Physical performance of the new hybrid PET∕CT Discovery-690.

V Bettinardi1, L Presotto, E Rapisarda, M Picchio, L Gianolli, M C Gilardi.   

Abstract

PURPOSE: The aim of this work was the assessment of the physical performance of the new hybrid PET∕CT system: Discovery-690.
METHODS: The Discovery-690 combines a lutetium-yttrium-orthosilicate (LYSO) block detector designed PET tomograph with a 64-slice CT scanner. The system is further characterized by a dedicated powerful computing platform implementing fully 3D-PET iterative reconstruction algorithms. These algorithms can account for time of flight (TOF) information and∕or a 3D model of the PET point spread function (PSF). PET physical performance was measured following NEMA NU-2-2007 procedures. Furthermore, specific tests were used: (i) to measure the energy and timing resolution of the PET system and (ii) to evaluate image quality, by using phantoms representing different clinical conditions (e.g., brain and whole body). Data processing and reconstructions were performed as required by standard procedures. Further reconstructions were carried out to evaluate the performance of the new reconstruction algorithms. In particular, four algorithms were considered for the reconstruction of the PET data: (i) HD = standard configuration, without TOF and PSF, (ii) TOF = HD + TOF, (iii) PSF = HD + PSF, and (iv) TOFPSF = HD + TOF + PSF.
RESULTS: The transverse (axial) spatial resolution values were 4.70 (4.74) mm and 5.06 (5.55) mm at 1 cm and 10 cm off axis, respectively. Sensitivity (average between 0 and 10 cm) was 7.5 cps∕kBq. The noise equivalent count rate (NECR) peak was 139.1 kcps at 29.0 kBq∕ml. The scatter fraction at the NECR peak was 37%. The correction accuracy for the dead time losses and random event counts had a maximum absolute error below the NECR peak of 2.09%. The average energy and timing resolution were 12.4% and 544.3 ps, respectively. PET image quality was evaluated with the NEMA IEC Body phantom by using four reconstruction algorithms (HD, TOF, PSF, and TOFPSF), as previously described. The hot contrast (after 3 iterations and for a lesion∕background activity ratio of 4:1) for the spheres of 10, 13, 17, and 22 mm was (HD) 29.8, 45.4, 55.4, and 68.1%; (TOF) 39.9, 53.5, 62.7, and 72.2%; (PSF) 28.3, 47.3, 60.4, and 71.8%; (TOFPSF) 43.8, 62.9, 70.6, and 76.4%. The cold contrast for the spheres of 28 and 37 mm was (HD) 62.4 and 65.2%; (TOF) 77.1 and 81.4%; (PSF) 62.0 and 65.2%; (TOFPSF) 77.3 and 81.6%. Similar hot and cold contrast trends were found during the analyses of other phantoms representing different clinical conditions (brain and whole body). Nevertheless, the authors observed a predominant role of either TOF or PSF, depending on the specific characteristics and dimensions of the phantoms.
CONCLUSIONS: Discovery-690 shows very good PET physical performance for all the standard NEMA NU-2-2007 measurements. Furthermore, the new reconstruction algorithms available for PET data (TOF and PSF) allow further improvements of the D-690 image quality performance both qualitatively and quantitatively.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21992359     DOI: 10.1118/1.3635220

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  101 in total

1.  Evaluation of Hamamatsu PET imaging modules for dedicated TOF-capable scanners.

Authors:  A Stolin; G Jaliparthi; R R Raylman; J Brefczynski-Lewis; S Majewski; J Qi; K Gong; S Dolinsky
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2019-01-23

2.  Comparative analysis of iterative reconstruction algorithms with resolution recovery and time of flight modeling for 18F-FDG cardiac PET: A multi-center phantom study.

Authors:  Roberta Matheoud; Michela Lecchi; Domenico Lizio; Camilla Scabbio; Claudio Marcassa; Lucia Leva; Angelo Del Sole; Carlo Rodella; Luca Indovina; Christian Bracco; Marco Brambilla; Orazio Zoccarato
Journal:  J Nucl Cardiol       Date:  2016-01-12       Impact factor: 5.952

Review 3.  Characterization of the impact to PET quantification and image quality of an anterior array surface coil for PET/MR imaging.

Authors:  Scott D Wollenweber; Gaspar Delso; Timothy Deller; David Goldhaber; Martin Hüllner; Patrick Veit-Haibach
Journal:  MAGMA       Date:  2013-06-26       Impact factor: 2.310

4.  Simultaneous reconstruction of attenuation and activity in cardiac PET can remove CT misalignment artifacts.

Authors:  L Presotto; E Busnardo; D Perani; L Gianolli; M C Gilardi; V Bettinardi
Journal:  J Nucl Cardiol       Date:  2015-08-15       Impact factor: 5.952

Review 5.  Innovations in Instrumentation for Positron Emission Tomography.

Authors:  Eric Berg; Simon R Cherry
Journal:  Semin Nucl Med       Date:  2018-03-12       Impact factor: 4.446

6.  DOI Determination by Rise Time Discrimination in Single-Ended Readout for TOF PET Imaging.

Authors:  R I Wiener; S Surti; J S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2013-06       Impact factor: 1.679

7.  Staging the axilla in breast cancer patients with ¹⁸F-FDG PET: how small are the metastases that we can detect with new generation clinical PET systems?

Authors:  Dimitri Bellevre; Cécile Blanc Fournier; Odile Switsers; Audrey Emmanuelle Dugué; Christelle Levy; Djelila Allouache; Cédric Desmonts; Hubert Crouet; Jean-Marc Guilloit; Jean-Michel Grellard; Nicolas Aide
Journal:  Eur J Nucl Med Mol Imaging       Date:  2014-02-22       Impact factor: 9.236

8.  Data-Driven Respiratory Gating Outperforms Device-Based Gating for Clinical 18F-FDG PET/CT.

Authors:  Matthew D Walker; Andrew J Morgan; Kevin M Bradley; Daniel R McGowan
Journal:  J Nucl Med       Date:  2020-04-03       Impact factor: 10.057

9.  Resolution Enhancement in PET Reconstruction Using Collimation.

Authors:  Scott D Metzler; Samuel Matej; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

10.  Impact of image reconstruction methods on quantitative accuracy and variability of FDG-PET volumetric and textural measures in solid tumors.

Authors:  Ali Ketabi; Pardis Ghafarian; Mohammad Amin Mosleh-Shirazi; Seyed Rabi Mahdavi; Arman Rahmim; Mohammad Reza Ay
Journal:  Eur Radiol       Date:  2018-10-02       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.