A tetrahedron beam computed tomography benchtop system with a multiple pixel field emission x-ray tube.

Xiaochao Xu1, Joshua Kim, Philip Laganis, Derek Schulze, Yongguang Liang, Tiezhi Zhang.   

Abstract

PURPOSE: To demonstrate the feasibility of Tetrahedron Beam Computed Tomography (TBCT) using a carbon nanotube (CNT) multiple pixel field emission x-ray (MPFEX) tube.
METHODS: A multiple pixel x-ray source facilitates the creation of novel x-ray imaging modalities. In a previous publication, the authors proposed a Tetrahedron Beam Computed Tomography (TBCT) imaging system which comprises a linear source array and a linear detector array that are orthogonal to each other. TBCT is expected to reduce scatter compared with Cone Beam Computed Tomography (CBCT) and to have better detector performance. Therefore, it may produce improved image quality for image guided radiotherapy. In this study, a TBCT benchtop system has been developed with an MPFEX tube. The tube has 75 CNT cold cathodes, which generate 75 x-ray focal spots on an elongated anode, and has 4 mm pixel spacing. An in-house-developed, 5-row CT detector array using silicon photodiodes and CdWO(4) scintillators was employed in the system. Hardware and software were developed for tube control and detector data acquisition. The raw data were preprocessed for beam hardening and detector response linearity and were reconstructed with an FDK-based image reconstruction algorithm.
RESULTS: The focal spots were measured at about 1 × 2 mm(2) using a star phantom. Each cathode generates around 3 mA cathode current with 2190 V gate voltage. The benchtop system is able to perform TBCT scans with a prolonged scanning time. Images of a commercial CT phantom were successfully acquired.
CONCLUSIONS: A prototype system was developed, and preliminary phantom images were successfully acquired. MPFEX is a promising x-ray source for TBCT. Further improvement of tube output is needed in order for it to be used in clinical TBCT systems.

Entities:  

Mesh:

Year:  2011        PMID: 21992368      PMCID: PMC3195373          DOI: 10.1118/1.3634043

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

1.  Cone-beam computed tomography with a flat-panel imager: effects of image lag.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2001-02       Impact factor: 4.071

3.  Effect of scattered radiation on image noise in cone beam CT.

Authors:  M Endo; T Tsunoo; N Nakamori; K Yoshida
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

4.  Optimization of x-ray imaging geometry (with specific application to flat-panel cone-beam computed tomography).

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

5.  Flat-panel cone-beam computed tomography for image-guided radiation therapy.

Authors:  David A Jaffray; Jeffrey H Siewerdsen; John W Wong; Alvaro A Martinez
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-08-01       Impact factor: 7.038

6.  Dose and image quality for a cone-beam C-arm CT system.

Authors:  Rebecca Fahrig; Robert Dixon; Thomas Payne; Richard L Morin; Arundhuti Ganguly; Norbert Strobel
Journal:  Med Phys       Date:  2006-12       Impact factor: 4.071

7.  Scanning-beam digital x-ray (SBDX) technology for interventional and diagnostic cardiac angiography.

Authors:  Michael A Speidel; Brian P Wilfley; Josh M Star-Lack; Joseph A Heanue; Michael S Van Lysel
Journal:  Med Phys       Date:  2006-08       Impact factor: 4.071

8.  A dynamic micro-CT scanner based on a carbon nanotube field emission x-ray source.

Authors:  G Cao; Y Z Lee; R Peng; Z Liu; R Rajaram; X Calderon-Colon; L An; P Wang; T Phan; S Sultana; D S Lalush; J P Lu; O Zhou
Journal:  Phys Med Biol       Date:  2009-03-25       Impact factor: 3.609

9.  Cone-beam-CT guided radiation therapy: technical implementation.

Authors:  Daniel Létourneau; John W Wong; Mark Oldham; Misbah Gulam; Lindsay Watt; David A Jaffray; Jeffrey H Siewerdsen; Alvaro A Martinez
Journal:  Radiother Oncol       Date:  2005-06       Impact factor: 6.280

10.  Design and characterization of a spatially distributed multibeam field emission x-ray source for stationary digital breast tomosynthesis.

Authors:  Xin Qian; Ramya Rajaram; Xiomara Calderon-Colon; Guang Yang; Tuyen Phan; David S Lalush; Jianping Lu; Otto Zhou
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

View more
  3 in total

1.  Technical note: algebraic iterative image reconstruction using a cylindrical image grid for tetrahedron beam computed tomography.

Authors:  Joshua Kim; Dan Ionascu; Tiezhi Zhang
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

2.  Studies of a prototype linear stationary x-ray source for tomosynthesis imaging.

Authors:  P R Schwoebel; John M Boone; Joe Shao
Journal:  Phys Med Biol       Date:  2014-04-17       Impact factor: 3.609

3.  Evaluation of algebraic iterative image reconstruction methods for tetrahedron beam computed tomography systems.

Authors:  Joshua Kim; Huaiqun Guan; David Gersten; Tiezhi Zhang
Journal:  Int J Biomed Imaging       Date:  2013-05-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.