Cone-beam-CT guided radiation therapy: technical implementation.

Daniel Létourneau1, John W Wong, Mark Oldham, Misbah Gulam, Lindsay Watt, David A Jaffray, Jeffrey H Siewerdsen, Alvaro A Martinez.   

Abstract

BACKGROUND AND
PURPOSE: X-ray volumetric imaging system (XVI) mounted on a linear accelerator is available for image guidance applications. In preparation for clinical implementation, phantom and patient imaging studies were conducted to determine the irradiation parameters that would trade-off image quality, patient dose and scanning time. PATIENTS AND METHODS: The XVI image quality and imaging dose were benchmarked against those obtained with a helical CT scanner for a head and body phantom. The irradiation parameters were varied including the total imaging dose, number of projections, field of view, reconstruction resolution and use of a scatter rejection grid. We characterized the image quality based on relative contrast, noise, contrast to noise ratio (CNR) and point spread function (PSF). XVI scans of pelvis, head and neck and lung patients were acquired and submitted to a range of observers to identify the favorable reconstruction parameters.
RESULTS: Phantom studies have demonstrated that a scatter rejection grid reduces photon scattering and improves the image uniformity. For the body phantom, the helical CT and the wide field XVI technique produce similar image quality, with surface doses of 0.025 and 0.044 Gy respectively. We have demonstrated that the local tomography technique improves the image contrast and the CNR while reducing the skin dose by 40-50% compared to the wide field technique. Clinical scans of head and neck, lung and prostate patients present good soft tissue contrast and excellent bone definition.
CONCLUSIONS: With adjustment of irradiation parameters and an imaging surface dose of less than 0.05 Gy, high quality XVI images can be obtained for a phantom simulating the body thickness. XVI is currently feasible for image-guided treatments of head and neck, torso and pelvic areas using soft tissue and bony structures.

Entities:  

Mesh:

Year:  2005        PMID: 15890424     DOI: 10.1016/j.radonc.2005.03.001

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  55 in total

1.  Image-based modeling of tumor shrinkage in head and neck radiation therapy.

Authors:  Ming Chao; Yaoqin Xie; Eduardo G Moros; Quynh-Thu Le; Lei Xing
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

Review 2.  Image-guided radiation therapy for muscle-invasive bladder cancer.

Authors:  Juliette Thariat; Shafak Aluwini; Qiong Pan; Mickael Caullery; Pierre-Yves Marcy; Martin Housset; Jean-Leon Lagrange
Journal:  Nat Rev Urol       Date:  2011-11-08       Impact factor: 14.432

3.  Evaluation of vascular supply with cone-beam computed tomography during intraarterial chemotherapy for a skull base tumor.

Authors:  Reiichi Ishikura; Kumiko Ando; Yuki Nagami; Satoshi Yamamoto; Koui Miura; Ajaya Raj Pande; Tosyiko Yamano; Shozo Hirota; Norio Nakao
Journal:  Radiat Med       Date:  2006-06

4.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

5.  Region-of-interest image reconstruction with intensity weighting in circular cone-beam CT for image-guided radiation therapy.

Authors:  Seungryong Cho; Erik Pearson; Charles A Pelizzari; Xiaochuan Pan
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

6.  Quality and accuracy of cone beam computed tomography gated by active breathing control.

Authors:  Bria P Thompson; Geoffrey D Hugo
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

7.  Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: a feasibility and accuracy study.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

8.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT.

Authors:  M J Daly; J H Siewerdsen; Y B Cho; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

9.  Optimizing monoscopic kV fluoro acquisition for prostate intrafraction motion evaluation.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Phys Med Biol       Date:  2008-12-10       Impact factor: 3.609

10.  Inferences about prostate intrafraction motion from pre- and posttreatment volumetric imaging.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-06-08       Impact factor: 7.038

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.