Literature DB >> 22149857

Commissioning a small-field biological irradiator using point, 2D, and 3D dosimetry techniques.

Joseph Newton1, Mark Oldham, Andrew Thomas, Yifan Li, John Adamovics, David G Kirsch, Shiva Das.   

Abstract

PURPOSE: To commission a small-field biological irradiator, the XRad225Cx from Precision x-Ray, Inc., for research use. The system produces a 225 kVp x-ray beam and is equipped with collimating cones that produce both square and circular radiation fields ranging in size from 1 to 40 mm. This work incorporates point, 2D, and 3D measurements to determine output factors (OF), percent-depth-dose (PDD) and dose profiles at multiple depths.
METHODS: Three independent dosimetry systems were used: ion-chambers (a farmer chamber and a micro-ionisation chamber), 2D EBT2 radiochromic film, and a novel 3D dosimetry system (DLOS∕PRESAGE®). Reference point dose rates and output factors were determined from in-air ionization chamber measurements for fields down to ∼13 mm using the formalism of TG61. PDD, profiles, and output factors at three separate depths (0, 0.5, and 2 cm), were determined for all field sizes from EBT2 film measurements in solid water. Several film PDD curves required a scaling correction, reflecting the challenge of accurate film alignment in very small fields. PDDs, profiles, and output factors were also determined with the 3D DLOS∕PRESAGE® system which generated isotropic 0.2 mm data, in scan times of 20 min.
RESULTS: Surface output factors determined by ion-chamber were observed to gradually drop by ∼9% when the field size was reduced from 40 to 13 mm. More dramatic drops were observed for the smallest fields as determined by EBT∼18% and ∼42% for the 2.5 mm and 1 mm fields, respectively. PRESAGE® and film output factors agreed well for fields <20 mm (where 3D data were available) with mean deviation of 2.2% (range 1%-4%). PDD values at 2 cm depth varied from ∼72% for the 40 mm field, down to ∼55% for the 1 mm field. EBT and PRESAGE® PDDs agreed within ∼3% in the typical therapy region (1-4 cm). At deeper depths the EBT curves were slightly steeper (2.5% at 5 cm). These results indicate good overall consistency between ion-chamber, EBT2 and PRESAGE® measured OFs, PDDs, and profiles.
CONCLUSIONS: The combination of independent 2D and 3D measurements was found to be valuable to ensure accurate and comprehensive commissioning. Film measurements were time consuming and challenging due to the difficulty of film alignment in small fields. PRESAGE® 3D measurements were comprehensive and efficient, because alignment errors are negligible, and all parameters for multiple fields could be obtained from a single dosimeter and scan. However, achieving accurate superficial data (within 4 mm) is not yet feasible due to optical surface artifacts.

Entities:  

Mesh:

Year:  2011        PMID: 22149857      PMCID: PMC3247930          DOI: 10.1118/1.3663675

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

1.  High resolution gel-dosimetry by optical-CT and MR scanning.

Authors:  M Oldham; J H Siewerdsen; A Shetty; D A Jaffray
Journal:  Med Phys       Date:  2001-07       Impact factor: 4.071

2.  Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation.

Authors:  H S Sakhalkar; M Oldham
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  3D dosimetry by optical-CT scanning.

Authors:  Mark Oldham
Journal:  J Phys Conf Ser       Date:  2006

4.  GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

Authors:  Sigrun Saur; Jomar Frengen
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

5.  Investigation of radiological properties and water equivalency of PRESAGE dosimeters.

Authors:  Tina Gorjiara; Robin Hill; Zdenka Kuncic; John Adamovics; Stephen Bosi; Jung-Ha Kim; Clive Baldock
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

6.  Radiation dose distributions in three dimensions from tomographic optical density scanning of polymer gels: I. Development of an optical scanner.

Authors:  J C Gore; M Ranade; M J Maryañski; R J Schulz
Journal:  Phys Med Biol       Date:  1996-12       Impact factor: 3.609

7.  A method to correct for stray light in telecentric optical-CT imaging of radiochromic dosimeters.

Authors:  Andrew Thomas; Joseph Newton; Mark Oldham
Journal:  Phys Med Biol       Date:  2011-06-30       Impact factor: 3.609

8.  Small SRS photon field profile dosimetry performed using a PinPoint air ion chamber, a diamond detector, a novel silicon-diode array (DOSI), and polymer gel dosimetry. Analysis and intercomparison.

Authors:  E Pappas; T G Maris; F Zacharopoulou; A Papadakis; S Manolopoulos; S Green; C Wojnecki
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

9.  Small field dose delivery evaluations using cone beam optical computed tomography-based polymer gel dosimetry.

Authors:  Timothy Olding; Oliver Holmes; Paul Dejean; Kim B McAuley; Ken Nkongchu; Giles Santyr; L John Schreiner
Journal:  J Med Phys       Date:  2011-01

10.  p53 controls radiation-induced gastrointestinal syndrome in mice independent of apoptosis.

Authors:  David G Kirsch; Philip M Santiago; Emmanuelle di Tomaso; Julie M Sullivan; Wu-Shiun Hou; Talya Dayton; Laura B Jeffords; Pooja Sodha; Kim L Mercer; Rhianna Cohen; Osamu Takeuchi; Stanley J Korsmeyer; Roderick T Bronson; Carla F Kim; Kevin M Haigis; Rakesh K Jain; Tyler Jacks
Journal:  Science       Date:  2009-12-17       Impact factor: 47.728

View more
  32 in total

1.  PRESAGE 3D dosimetry accurately measures Gamma Knife output factors.

Authors:  Slade J Klawikowski; James N Yang; John Adamovics; Geoffrey S Ibbott
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  On the feasibility of polyurethane based 3D dosimeters with optical CT for dosimetric verification of low energy photon brachytherapy seeds.

Authors:  Justus Adamson; Yun Yang; Titania Juang; Kelsey Chisholm; Leith Rankine; John Adamovics; Fang Fang Yin; Mark Oldham
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

Review 3.  A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy.

Authors:  S O'Keeffe; D McCarthy; P Woulfe; M W D Grattan; A R Hounsell; D Sporea; L Mihai; I Vata; G Leen; E Lewis
Journal:  Br J Radiol       Date:  2015-03-11       Impact factor: 3.039

4.  Commissioning a CT-compatible LDR tandem and ovoid applicator using Monte Carlo calculation and 3D dosimetry.

Authors:  Justus Adamson; Joseph Newton; Yun Yang; Beverly Steffey; Jing Cai; John Adamovics; Mark Oldham; Junzo Chino; Oana Craciunescu
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

5.  p53 functions in endothelial cells to prevent radiation-induced myocardial injury in mice.

Authors:  Chang-Lung Lee; Everett J Moding; Kyle C Cuneo; Yifan Li; Julie M Sullivan; Lan Mao; Iman Washington; Laura B Jeffords; Rafaela C Rodrigues; Yan Ma; Shiva Das; Christopher D Kontos; Yongbaek Kim; Howard A Rockman; David G Kirsch
Journal:  Sci Signal       Date:  2012-07-24       Impact factor: 8.192

6.  Fast dynamic electron paramagnetic resonance (EPR) oxygen imaging using low-rank tensors.

Authors:  Anthony G Christodoulou; Gage Redler; Bryan Clifford; Zhi-Pei Liang; Howard J Halpern; Boris Epel
Journal:  J Magn Reson       Date:  2016-07-15       Impact factor: 2.229

7.  Adaptation, Commissioning, and Evaluation of a 3D Treatment Planning System for High-Resolution Small-Animal Irradiation.

Authors:  Jeho Jeong; Qing Chen; Robert Febo; Jie Yang; Hai Pham; Jian-Ping Xiong; Pat B Zanzonico; Joseph O Deasy; John L Humm; Gig S Mageras
Journal:  Technol Cancer Res Treat       Date:  2015-05-06

8.  A precision 3D conformal treatment technique in rats: Application to whole-brain radiotherapy with hippocampal avoidance.

Authors:  Suk W Yoon; Christina K Cramer; Devin A Miles; Michael H Reinsvold; Kyeung M Joo; David G Kirsch; Mark Oldham
Journal:  Med Phys       Date:  2017-09-30       Impact factor: 4.071

9.  Development and dosimetry of a small animal lung irradiation platform.

Authors:  Ross McGurk; Caroline Hadley; Isabel L Jackson; Zeljko Vujaskovic
Journal:  Health Phys       Date:  2012-10       Impact factor: 1.316

10.  Automated measurement of microcirculatory blood flow velocity in pulmonary metastases of rats.

Authors:  Gert Blueschke; Gabi Hanna; Andrew N Fontanella; Gregory M Palmer; Alina Boico; Hooney Min; Mark W Dewhirst; David C Irwin; Yulin Zhao; Thies Schroeder
Journal:  J Vis Exp       Date:  2014-11-30       Impact factor: 1.355

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.