| Literature DB >> 28470956 |
Youngseob Seo1,2, Zhiyue J Wang2.
Abstract
OBJECTIVE: The purpose of this study was to measure specific absorption rate (SAR) during MRI scanning using a human torso phantom through quantification of diffusion coefficients independently of those reported by the scanner software for five 1.5 and 3 T clinical MRI systems from different vendors.Entities:
Keywords: MR safety; diffusion coefficient; diffusion tensor imaging; human torso phantom; specific absorption rate
Mesh:
Year: 2017 PMID: 28470956 PMCID: PMC5875836 DOI: 10.1002/acm2.12095
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Image acquisition parameters at 1.5 and 3.0 T
| Image sequence | |||
|---|---|---|---|
| T1w TSE | T1 TIRM | T2w TSE | |
| TR/TE [ms] | 800/10 | 1150/9.3 | 4710/110 |
| TI [ms] | n/a | 220 | n/a |
| Field of view [mm2] | 400 × 400 | 400 × 400 | 400 × 400 |
| No. of slices | 10 | 8 | 10 |
| Slice thickness [mm] | 6 | 6 | 6 |
| Acquisition matrix | 400 × 400 | 256 × 256 | 200 × 154 (reconstructed to 400 × 400) |
| Voxel size [mm3] | 1 × 1 × 6 | 1.56 × 1.56 × 6 | 1 × 1 × 6 |
| Slice orientation | Transverse | Transverse | Transverse |
| Phase‐encoding direction | AP | AP | AP |
| NSA | 2 | 1 | 4 |
| Total scan time (GE 1.5 T/ Philips 1.5 T) | 5 min 12 s/4 min 56 s | 4 min 18 s | 4 min 23 s |
| Parallel imaging method | No | GRAPPA for Siemens 3 T | No |
| Bandwidth [Hz/pixel] (GE 1.5 T/ Philips 1.5 T) | 260/290 | 260 | 334 |
TSE, Turbo spin echo; TIRM, Turbo inversion recovery magnitude.
Figure 1Phantom morphology mimics the shape of the human torso.
Figure 2Location of four optic fiber temperature sensors in the phantom periphery, used to measure the initial temperatures of the phantom and the time taken to reach equilibrium with the environment.
Figure 3The apparent diffusion coefficient within an ROI (red) which was manually placed on the periphery of the phantom was quantified using a set of diffusion‐weighted images.
DTI acquisition parameters for each scanner
| GE 1.5 T | Philips 1.5 T | Siemens 3 T | Philips 3 T | |
|---|---|---|---|---|
| b‐values [s mm‐2] | 0, 1000 | 0, 1000 | 0, 1000 | 0, 1000 |
| No. of gradient directions | 30 | 30 | 30 | 30 |
| TE [msec] | 80 | 82 | 88 | 74 |
| TR [msec] | 9218 | 8000 | 7100 | 6844 |
| FoV [mm2] | 400 × 400 | 380 × 380 | 400 × 400 | 400 × 400 |
| Acquisition matrix | 128 × 128 | 128 × 128 | 128 × 128 | 128 × 128 |
| Slice thickness [mm] | 4 | 4 | 4 | 4 |
| No. of slices | 16 | 14 | 14 | 14 |
| Parallel imaging method | ASSET | SENSE | GRAPPA | SENSE |
| Acquisition time | 5 min 12 s | 4 min 56 s | 4 min 18 s | 4 min 23 s |
Figure 4Temperature change derived from the difference between mean diffusivity (MD) maps is shown for one Philips 3T scanner.
Phantom temperatures (mean ± SD) measured via optic fiber sensors and times to equilibrate with temperature inside a magnet bore at 3 T
| Temperature inside magnet bore = 22.8 ± 0.6°C | |||
|---|---|---|---|
| Sensor No. | Initial temperature | Final temperature | Time to equilibrate with the environment |
| OFS#1 | 28.4 ± 0.5°C | 22.7 ± 0.6°C | 4 hrs 08 min |
| OFS#2 | 28.3 ± 0.6°C | 22.8 ± 0.5°C | 4 hrs 02 min |
| OFS#3 | 28.3 ± 0.6°C | 22.8 ± 0.6°C | 4 hrs 07 min |
| OFS#4 | 28.2 ± 0.5°C | 22.7 ± 0.6°C | 4 hrs 11 min |
Temperatures obtained by diffusion coefficients and measured via optic fiber temperature sensors before and after high SAR image sequence on one of two same model Philips 3 T (T2w TSE and MRI system‐reported SAR value = 1.5 W kg‐1)
| Before high SAR image sequence | After high SAR image sequence | Temperature difference between before and after high SAR image sequence | ||
|---|---|---|---|---|
| Position #1 | Mean diffusion coefficient [× 10−3 mm2 s−1] | 2.1462 | 2.1553 | |
| Temperature obtained by mean diffusion coefficient [K] | 295.41 | 295.57 | 0.16 | |
| Temperature measured via temperature sensor [°C] | 24.18 | 24.35 | 0.17 | |
| Position #2 | Mean diffusion coefficient [× 10−3 mm2 s−1] | 2.1473 | 2.1542 | |
| Temperature obtained by mean diffusion coefficient [K] | 295.43 | 295.56 | 0.13 | |
| Temperature measured via temperature sensor [°C] | 24.01 | 24.15 | 0.14 | |
| Position #3 | Mean diffusion coefficient [× 10−3 mm2 s−1] | 2.1473 | 2.1533 | |
| Temperature obtained by mean diffusion coefficient [K] | 295.43 | 295.54 | 0.11 | |
| Temperature measured via temperature sensor [°C] | 24.16 | 24.28 | 0.12 | |
| Position #4 | Mean diffusion coefficient [× 10−3 mm2 s−1] | 2.1484 | 2.1583 | |
| Temperature obtained by mean diffusion coefficient [K] | 295.45 | 295.63 | 0.18 | |
| Temperature measured via temperature sensor [°C] | 24.21 | 24.38 | 0.17 |
SAR measurement results using MR DTI from four MRI systems
| Image sequence | GE 1.5 T | Philips 1.5 T | Siemens 3 T | Philips 3 T | Philips 3 T |
|---|---|---|---|---|---|
| T1w TSE | T1w TSE | T1 TIRM | T2w TSE | T2w TSE | |
| Temperature difference between before and after the high SAR image sequence [°C] (mean ± SD) | 0.08 ± 0.01 | 0.07 ± 0.01 | 0.11 ± 0.01 | 0.18 ± 0.02 | 0.18 ± 0.02 |
| Measured SAR value induced by a DTI scan [W kg‐1] (mean±SD) | 0.11 ± 0.01 | 0.21 ± 0.02 | 0.48 ± 0.04 | 0.53 ± 0.05 | 0.51 ± 0.05 |
| High SAR sequence scan time [sec] | 168 | 169 | 186 | 320 | 320 |
| MRI system‐reported SAR value [W kg‐1] | 1.58 | 1.48 | 2.5 | 1.5 | 1.5 |
| SAR value measured for the high SAR image sequence [W kg‐1] (mean ± SD) | 1.38 ± 0.09 | 1.39 ± 0.12 | 1.96 ± 0.18 | 1.94 ± 0.18 | 1.96 ± 0.19 |
| Percentage difference between measured and reported SAR values [%] | −13.5 | −6.3 | −24.2 | 25.6 | 26.6 |
TSE, Turbo spin echo; TIRM, Turbo inversion recovery magnitude.
The sign ‘‐’ means that measured SAR value was lower than the reported one.