Question 2. Should carbon dioxide detectors be used to check correct placement of endotracheal tubes in preterm and term neonates?

Harsha Gowda1.   

Abstract

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22080462      PMCID: PMC3223393          DOI: 10.1136/archdischild-2011-300893

Source DB:  PubMed          Journal:  Arch Dis Child        ISSN: 0003-9888            Impact factor:   3.791


× No keyword cloud information.

Scenario

The Pedi-Cap device (Covidien, Mansfield, Massachusetts, USA) is frequently used in neonatal resuscitation to check the position of the endotracheal (ET) tube in term and preterm neonates. As a paediatric trainee having worked in various regions of the UK you note a huge variability in this practice. Clinical assessment of chest expansion and air entry, with improvement in saturations, colour and heart rate have been used for decades and work well. Is the Pedi-Cap superior to clinical assessment for checking the position of the ET tube?

Structured clinical question

During intubation of neonates [patients], is a carbon dioxide detector [intervention] better than clinical assessment [comparison] to detect correct endotracheal tube placement [outcome]?

Search strategy and outcome

Medline (1948–April week 1, 2011) and Embase (1947–15 April 2011) using the Ovid interface and the Cochrane Library were searched using the search terms: Neonate/Newborn/Preterm/Infants/Babies AND Carbon dioxide detector/CO2 detector/Pedi-Cap/Capnography/End tidal CO2 AND Intubation/Endotracheal tube; limits: Humans. The search of Medline yielded 41 articles and Embase yielded 43 articles (2 unique articles). No relevant reviews were found in the Cochrane Library. Thirty-nine articles were excluded, leaving four well conducted prospective studies for review (table 2).
Table 2

Should carbon dioxide detectors be used to check correct placement of endotracheal tubes in preterm and term neonates?

CitationStudy groupStudy typeOutcomeKey resultsComments
Hosono et al754 intubations in 40 neonates in the delivery room were analysedProspective cohortETT placement was compared using Pedi-Cap by an investigator not involved in the resuscitation, and by evaluation of clinical parameters by a resuscitation team unaware of the ETCO2 dataCapnography: sensitivity and specificity 100%Resuscitation team blinded to ETCO2 result. Investigator not involved in resuscitation
Clinical: sensitivity 92.5% and specificity 82.4%
Mean birth weight 839±263 gMean time for capnographic determination was significantly less than for clinical determination for both tracheal 7.5 (±1.3) s vs 17.0 (±3.4) s (p<0.01) and oesophageal intubation 6.5 (±0.7) s vs 19.9 (±1.8) s (p<0.01)
Mean gestational age 27.0±2.5 weeks
The clinical parameter was used as the gold standard
The times taken to detect accurate placement of the ETT using ETCO2 vs clinical determination of tracheal or oesophageal tube placement were compared
Repetto et al1027 intubations in 16 patients were analysedProspective cohortThe times taken using ETCO2 and clinical determinations of ETT placement in the delivery room were comparedThe median (range) times required for capnographic and clinical determination of tracheal intubation were 9 (4–26) s vs 35 (18–70) s (p<0.001), and for oesophageal intubation were 9 (4–17) s vs 30 (25–111) s (p=0.001)High rate of oesophageal intubation (11/27=40%)
Birth weight 575–2040 gOnly delivery room intubations
Hand-held, portable CO2 monitor providing CO2 readout was used instead of Pedi-Cap
Gestational age 23–34 weeks
Investigator not involved in resuscitation, or resuscitation team unaware of the ETCO2 data
Aziz et al445 newborns (450–4620 g) who needed endotracheal intubationProspective cohortAccuracy and ease of the Pedi-CapCorrelated in 30 of 33 patients (sensitivity 91%, specificity 100%)Resuscitation team blinded to the colour status of the Pedi-Cap
Pedi-Cap correlation with clinical evaluation and radiography findings for endotracheal intubation
Correlated in 12 of 12 patients (sensitivity, specificity, and PPV and NPV were all 100%)
Three false negatives with severe cardiorespiratory depression
34 (53.3%) were intubated in the delivery room and 21 (46.7%) in the NICU
Clinical evaluation: 0–90 s (mean 39.7±15.3 s)Negative result in CPR is not assessable as four infants needing most resuscitation were excluded
Pedi-Cap correlation with clinical evaluation for ETT in the oesophagusETCO2 detector: 4–12 s (mean 8.1±2.9 s) (p<0.001)
No measure of success of resuscitation
Comparison of time required to determine tube position by clinical evaluation and ETCO2
Roberts et al11100 intubations in 55 neonates in the NICU were studiedProspective cohortCapnography and clinical examination for identification of tube position were analysed40/100 intubation attempts resulted in oesophageal intubationUseful observational study and the intubating clinicians were blinded to the capnography
Capnography correctly identified oesophageal tube placement in 39/40 and did so in a mean of 1.6±2.4 s. Capnography failed to identify successful endotracheal intubation on only one occasion
Study was carried out in a neonatal unit and not in the delivery room
Mean birth weight 1419±811 gThe time required for establishing by clinical confirmation whether the tube was in the trachea or not was compared to that required for capnography
A hand-held, portable CO2 monitor was used instead of Pedi-Cap
Mean gestational age 28.5±4.0 weeks
Clinical indicators of tube position required 97.1+/–92.6 s to identify oesophageal intubation and failed to identify successful endotracheal intubation in 5 of 60 cases

ETCO2, end tidal CO2; ETT, endotracheal tube; NICU, neonatal intensive care unit; NPV, negative predictive value; PPV, positive predictive value.

Should carbon dioxide detectors be used to check correct placement of endotracheal tubes in preterm and term neonates? ETCO2, end tidal CO2; ETT, endotracheal tube; NICU, neonatal intensive care unit; NPV, negative predictive value; PPV, positive predictive value. There were no randomised controlled trials.

Commentary

Proper placement of the ET tube during resuscitation can be difficult, especially in neonates, and evidence suggests a significant rate of oesophageal intubation when neonatal tracheal intubation is attempted: the rates of successful intubation at the first attempt vary from 24% in junior trainees to 86% in consultants.1 Direct laryngoscopy and observation of the ET tube passing between the vocal cords is the standard criterion for verifying ET intubation.2 Detection of end-tidal carbon dioxide, however, serves as a valuable adjunct to confirm ET intubation, detect inadvertent oesophageal intubation and monitor for accidental tracheal extubation. Many studies have shown the colorimetric ETCO2 (end-tidal carbon dioxide) detector to be sensitive and specific in confirming ET intubation in haemodynamically stable adults and children.3 However, there has always been a concern that carbon dioxide in a neonatal small tidal volume may be diluted in the large dead space of the early versions of these detectors, resulting in false negative results (ie, indicating oesophageal placement despite the correct intratracheal position of the ET tube). Therefore, a specific paediatric disposable colorimetric ETCO2 detector (Pedi-Cap) with an internal volume of 3 ml is used during neonatal intubation.4 In direct comparisons in adults, capnography was superior to clinical assessment but no single technique was perfect, and capnography was found to be less accurate in cardiac arrest.5 In paediatric patients weighing more than 2 kg and with spontaneous circulation, detection of exhaled carbon dioxide confirmed tracheal tube position in all cases, but during cardiac arrest the possibility of a false negative result required further confirmation of tracheal tube position.6 Four good quality neonatal studies4 7 10 11 found that capnography/Pedi-Cap identified tracheal tube position more rapidly than clinical assessment. In all studies direct visualisation of tracheal tube position (or clinical assessment) was used as the final ‘gold standard’. Hosono et al compared capnography with defined clinical assessments. Capnography was completely accurate in all babies studied, all of whom had spontaneous circulation and were less than 32 weeks gestation. This study also had a well defined method for defining tracheal tube position.7 All studies utilised a separate team to measure exhaled carbon dioxide, with the clinical team blinded to the measurements, and all four examined neonates with spontaneous circulation. Several cases of false negatives in neonates as well as false negatives occurring in adult and paediatric cardiac arrest have been reported.8 Therefore, capnography should be interpreted carefully in extremely small neonates or in those in whom extensive resuscitation is required. All studies showed that detection of exhaled carbon dioxide confirms tracheal intubation in neonates with a cardiac output more rapidly and more accurately than clinical assessment alone. False negatives may occur in very low birthweight neonates and in those in cardiac arrest.8 False positives may occur in the presence of colorimetric devices contaminated with epinephrine (adrenaline), surfactant or atropine.9 There is no comparative information to recommend any one method for the detection of exhaled carbon dioxide in the neonatal population. It appears important to use ETCO2 detection during neonatal intubation. Detection of exhaled carbon dioxide confirms tracheal intubation in neonates with a cardiac output more rapidly and more accurately than clinical assessment alone. (Grade B) False negatives may occur in neonates with cardiac arrest. (Grade C) It is unclear if false positives occur with colorimetric devices contaminated with epinephrine, surfactant or atropine. (Grade D)
  11 in total

Review 1.  Esophageal intubation: a review of detection techniques.

Authors:  P K Birmingham; F W Cheney; R J Ward
Journal:  Anesth Analg       Date:  1986-08       Impact factor: 5.108

2.  The pediatric disposable end-tidal carbon dioxide detector role in endotracheal intubation in newborns.

Authors:  H F Aziz; J B Martin; J J Moore
Journal:  J Perinatol       Date:  1999-03       Impact factor: 2.521

3.  The use of capnography for recognition of esophageal intubation in the neonatal intensive care unit.

Authors:  W A Roberts; W M Maniscalco; A R Cohen; R S Litman; A Chhibber
Journal:  Pediatr Pulmonol       Date:  1995-05

4.  Nellcor Stat Cap differentiates oesophageal from tracheal intubation.

Authors:  P D Sutherland; M Quinn
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1995-11       Impact factor: 5.747

5.  Use of capnography in the delivery room for assessment of endotracheal tube placement.

Authors:  J E Repetto; P K Donohue PA-C; S F Baker; L Kelly; L M Nogee
Journal:  J Perinatol       Date:  2001 Jul-Aug       Impact factor: 2.521

6.  Colorimetric end-tidal carbon dioxide detectors in the delivery room: strengths and limitations. A case report.

Authors:  C Omar F Kamlin; Colm P F O'Donnell; Peter G Davis; Colin J Morley
Journal:  J Pediatr       Date:  2005-10       Impact factor: 4.406

7.  Comparison of three different methods to confirm tracheal tube placement in emergency intubation.

Authors:  Stefek Grmec
Journal:  Intensive Care Med       Date:  2002-04-30       Impact factor: 17.440

8.  False-positive results on colorimetric carbon dioxide analysis in neonatal resuscitation: potential for serious patient harm.

Authors:  S M Hughes; B L Blake; S L Woods; C U Lehmann
Journal:  J Perinatol       Date:  2007-12       Impact factor: 2.521

9.  Validity of a disposable end-tidal CO2 detector in verifying endotracheal tube placement in infants and children.

Authors:  M S Bhende; A E Thompson; D R Cook; A L Saville
Journal:  Ann Emerg Med       Date:  1992-02       Impact factor: 5.721

10.  A role of end-tidal CO(2) monitoring for assessment of tracheal intubations in very low birth weight infants during neonatal resuscitation at birth.

Authors:  Shigeharu Hosono; Ikuhiro Inami; Hidetoshi Fujita; Michiyoshi Minato; Shigeru Takahashi; Hideo Mugishima
Journal:  J Perinat Med       Date:  2009       Impact factor: 1.901