The TALON removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy.

B J Salter1, M Fuss, D G Vollmer, A Sadeghi, C A Bogaev, D A Cheek, T S Herman, J M Hevezi.   

Abstract

PURPOSE: To present the TALON removable head frame system as an immobilization device for single-fraction intensity-modulated stereotactic radiosurgery (IMRS) and fractionated stereotactic intensity-modulated radiotherapy (FS-IMRT); and to evaluate the repositioning accuracy by measurement of anatomic landmark coordinates in repeated computed tomography (CT) examinations. METHODS AND MATERIALS: Nine patients treated by fractionated stereotactic intensity-modulated radiotherapy underwent repeated CTs during their treatment courses. We evaluated anatomic landmark coordinates in a total of 26 repeat CT data sets and respective x, y, and z shifts relative to their positions in the nine treatment-planning reference CTs. An iterative optimization algorithm was employed using a root mean square scoring function to determine the best-fit orientation of subsequent sets of anatomic landmark measurements relative to the original image set. This allowed for the calculation of the x, y, and z components of translation of the target isocenter for each repeat CT. In addition to absolute target isocenter translation, the magnitude (sum vector) of isocenter motion and the patient/target rotation about the three principal axes were calculated.
RESULTS: Anatomic landmark analysis over a treatment course of 6 weeks revealed a mean target isocenter translation of 0.95 +/- 0.55, 0.58 +/- 0.46, and 0.51 +/- 0.38 mm in x, y, and z directions, respectively. The mean magnitude of isocenter translation was 1.38 +/- 0.48 mm. The 95% confidence interval ([CI], mean translation plus two standard deviations) for repeated isocenter setup accuracy over the 6-week period was 2.34 mm. Average rotations about the x, y, and z axes were 0.41 +/- 0.36, 0.29 +/- 0.25, and 0.18 +/- 0.15 degrees, respectively. Analysis of the accuracy of the first repeated setup control, representative of single-fraction stereotactic radiosurgery situations, resulted in a mean target isocenter translation in the x, y, and z directions of 0.52 +/- 0.38, 0.56 +/- 0.30, and 0.46 +/- 0.25 mm, respectively. The mean magnitude of isocenter translation was 0.99 +/- 0.28 mm. The 95% confidence interval for these radiosurgery situations was 1.55 mm. Average rotations at first repeated setup control about the x, y, and z axes were 0.24 +/- 0.19, 0.19 +/- 0.17, and 0.19 +/- 0.12 degrees, respectively.
CONCLUSION: The TALON relocatable head frame was seen to be well suited for immobilization and repositioning of single-fraction stereotactic radiosurgery treatments. Because of its unique removable design, the system was also seen to provide excellent repeat immobilization and alignment for fractionated stereotactic applications. The exceptional accuracy for the single-fraction stereotactic radiosurgical application of the system was seen to deteriorate only slightly over a 6-week fractionated stereotactic treatment course.

Entities:  

Mesh:

Year:  2001        PMID: 11567832     DOI: 10.1016/s0360-3016(01)01670-4

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

Review 1.  Radiation medicine innovations for the new millenium.

Authors:  Dwight E Heron; Karen D Godette; Ray A Wynn; V Elayne Arterbery; Oscar A Streeter; Mack Roach; Joseph R Simpson; Melissa Blough; Charles R Thomas
Journal:  J Natl Med Assoc       Date:  2003-01       Impact factor: 1.798

2.  Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.

Authors:  Shidong Li; Lawrence R Kleinberg; Daniele Rigamonti; Moody D Wharam; Abdul Rashid; Juan Jackson; David Djajaputra; Shenjen He; Tunisia Creasey; Theodore L DeWeese
Journal:  Technol Cancer Res Treat       Date:  2010-12

3.  Rotational and translational reproducibility of newly developed Leksell frame-based relocatable fixation system.

Authors:  Etsuo Kunieda; Hossain M Deloar; Masayuki Kitamura; Osamu Kawaguchi; Hideyuki Shiba; Atsuya Takeda; Takatsugu Kawase; Satoshi Seki; Naoyuki Shigematsu; Atsushi Kubo
Journal:  Radiat Med       Date:  2006-08

4.  Relocatable fixation systems in intracranial stereotactic radiotherapy. Accuracy of serial CT scans and patient acceptance in a randomized design.

Authors:  A Theelen; J Martens; G Bosmans; R Houben; J J Jager; I Rutten; P Lambin; A W Minken; B G Baumert
Journal:  Strahlenther Onkol       Date:  2011-12-24       Impact factor: 3.621

5.  Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.

Authors:  Jürgen Wilbert; Matthias Guckenberger; Bülent Polat; Otto Sauer; Michael Vogele; Michael Flentje; Reinhart A Sweeney
Journal:  Radiat Oncol       Date:  2010-05-26       Impact factor: 3.481

6.  Clinical outcomes of patients treated with a second course of stereotactic radiosurgery for locally or regionally recurrent brain metastases after prior stereotactic radiosurgery.

Authors:  Daniel H Kim; Timothy E Schultheiss; Eric H Radany; Behnam Badie; Richard D Pezner
Journal:  J Neurooncol       Date:  2013-06-28       Impact factor: 4.130

7.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

8.  Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging.

Authors:  Imad Ali; Jesse Tubbs; Kerry Hibbitts; Ozer Algan; Spencer Thompson; Terence Herman; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2010-05-20       Impact factor: 2.102

9.  Analysis of patient repositioning accuracy in precision radiation therapy using automated image fusion.

Authors:  James L Robar; Brenda G Clark; Jason W Schella; Chang Seon Kim
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

10.  Quantitative evaluation of patient setup uncertainty of stereotactic radiotherapy with the frameless 6D ExacTrac system using statistical modeling.

Authors:  Vance Keeling; Sabbir Hossain; Hosang Jin; Ozer Algan; Salahuddin Ahmad; Imad Ali
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.