Yasser Y El-Sayed; Edward T Riley; Alexander J Butwick
Journal:  Anesth Analg       Date:  2017-03       Impact factor: 5.108

5.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

6.  Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants.

Authors:  C Nicholas; R George; S Sardesai; M Durand; R Ramanathan; R Cayabyab
Journal:  J Perinatol       Date:  2015-03-05       Impact factor: 2.521

7.  The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.

Authors:  Junichi Saito; Masato Kitayama; Masafumi Oishi; Tomoyuki Kudo; Masahiro Sawada; Hiroshi Hashimoto; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2014-06-28       Impact factor: 2.078

8.  Continuous noninvasive hemoglobin monitor from pulse ox: ready for prime time?

Authors:  Bellal Joseph; Pantelis Hadjizacharia; Hassan Aziz; Kara Snyder; Julie Wynne; Narong Kulvatunyou; Andrew Tang; Terence O'Keeffe; Rifat Latifi; Randall Friese; Peter Rhee
Journal:  World J Surg       Date:  2013-03       Impact factor: 3.352

9.  Accuracy of non-invasive haemoglobin measurements in patients undergoing transurethral resection of prostate surgery.

Authors:  Naveen Selvaraj; Gaurav Jain; Debendra Kumar Tripathy; Ankur Mittal; Haritha Indulekha
Journal:  Indian J Anaesth       Date:  2021-05-10

10.  Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children-a method comparison study.

Authors:  Aditya Bhat; Amit Upadhyay; Vijay Jaiswal; Deepak Chawla; Dharamveer Singh; Mithilesh Kumar; C P Yadav
Journal:  Eur J Pediatr       Date:  2015-08-19       Impact factor: 3.183

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.