Masoud Varshowsaz1, Sepideh Goorang1, Sara Ehsani1, Zeynab Azizi2, Sepideh Rahimian3.
Abstract
OBJECTIVES: Bone quality and quantity assessment is one of the most important steps in implant treatment planning. Different methods such as computed tomography (CT) and recently suggested cone beam computed tomography (CBCT) with lower radiation dose and less time and cost are used for bone density assessment. This in vitro study aimed to compare the tissue density values in Hounsfield units (HUs) in CBCT and CT scans of different tissue phantoms with two different thicknesses, two different image acquisition settings and in three locations in the phantoms.Entities:
Keywords: Bone Density; Cone-Beam Computed Tomography; Tomography, X-Ray Computed
Year: 2016 PMID: 27928239 PMCID: PMC5139928
Source DB: PubMed Journal: J Dent (Tehran) ISSN: 1735-2150
Fig. 1:Different tissue phantoms placed in a cylindrical container as the water phantom.
Exposure parameters for different units
| 100 kVp, 140 mAs | 100 kVp, 100 mAs | |
| 85 kVp, 15 mAs | 85 kVp, 10 mAs | |
| 110 kVp, 4 mAs | 110 kVp, 1 mAs | |
| 84 kVp, 12 mAs | 84 kVp, 8 mAs |
Fig. 2:Evaluation of HUs in hard tissue phantom with Merge e-Film software in three times of scanning in full thickness phantom.
Comparison of voxel values in CBCT systems with Hounsfield units in CT in full thickness phantoms
| Scanora 3D | NS | Scanora 3D | S | |
| NewTom VG | S except for point P | NewTom VG | S except for point M | |
| Pro-Max 3D | S | Pro-Max 3D | S | |
| Scanora 3D | NS | Scanora 3D | S | |
| NewTom VG | NS | NewTom VG | S | |
| Pro-Max 3D | NS | Pro-Max 3D | S | |
| Scanora 3D | NS | Scanora 3D | NS | |
| NewTom VG | NS except for point M | NewTom VG | S | |
| Pro-Max 3D | NS | Pro-Max 3D | NS except for point P | |
| Scanora 3D | S except for point M | Scanora 3D | NS | |
| NewTom VG | S | NewTom VG | NS except for point P | |
| Pro-Max 3D | S | Pro-Max 3D | S | |
NS: Non-significant S: Significant
Comparison of voxel values in CBCT systems with Hounsfield units in CT in half thickness phantoms
| Scanora 3D | S | Scanora 3D | S | |
| NewTom VG | S | NewTom VG | S | |
| Pro-Max 3D | S | Pro-Max 3D | S | |
| Scanora 3D | S except for point M | Scanora 3D | S except for point M | |
| NewTom VG | S | NewTom VG | S | |
| Pro-Max 3D | S | Pro-Max 3D | S | |
| Scanora 3D | NS | Scanora 3D | NS except for point M | |
| NewTom VG | NS except for point INT | NewTom VG | S | |
| Pro-Max 3D | NS | Pro-Max 3D | NS except for point M | |
| Scanora 3D | S | Scanora 3D | S | |
| NewTom VG | S | NewTom VG | S | |
| Pro-Max 3D | S | Pro-Max 3D | S | |
NS: Non-significant S: Significant