Characterization of the deflections of a catheter steered using a magnetic resonance imaging system.

Frederick P Gosselin1, Viviane Lalande, Sylvain Martel.   

Abstract

PURPOSE: The authors quantify the deflections of a catheter and a guidewire in MR setting with different designs of ferromagnetic tips and a system of high gradient coils which can generate gradients, and thus forces, 20 times larger than a conventional scanner.
METHODS: Different designs of catheter tips are experimentally tested in an effort to maximize the deflections. One to two ferromagnetic spheres are attached at the distal tip of the catheter (or guidewire) with different spacing between the spheres. The effect of dipole-dipole interaction on the steering of the catheter is studied through experimentation and theoretical modeling. The effect of using many spheres on the artefact generated in fast imaging sequences is also investigated.
RESULTS: A catheter and a guidewire are successfully steered by applying magnetic gradients inside a magnetic resonance scanner. More ferromagnetic material allows for larger magnetic forces, however, the use of two ferromagnetic spheres introduces undesired dipole-dipole interactions. Two ferromagnetic spheres generate a single larger artefact as they are close together.
CONCLUSIONS: By varying the distance between the two ferromagnetic spheres, a balance can be struck between the need to minimize the size of the tip and the undesirable dipole-dipole interaction.

Entities:  

Mesh:

Year:  2011        PMID: 21978043     DOI: 10.1118/1.3622599

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy.

Authors:  Parham Moftakhar; Prasheel Lillaney; Aaron D Losey; Daniel L Cooke; Alastair J Martin; Bradford R H Thorne; Ronald L Arenson; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  Radiology       Date:  2015-06-01       Impact factor: 11.105

2.  Design of a Magnetic Resonance Imaging Guided Magnetically Actuated Steerable Catheter.

Authors:  Taoming Liu; Nate Lombard Poirot; Tipakorn Greigarn; M Cenk Çavuşoğlu
Journal:  J Med Device       Date:  2017-05-03       Impact factor: 0.582

3.  Endovascular MR-guided Renal Embolization by Using a Magnetically Assisted Remote-controlled Catheter System.

Authors:  Prasheel V Lillaney; Jeffrey K Yang; Aaron D Losey; Alastair J Martin; Daniel L Cooke; Bradford R H Thorne; David C Barry; Andrew Chu; Carol Stillson; Loi Do; Ronald L Arenson; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  Radiology       Date:  2016-03-28       Impact factor: 11.105

4.  State of the Art and Future Opportunities in MRI-Guided Robot-Assisted Surgery and Interventions.

Authors:  Hao Su; Ka-Wai Kwok; Kevin Cleary; Iulian Iordachita; M Cenk Cavusoglu; Jaydev P Desai; Gregory S Fischer
Journal:  Proc IEEE Inst Electr Electron Eng       Date:  2022-05-03       Impact factor: 14.910

5.  Modeling and Validation of the Three-Dimensional Deflection of an MRI-Compatible Magnetically Actuated Steerable Catheter.

Authors:  Taoming Liu; Nate Lombard Poirot; Dominique Franson; Nicole Seiberlich; Mark A Griswold; M Cenk Cavusoglu
Journal:  IEEE Trans Biomed Eng       Date:  2015-12-22       Impact factor: 4.538

6.  Magnetic catheter manipulation in the interventional MR imaging environment.

Authors:  Mark W Wilson; Alastair B Martin; Prasheel Lillaney; Aaron D Losey; Erin J Yee; Anthony Bernhardt; Vincent Malba; Lee Evans; Ryan Sincic; Maythem Saeed; Ronald L Arenson; Steven W Hetts
Journal:  J Vasc Interv Radiol       Date:  2013-06       Impact factor: 3.464

7.  Magnetically assisted remote-controlled endovascular catheter for interventional MR imaging: in vitro navigation at 1.5 T versus X-ray fluoroscopy.

Authors:  Aaron D Losey; Prasheel Lillaney; Alastair J Martin; Daniel L Cooke; Mark W Wilson; Bradford R H Thorne; Ryan S Sincic; Ronald L Arenson; Maythem Saeed; Steven W Hetts
Journal:  Radiology       Date:  2014-02-12       Impact factor: 11.105

8.  Endovascular catheter for magnetic navigation under MR imaging guidance: evaluation of safety in vivo at 1.5T.

Authors:  S W Hetts; M Saeed; A J Martin; L Evans; A F Bernhardt; V Malba; F Settecase; L Do; E J Yee; A Losey; R Sincic; P Lillaney; S Roy; R L Arenson; M W Wilson
Journal:  AJNR Am J Neuroradiol       Date:  2013-07-11       Impact factor: 3.825

Review 9.  Remote control catheter navigation: options for guidance under MRI.

Authors:  Leah Muller; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  J Cardiovasc Magn Reson       Date:  2012-06-01       Impact factor: 5.364

10.  Heat-Mitigated Design and Lorentz Force-Based Steering of an MRI-Driven Microcatheter toward Minimally Invasive Surgery.

Authors:  Martin Francis Phelan; Mehmet Efe Tiryaki; Jelena Lazovic; Hunter Gilbert; Metin Sitti
Journal:  Adv Sci (Weinh)       Date:  2022-02-03       Impact factor: 16.806

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.