Biomechanical analysis of the effects of implant diameter and bone quality in short implants placed in the atrophic posterior maxilla.

Shih-Hao Chang1, Chun-Li Lin, Shue-Sang Hsue, Yang-Sung Lin, Shiang-Rung Huang.   

Abstract

Short dental implant (SDI) placement has been proposed as an alternative to reduce the surgical risks related to the advanced grafting procedures. The aim of this study was to simulate the biomechanical behaviors and influences of SDI diameters under various conditions of bone quality by using a validated finite element (FE) model for simulation. The CT image and CAD system were combined to construct the FE models with 6 mm length SDIs for 6, 7 and 8 mm diameters under three types of bone qualities, from normal to osteoporotic. The simulated results showed that implant diameter did not influence the von Mises strains of bone under the vertical load. The bone strains increased about 58.58% in the bone of least density under lateral load. Lateral loads induced high bone strain and implant stress than vertical loads. The bone strains of 7 mm- and 8 mm-diameter short implants were not different, and both were about 52% and 66% compared to those of 6 mm-wide short implant under lateral loads. The von Mises stress of the SDIs and their compartments were all less than the yield stress of the material under vertical and lateral loads. SDIs with diameter of 7 mm or above may have better mechanical transmission in the same length at feasible condition. Attaining a proper occlusal scheme design or selective occlusal adjustments to reduce the lateral occlusal force upon the SDIs is recommended.
Copyright © 2011 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21807548     DOI: 10.1016/j.medengphy.2011.07.005

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  10 in total

1.  Posterior partially edentulous jaws, planning a rehabilitation with dental implants.

Authors:  Douglas R Monteiro; Emily V F Silva; Eduardo P Pellizzer; Osvaldo Magro Filho; Marcelo C Goiato
Journal:  World J Clin Cases       Date:  2015-01-16       Impact factor: 1.337

2.  Influence of implant number, length, and tilting degree on stress distribution in atrophic maxilla: a finite element study.

Authors:  Zeynep Gümrükçü; Yavuz Tolga Korkmaz
Journal:  Med Biol Eng Comput       Date:  2017-11-09       Impact factor: 2.602

3.  Success of dental implants in patients with large bone defect and analysis of risk factors for implant failure: a non-randomized retrospective cohort study.

Authors:  Takumi Hasegawa; Aki Sasaki; Izumi Saito; Satomi Arimoto; Nanae Yatagai; Yujiro Hiraoka; Daisuke Takeda; Yasumasa Kakei; Masaya Akashi
Journal:  Clin Oral Investig       Date:  2021-11-16       Impact factor: 3.573

4.  Implant surface alters compartmental-specific contributions to fixation strength in rats.

Authors:  Frank C Ko; Matthew J Meagher; Maleeha Mashiatulla; Ryan D Ross; Amarjit S Virdi; Dale R Sumner
Journal:  J Orthop Res       Date:  2019-12-25       Impact factor: 3.494

5.  Internal oblique line implants in severe mandibular atrophies.

Authors:  Argimiro Hernández-Suarez; Luis-Guillermo Oliveros-López; María-Ángeles Serrera-Figallo; Celia Vázquez-Pachón; Daniel Torres-Lagares; José-Luis Gutiérrez-Pérez
Journal:  J Clin Exp Dent       Date:  2020-12-01

6.  Short implants in partially edentuolous maxillae and mandibles: a 10 to 20 years retrospective evaluation.

Authors:  Diego Lops; Eriberto Bressan; Gianluca Pisoni; Niccolò Cea; Boris Corazza; Eugenio Romeo
Journal:  Int J Dent       Date:  2012-07-09

7.  Photoelastic analysis of fixed partial prosthesis crown height and implant length on distribution of stress in two dental implant systems.

Authors:  Evandro Portela Figueirêdo; Eder Alberto Sigua-Rodriguez; Marcele Jardim Pimentel; Ana Regina Oliveira Moreira; Mauro Antônio de Arruda Nóbilo; José Ricardo de Albergaria-Barbosa
Journal:  Int J Dent       Date:  2014-10-08

8.  Influence of Implant Design (Cylindrical and Conical) in the Load Transfer Surrounding Long (13mm) and Short (7mm) Length Implants: A Photoelastic Analysis.

Authors:  Sergio Alexandre Gehrke; Victor Lourenço Frugis; Jamil Awad Shibli; Maria Piedad Ramirez Fernandez; José Eduardo Maté Sánchez de Val; José Luis Calvo Girardo; Silvio Taschieri; Stefano Corbella
Journal:  Open Dent J       Date:  2016-09-30

9.  GPU accelerated voxel-driven forward projection for iterative reconstruction of cone-beam CT.

Authors:  Yi Du; Gongyi Yu; Xincheng Xiang; Xiangang Wang
Journal:  Biomed Eng Online       Date:  2017-01-05       Impact factor: 2.819

Review 10.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.