Literature DB >> 1739199

Validity of a disposable end-tidal CO2 detector in verifying endotracheal tube placement in infants and children.

M S Bhende1, A E Thompson, D R Cook, A L Saville.   

Abstract

STUDY
OBJECTIVE: To examine the validity of a disposable, colorimetric end-tidal CO2 detector in verifying endotracheal tube (ETT) placement in infants and children.
DESIGN: The detector was studied prospectively in 151 intubations.
SETTING: Operating room, ICU, and emergency department of a children's hospital. PARTICIPANTS: One hundred thirty-seven children undergoing endotracheal intubation for anesthesia (52), respiratory support (76), or CPR (23).
INTERVENTIONS: After endotracheal intubation, tube position was verified, the detector was attached, and readings were obtained. MEASUREMENTS AND
RESULTS: The detector correctly identified tube position (trachea, 124; esophagus, four) in all 120 patients who were not in cardiac arrest (P less than .01). In the cardiac arrest setting, all six esophageal intubations were correctly identified, but two of the 17 tracheal intubations were incorrectly interpreted as esophageal intubations (P less than .01).
CONCLUSION: The detector accurately identifies ETT position in children with spontaneous circulation who weigh more than 2 kg. During CPR, a positive test correctly indicates that the ETT is in the airway, but a negative result (suggesting esophageal placement) requires an alternate means of confirming ETT position.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1739199     DOI: 10.1016/s0196-0644(05)80148-9

Source DB:  PubMed          Journal:  Ann Emerg Med        ISSN: 0196-0644            Impact factor:   5.721


  11 in total

1.  Part 10: Pediatric basic and advanced life support: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations.

Authors:  Monica E Kleinman; Allan R de Caen; Leon Chameides; Dianne L Atkins; Robert A Berg; Marc D Berg; Farhan Bhanji; Dominique Biarent; Robert Bingham; Ashraf H Coovadia; Mary Fran Hazinski; Robert W Hickey; Vinay M Nadkarni; Amelia G Reis; Antonio Rodriguez-Nunez; James Tibballs; Arno L Zaritsky; David Zideman
Journal:  Circulation       Date:  2010-10-19       Impact factor: 29.690

2.  Pediatric basic and advanced life support: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science with Treatment Recommendations.

Authors:  Monica E Kleinman; Allan R de Caen; Leon Chameides; Dianne L Atkins; Robert A Berg; Marc D Berg; Farhan Bhanji; Dominique Biarent; Robert Bingham; Ashraf H Coovadia; Mary Fran Hazinski; Robert W Hickey; Vinay M Nadkarni; Amelia G Reis; Antonio Rodriguez-Nunez; James Tibballs; Arno L Zaritsky; David Zideman
Journal:  Pediatrics       Date:  2010-10-18       Impact factor: 7.124

Review 3.  Are carbon dioxide detectors useful in neonates?

Authors:  E J Molloy; K Deakins
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2006-07       Impact factor: 5.747

4.  A novel airway device with tactile sensing capabilities for verifying correct endotracheal tube placement.

Authors:  Pauwel Goethals; Harshu Chaobal; Dominiek Reynaerts; David Schaner
Journal:  J Clin Monit Comput       Date:  2013-09-25       Impact factor: 2.502

5.  Combined use of the esophageal-tracheal Combitube with a colorimetric carbon dioxide detector for emergency intubation/ventilation.

Authors:  B D Butler; T Little; S Drtil
Journal:  J Clin Monit       Date:  1995-09

6.  T-piece gas flow palpation as a clinical indicator of endotracheal intubation in neonates.

Authors:  Peter M Moran; Colin Patrick Hawkes; Eugene Michael Dempsey; C Anthony Ryan
Journal:  Eur J Pediatr       Date:  2012-12-28       Impact factor: 3.183

Review 7.  Question 2. Should carbon dioxide detectors be used to check correct placement of endotracheal tubes in preterm and term neonates?

Authors:  Harsha Gowda
Journal:  Arch Dis Child       Date:  2011-12       Impact factor: 3.791

Review 8.  WITHDRAWN: Techniques to ascertain correct endotracheal tube placement in neonates.

Authors:  Georg M Schmölzer; Charles Christoph C Roehr
Journal:  Cochrane Database Syst Rev       Date:  2018-07-05

9.  A feasibility study on bedside upper airway ultrasonography compared to waveform capnography for verifying endotracheal tube location after intubation.

Authors:  Osman Adi; Tan Wan Chuan; Manikam Rishya
Journal:  Crit Ultrasound J       Date:  2013-07-04

Review 10.  Current methodological and technical limitations of time and volumetric capnography in newborns.

Authors:  Gerd Schmalisch
Journal:  Biomed Eng Online       Date:  2016-08-30       Impact factor: 2.819

View more