A CCD-based optical CT scanner for high-resolution 3D imaging of radiation dose distributions: equipment specifications, optical simulations and preliminary results.

S J Doran1, K K Koerkamp, M A Bero, P Jenneson, E J Morton, W B Gilboy.   

Abstract

Methods based on magnetic resonance imaging for the measurement of three-dimensional distributions of radiation dose are highly developed. However, relatively little work has been done on optical computed tomography (OCT). This paper describes a new OCT scanner based on a broad beam light source and a two-dimensional charge-coupled device (CCD) detector. A number of key design features are discussed including the light source; the scanning tank, turntable and stepper motor control; the diffuser screen onto which images are projected and the detector. It is shown that the non-uniform pixel sensitivity of the low-cost CCD detector used and the granularity of the diffuser screen lead to a serious ring artefact in the reconstructed images. Methods are described for eliminating this. The problems arising from reflection and refraction at the walls of the gel container are explained. Optical ray-tracing simulations are presented for cylindrical containers with a variety of radii and verified experimentally. Small changes in the model parameters lead to large variations in the signal intensity observed in the projection data. The effect of imperfect containers on data quality is discussed and a method based on a 'correction scan' is shown to be successful in correcting many of the related image artefacts. The results of two tomography experiments are presented. In the first experiment, a radiochromic Fricke gel sample was exposed four times in different positions to a 100 kVp x-ray beam perpendicular to the plane of imaging. Images of absorbed dose with slice thickness of 140 microm were acquired. with 'true' in-plane resolution of 560 x 560 microm2 at the edge of the 72 mm field of view and correspondingly higher resolution at the centre. The nominal doses measured correlated well with the known exposure times. The second experiment demonstrated the well known phenomenon of diffusion in the dosemeter gels and yielded a value of (0.12 +/- 0.02) mm2 s(-1) for the diffusion coefficient of the xylenol orange/iron complex. Finally, the overall implications of the above findings for dosimetry using OCT are discussed.

Entities:  

Mesh:

Year:  2001        PMID: 11768500     DOI: 10.1088/0031-9155/46/12/309

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  29 in total

1.  Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion.

Authors:  Mark Oldham; Leonard Kim
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

2.  Optical-CT scanning of polymer gels.

Authors:  M Oldham
Journal:  J Phys Conf Ser       Date:  2004

3.  Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation.

Authors:  H S Sakhalkar; M Oldham
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

4.  Quality assurance in 3D dosimetry by optical-CT.

Authors:  Pengyi Guo; John Adamovics; Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006-12-01

5.  A dual-purpose CCD based micro-optical-CT scanning system.

Authors:  M Oldham; H Sakhalkar; P Guo
Journal:  J Phys Conf Ser       Date:  2006-12-01

6.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

Review 7.  Optical clearing of unsectioned specimens for three-dimensional imaging via optical transmission and emission tomography.

Authors:  Mark Oldham; Harshad Sakhalkar; Tim Oliver; G Allan Johnson; Mark Dewhirst
Journal:  J Biomed Opt       Date:  2008 Mar-Apr       Impact factor: 3.170

8.  How effective can optical-CT 3D dosimetry be without refractive fluid matching?

Authors:  L Rankine
Journal:  J Phys Conf Ser       Date:  2013

9.  Sensitivity calibration procedures in optical-CT scanning of BANG 3 polymer gel dosimeters.

Authors:  Y Xu; Cheng-Shie Wuu; Marek J Maryanski
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

10.  A Convex Reconstruction Model for X-ray Tomographic Imaging with Uncertain Flat-fields.

Authors:  Hari Om Aggrawal; Martin S Andersen; Sean Rose; Emil Y Sidky
Journal:  IEEE Trans Comput Imaging       Date:  2017-07-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.