A quality assurance protocol for diffusion tensor imaging using the head phantom from American College of Radiology.

Zhiyue J Wang1, Youngseob Seo, Jonathan M Chia, Nancy K Rollins.   

Abstract

PURPOSE: To propose a quality assurance procedure for routine clinical diffusion tensor imaging (DTI) using the widely available American College of Radiology (ACR) head phantom.
METHODS: Analysis was performed on the data acquired at 1.5 and 3.0 T on whole body clinical MRI scanners using the ACR phantom and included the following: (1) the signal-to-noise ratio (SNR) at the center and periphery of the phantom, (2) image distortion by EPI readout relative to spin echo imaging, (3) distortion of high-b images relative to the b= 0 image caused by diffusion encoding, and (4) determination of fractional anisotropy (FA) and mean diffusivity (MD) measured with region-of-interest (ROI) and pixel-based approaches. Reproducibility of the measurements was assessed by five repetitions of data acquisition on each scanner.
RESULTS: The SNR at the phantom center was approximately half of that near the periphery at both 1.5 and 3 T. The image distortion by the EPI readout was up to 7 mm at 1.5 T and 10 mm at 3 T. The typical distortion caused by eddy currents from diffusion encoding was on the order of 0.5 mm. The difference between ROI-based and pixel-based MD quantification was 1.4% at 1.5 T and 0.3% at 3 T. The ROI-based MD values were in close agreement (within 2%) with the reference values. The ROI-based FA values were approximately a factor of 10 smaller than pixel-based values and less than 0.01. The measurement reproducibility was sufficient for quality assurance (QA) purposes.
CONCLUSIONS: This QA approach is simple to perform and evaluates key aspects of the scanner performance for DTI data acquisition using a widely available phantom.

Entities:  

Mesh:

Year:  2011        PMID: 21859042      PMCID: PMC3145225          DOI: 10.1118/1.3595111

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

Review 1.  Report on a multicenter fMRI quality assurance protocol.

Authors:  Lee Friedman; Gary H Glover
Journal:  J Magn Reson Imaging       Date:  2006-06       Impact factor: 4.813

Review 2.  Artifacts and pitfalls in diffusion MRI.

Authors:  Denis Le Bihan; Cyril Poupon; Alexis Amadon; Franck Lethimonnier
Journal:  J Magn Reson Imaging       Date:  2006-09       Impact factor: 4.813

3.  Use of capillaries in the construction of an MRI phantom for the assessment of diffusion tensor imaging: demonstration of performance.

Authors:  Nathan Yanasak; Jerry Allison
Journal:  Magn Reson Imaging       Date:  2006-10-05       Impact factor: 2.546

4.  EPI distortion correction from a simultaneously acquired distortion map using TRAIL.

Authors:  Andrew N Priest; Enrico De Vita; David L Thomas; Roger J Ordidge
Journal:  J Magn Reson Imaging       Date:  2006-04       Impact factor: 4.813

Review 5.  Diffusion tensor imaging of the brain.

Authors:  Andrew L Alexander; Jee Eun Lee; Mariana Lazar; Aaron S Field
Journal:  Neurotherapeutics       Date:  2007-07       Impact factor: 7.620

6.  Phantom-based evaluation of geometric distortions in functional magnetic resonance and diffusion tensor imaging.

Authors:  Seppo Mattila; Ville Renvall; Jaana Hiltunen; Douglas Kirven; Raimo Sepponen; Riitta Hari; Antti Tarkiainen
Journal:  Magn Reson Med       Date:  2007-04       Impact factor: 4.668

7.  Effects of signal-to-noise ratio on the accuracy and reproducibility of diffusion tensor imaging-derived fractional anisotropy, mean diffusivity, and principal eigenvector measurements at 1.5 T.

Authors:  Jonathan A D Farrell; Bennett A Landman; Craig K Jones; Seth A Smith; Jerry L Prince; Peter C M van Zijl; Susumu Mori
Journal:  J Magn Reson Imaging       Date:  2007-09       Impact factor: 4.813

8.  Comparison of EPI distortion correction methods in diffusion tensor MRI using a novel framework.

Authors:  M Wu; L C Chang; L Walker; H Lemaitre; A S Barnett; S Marenco; C Pierpaoli
Journal:  Med Image Comput Comput Assist Interv       Date:  2008

9.  Influence of multichannel combination, parallel imaging and other reconstruction techniques on MRI noise characteristics.

Authors:  Olaf Dietrich; José G Raya; Scott B Reeder; Michael Ingrisch; Maximilian F Reiser; Stefan O Schoenberg
Journal:  Magn Reson Imaging       Date:  2008-04-28       Impact factor: 2.546

Review 10.  Clinical applications of diffusion tensor imaging and tractography in children.

Authors:  Nancy K Rollins
Journal:  Pediatr Radiol       Date:  2007-06-28
View more
  13 in total

1.  Integration of routine QA data into mega-analysis may improve quality and sensitivity of multisite diffusion tensor imaging studies.

Authors:  Peter Kochunov; Erin W Dickie; Joseph D Viviano; Jessica Turner; Peter B Kingsley; Neda Jahanshad; Paul M Thompson; Meghann C Ryan; Els Fieremans; Dmitry Novikov; Jelle Veraart; Elliot L Hong; Anil K Malhotra; Robert W Buchanan; Sofia Chavez; Aristotle N Voineskos
Journal:  Hum Brain Mapp       Date:  2017-11-27       Impact factor: 5.038

2.  Signal-to-noise assessment for diffusion tensor imaging with single data set and validation using a difference image method with data from a multicenter study.

Authors:  Zhiyue J Wang; Jonathan M Chia; Shaheen Ahmed; Nancy K Rollins
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

3.  A fiber coherence index for quality control of B-table orientation in diffusion MRI scans.

Authors:  Kurt G Schilling; Fang-Cheng Yeh; Vishwesh Nath; Colin Hansen; Owen Williams; Susan Resnick; Adam W Anderson; Bennett A Landman
Journal:  Magn Reson Imaging       Date:  2019-01-22       Impact factor: 2.546

4.  A novel DTI-QA tool: Automated metric extraction exploiting the sphericity of an agar filled phantom.

Authors:  Sofia Chavez; Joseph Viviano; Mojdeh Zamyadi; Peter B Kingsley; Peter Kochunov; Stephen Strother; Aristotle Voineskos
Journal:  Magn Reson Imaging       Date:  2017-10-17       Impact factor: 2.546

5.  Signal-to-noise ratio assessment of muscle diffusion tensor imaging using single image set and validation by the difference image method.

Authors:  Zhiyue J Wang; Jin Yamamura; Sarah Keller
Journal:  Br J Radiol       Date:  2019-07-26       Impact factor: 3.039

6.  MR scanner systems should be adequately characterized in diffusion-MRI of the breast.

Authors:  Marco Giannelli; Roberto Sghedoni; Chiara Iacconi; Mauro Iori; Antonio Claudio Traino; Maria Guerrisi; Mario Mascalchi; Nicola Toschi; Stefano Diciotti
Journal:  PLoS One       Date:  2014-01-28       Impact factor: 3.240

7.  Early detection of neonatal hypoxic-ischemic white matter injury: an MR diffusion tensor imaging study.

Authors:  Youngseob Seo; Geun-Tae Kim; Jin Wook Choi
Journal:  Neuroreport       Date:  2017-09-06       Impact factor: 1.837

8.  MRI scanner-independent specific absorption rate measurements using diffusion coefficients.

Authors:  Youngseob Seo; Zhiyue J Wang
Journal:  J Appl Clin Med Phys       Date:  2017-05-04       Impact factor: 2.102

9.  MRI phantoms - are there alternatives to agar?

Authors:  Alexandra Hellerbach; Verena Schuster; Andreas Jansen; Jens Sommer
Journal:  PLoS One       Date:  2013-08-05       Impact factor: 3.240

10.  Changes in lumbosacral spinal nerve roots on diffusion tensor imaging in spinal stenosis.

Authors:  Zhong-Jun Hou; Yong Huang; Zi-Wen Fan; Xin-Chun Li; Bing-Yi Cao
Journal:  Neural Regen Res       Date:  2015-11       Impact factor: 5.135

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.