Biomechanical evaluation of the X-Stop device for surgical treatment of lumbar spinal stenosis.

Zongmiao Wan1, Shaobai Wang, Michal Kozánek, Peter G Passias, Frederick L Mansfield, Kirkham B Wood, Guoan Li.   

Abstract

STUDY
DESIGN: Controlled experimental study.
OBJECTIVE: To evaluate the kinematical effects of X-Stop device on the spinal process at the operated and the adjacent segments before and after X-Stop surgeries during various weight-bearing postures in elderly patients with lumbar spine stenosis. SUMMARY OF BACKGROUND DATA: The mechanism of interspinous process (ISP) devices is to directly distract the ISP of the implanted level to indirectly decompress the intervertebra foramen and spinal canal. Few studies have investigated the changes of ISP gap caused by X-Stop implantation using magnetic resonance imaging or radiography, but the effect of X-Stop surgery on the kinematics of spinous processes during functional activities is still unclear.
METHODS: Eight patients were tested before and, on average, 7 months after surgical implantation of the X-Stop devices using a combined computed tomography/magnetic resonance imaging and dual fluoroscopic imaging system during weight-bearing standing, flexion-extension, left-right bending, and left-right twisting positions of the torso. The shortest distances of the ISPs at the operated and the adjacent levels were measured using iterative closest point method and was dissected into vertical (gap) and horizontal (lateral translation) components.
RESULTS: At the operated levels, the shortest vertical ISP distances (gap) significantly (P<0.05) increased by 1.5 mm during standing, 1.2 mm during left twist, 1.3 mm during extension, and 1.1 mm during flexion, whereas they also increased yet not significantly (P>0.05) in right twist, left bend, and right bend after the X-Stop implantation. The lateral translations were not significantly affected. At both cephalad and caudad adjacent levels, the ISP distances (vertical and horizontal) were not significantly affected during all postures after X-Stop implantation.
CONCLUSION: The findings of this study indicate that implantation of the X-Stop devices can effectively distract the ISP space at the diseased level without causing apparent kinematic changes at the adjacent segments during the studied postures.

Entities:  

Mesh:

Year:  2012        PMID: 22015626      PMCID: PMC3740369          DOI: 10.1097/BSD.0b013e318227eb84

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  23 in total

1.  Determination of inter-spinous process distance in the lumbar spine. Evaluation of reference population to facilitate detection of severe trauma.

Authors:  P Neumann; Y Wang; J Kärrholm; H Malchau; A Nordwall
Journal:  Eur Spine J       Date:  1999       Impact factor: 3.134

2.  The effects of an interspinous implant on intervertebral disc pressures.

Authors:  Kyle E Swanson; Derek P Lindsey; Ken Y Hsu; James F Zucherman; Scott A Yerby
Journal:  Spine (Phila Pa 1976)       Date:  2003-01-01       Impact factor: 3.468

3.  The effect of an interspinous process implant on facet loading during extension.

Authors:  Craig M Wiseman; Derek P Lindsey; Amy D Fredrick; Scott A Yerby
Journal:  Spine (Phila Pa 1976)       Date:  2005-04-15       Impact factor: 3.468

4.  In vivo articular cartilage contact kinematics of the knee: an investigation using dual-orthogonal fluoroscopy and magnetic resonance image-based computer models.

Authors:  Guoan Li; Louis E DeFrate; Sang Eun Park; Thomas J Gill; Harry E Rubash
Journal:  Am J Sports Med       Date:  2005-01       Impact factor: 6.202

5.  Anatomic changes of the spinal canal and intervertebral foramen associated with flexion-extension movement.

Authors:  A Inufusa; H S An; T H Lim; T Hasegawa; V M Haughton; B H Nowicki
Journal:  Spine (Phila Pa 1976)       Date:  1996-11-01       Impact factor: 3.468

6.  A multicenter, prospective, randomized trial evaluating the X STOP interspinous process decompression system for the treatment of neurogenic intermittent claudication: two-year follow-up results.

Authors:  James F Zucherman; Ken Y Hsu; Charles A Hartjen; Thomas F Mehalic; Dante A Implicito; Michael J Martin; Donald R Johnson; Grant A Skidmore; Paul P Vessa; James W Dwyer; Stephen T Puccio; Joseph C Cauthen; Richard M Ozuna
Journal:  Spine (Phila Pa 1976)       Date:  2005-06-15       Impact factor: 3.468

7.  The treatment mechanism of an interspinous process implant for lumbar neurogenic intermittent claudication.

Authors:  Joshua C Richards; Sharmila Majumdar; Derek P Lindsey; Gary S Beaupré; Scott A Yerby
Journal:  Spine (Phila Pa 1976)       Date:  2005-04-01       Impact factor: 3.468

8.  Back pain, neurogenic symptoms, and physical function in relation to spondylolisthesis among elderly men.

Authors:  Patrick J Denard; Kathleen F Holton; Jessica Miller; Howard A Fink; Deborah M Kado; Lynn M Marshall; Jung U Yoo
Journal:  Spine J       Date:  2010-10       Impact factor: 4.166

9.  A prospective randomized multi-center study for the treatment of lumbar spinal stenosis with the X STOP interspinous implant: 1-year results.

Authors:  J F Zucherman; K Y Hsu; C A Hartjen; T F Mehalic; D A Implicito; M J Martin; D R Johnson; G A Skidmore; P P Vessa; J W Dwyer; S Puccio; J C Cauthen; R M Ozuna
Journal:  Eur Spine J       Date:  2003-12-19       Impact factor: 3.134

10.  An interspinous process distractor (X STOP) for lumbar spinal stenosis in elderly patients: preliminary experiences in 10 consecutive cases.

Authors:  Jangbo Lee; Kazutoshi Hida; Toshitaka Seki; Yoshinobu Iwasaki; Akino Minoru
Journal:  J Spinal Disord Tech       Date:  2004-02
View more
  8 in total

1.  Motion characteristics of the lumbar spinous processes with degenerative disc disease and degenerative spondylolisthesis.

Authors:  Qi Yao; Shaobai Wang; Jae-Hyuk Shin; Guoan Li; Kirkham Wood
Journal:  Eur Spine J       Date:  2013-08-02       Impact factor: 3.134

2.  Biomechanical analysis of a new lumbar interspinous device with optimized topology.

Authors:  Chen-Sheng Chen; Shih-Liang Shih
Journal:  Med Biol Eng Comput       Date:  2018-01-06       Impact factor: 2.602

3.  Segmental spinal canal volume in patients with degenerative spondylolisthesis.

Authors:  Jun Miao; Shaobai Wang; Won Man Park; Qun Xia; Xiutong Fang; Martin P Torriani; Kirkham B Wood; Guoan Li
Journal:  Spine J       Date:  2013-03-27       Impact factor: 4.166

4.  Magnetic resonance imaging on disc degeneration changes after implantation of an interspinous spacer and fusion of the adjacent segment.

Authors:  Xiaokang Liu; Yingjie Liu; Xiaofeng Lian; Jianguang Xu
Journal:  Int J Clin Exp Med       Date:  2015-04-15

5.  Biomechanical effect of Coflex and X-STOP spacers on the lumbar spine: a finite element analysis.

Authors:  Zhiyuan Guo; Guangfei Liu; Lu Wang; Yuejiang Zhao; Ye Zhao; Shouliang Lu; Cai Cheng
Journal:  Am J Transl Res       Date:  2022-07-15       Impact factor: 3.940

Review 6.  Biomechanics of interspinous devices.

Authors:  Paolo D Parchi; Gisberto Evangelisti; Antonella Vertuccio; Nicola Piolanti; Lorenzo Andreani; Valentina Cervi; Christian Giannetti; Giuseppe Calvosa; Michele Lisanti
Journal:  Biomed Res Int       Date:  2014-07-09       Impact factor: 3.411

7.  Kinematic characteristics of lumbar spinous processes during axial rotation in patients with lumbar degenerative disc disease lateral lumbar interbody fusion and intervention.

Authors:  BaoLiang Zhang; Wenhan Huang; Hong Xia; Xinglong Feng
Journal:  BMC Musculoskelet Disord       Date:  2017-04-04       Impact factor: 2.362

8.  Dynamic stabilization for challenging lumbar degenerative diseases of the spine: a review of the literature.

Authors:  Tuncay Kaner; Ali Fahir Ozer
Journal:  Adv Orthop       Date:  2013-04-15
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.