EPI distortion correction from a simultaneously acquired distortion map using TRAIL.

Andrew N Priest1, Enrico De Vita, David L Thomas, Roger J Ordidge.   

Abstract

PURPOSE: To develop a method for shot-by-shot distortion correction of single-shot echo-planar imaging (EPI) that is capable of correcting each image individually using a distortion measurement performed during acquisition of the image itself.
MATERIALS AND METHODS: The recently-introduced method known as two reduced acquisitions interleaved (TRAIL) was extended to measure the distribution of the main magnetic field B0 with each shot. This corresponded to a map of distortion, and allowed distortion to be corrected in the acquired images.
RESULTS: Distortion-corrected images were demonstrated in the human brain. The distortion field could be directly visualized using the "stripe" distribution imposed by the TRAIL pulse sequence. This confirmed the success of the correction. Over a time-course measurement of 10 images, variance was reduced by using shot-by-shot distortion correction compared to correction with a constant field map.
CONCLUSION: Shot-by-shot distortion correction may be performed for EPI images acquired using an extension of the TRAIL technique, ensuring that the correction reflects the actual distortion pattern and not merely a previously measured, but possibly no longer valid, distortion field. This avoids errors due to changes in the distortion field or misregistration of a previously measured distortion map resulting from subject motion. 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2006        PMID: 16514595     DOI: 10.1002/jmri.20508

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  Distortion correction for diffusion-weighted MRI tractography and fMRI in the temporal lobes.

Authors:  Karl V Embleton; Hamied A Haroon; David M Morris; Matthew A Lambon Ralph; Geoff J M Parker
Journal:  Hum Brain Mapp       Date:  2010-10       Impact factor: 5.038

2.  A quality assurance protocol for diffusion tensor imaging using the head phantom from American College of Radiology.

Authors:  Zhiyue J Wang; Youngseob Seo; Jonathan M Chia; Nancy K Rollins
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

3.  Regularized field map estimation in MRI.

Authors:  Amanda K Funai; Jeffrey A Fessler; Desmond T B Yeo; Valur T Olafsson; Douglas C Noll
Journal:  IEEE Trans Med Imaging       Date:  2008-10       Impact factor: 10.048

4.  Efficient correction of inhomogeneous static magnetic field-induced distortion in Echo Planar Imaging.

Authors:  Dominic Holland; Joshua M Kuperman; Anders M Dale
Journal:  Neuroimage       Date:  2009-11-26       Impact factor: 6.556

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.