| Literature DB >> 21587192 |
Srijit Kamath1, William Song, Alexei Chvetsov, Shuichi Ozawa, Haibin Lu, Sanjiv Samant, Chihray Liu, Jonathan G Li, Jatinder R Palta.
Abstract
The purpose of this study is to evaluate and compare image quality characteristics for two commonly used and commercially available CBCT systems: the X-ray Volumetric Imager and the On-Board Imager. A commonly used CATPHAN image quality phantom was used to measure various image quality parameters, namely, pixel value stability and accuracy, noise, contrast to noise ratio (CNR), high-contrast resolution, low contrast resolution and image uniformity. For the XVI unit, we evaluated the image quality for four manufacturer-supplied protocols as a function of mAs. For the OBI unit, we did the same for the full-fan and half-fan scanning modes, which were respectively used with the full bow-tie and half bow-tie filters. For XVI, the mean pixel values of regions of interest were found to generally decrease with increasing mAs for all protocols, while they were relatively stable with mAs for OBI. Noise was slightly lower on XVI and was seen to decrease with increasing mAs, while CNR increased with mAs for both systems. For XVI and OBI, the high-contrast resolution was approximately limited by the pixel resolution of the reconstructed image. On OBI images, up to 6 and 5 discs of 1% and 0.5% contrast, respectively, were visible for a high mAs setting using the full-fan mode, while none of the discs were clearly visible on the XVI images for various mAs settings when the medium resolution reconstruction was used. In conclusion, image quality parameters for XVI and OBI have been quantified and compared for clinical protocols under various mAs settings. These results need to be viewed in the context of a recent study that reported the dose-mAs relationship for the two systems and found that OBI generally delivered higher imaging doses than XVI.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21587192 PMCID: PMC5718664 DOI: 10.1120/jacmp.v12i2.3435
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
List of the relevant parameters used in the scanning with the XVI unit. For the acquisition gantry angle, cw means clockwise rotation and ccw means counter‐clockwise rotation.
|
| Head & Neck | Prostate | Pelvis |
|
|---|---|---|---|---|
| kV Collimator | S20 | M10 | M20 | L20 |
| kVp | 100 | 120 | 120 | 120 |
| (mA,ms/frame) | (10,10), (10,20), (25,10), (20,20) | (25,20), (25,40), (40,32), (50,40) | (25,20), (25,40), (40,32), (40,40) | (25,20), (25,40), (40,32), (50,40) |
| # frames | 361 | 650 | 650 | 650 |
| Total mAs | 36.1, 72.2, 90.25, 144.4 | 325, 650, 832, 1300 | 325, 650, 832, 1040 | 325, 650, 832, 1300 |
| Acquisition Gantry Angle | 260° to 100° | 183° to 179° | 183° to 179° | 183° to 179° |
| Acquisition Time |
|
|
|
|
| Voxel Size |
|
|
|
|
List of the relevant parameters used in the scanning with the OBI. For the acquisition gantry angle, cw means clockwise rotation and ccw means counter‐clockwise rotation.
|
| Full‐fan |
|
|---|---|---|
| Filter | Full Bow‐tie | Half Bow‐tie |
| kVp | 125 | 125 |
| (mA, ms/frame) | (10,10), (25,20), (50,20), (50,30), (80,25), (80,32) | (10,10), (25,20), (50,20), (50,30), (80,25), (80,32) |
| # frames | 630 | 630 |
| Total mAs | 63, 315, 630, 945, 1260, 1612.8 | 63, 315, 630, 945, 1260, 1612.8 |
| Acquisition Gantry Angle | 175.5° to 182.5° | 175.5° to 182.5° |
| Acquisition Time |
|
|
| Voxel Size |
|
|
Figure 1The two major commercial CBCT systems evaluated: (a) the OBI, (b) the XVI, with their corresponding CATPHAN phantom (small phantom) setup shown, and (c) the CATPHAN phantom with the annulus surrounding it (large phantom).
Figure 2Three of the modules of the CATPHAN phantom (model: CTP 504) and the image quality parameters we have measured using them: (a) the CTP 404 module for pixel value stability and contrast‐to‐noise ratio, (b) the CTP 528 module for high contrast resolution test, and (c) the CTP 515 module for low contrast resolution test.
Materials present in CTP 404 and their nominal CT numbers.
|
|
|
|---|---|
| Air |
|
| PMP |
|
| LDPE |
|
| Water | 0 |
| Polystyrene |
|
| Acrylic | 120 |
| Derlin | 340 |
| Teflon | 990 |
Figure 3Mean pixel values in ROIs of various inserts of the CTP 404 module as the mAs is varied for the OBI scanning protocols: (a) full‐fan (small phantom (SP)), (b) half‐fan (small phantom (SP)) and (c) half‐fan (large phantom (LP)).
Figure 4Mean pixel values in ROIs of various inserts of the CTP 404 module as the mAs is varied for the four XVI scanning protocols using the small phantom (SP): (a) head and neck, (b) prostate, (c) pelvis, and (d) chest.
Figure 5Mean pixel values in ROIs of various inserts of the CTP 404 module as the mAs is varied for three of the XVI scanning protocols using the large phantom (LP): (a) prostate, (b) pelvis, and (c) chest.
Figure 6Noise averaged over the five ROIs chosen in the CTP 486 module as the mAs is varied, using the small phantom (SP) and large phantom (LP), for (a) OBI and (b) XVI.
Figure 7Contrast‐to‐noise ratio measured using the polystyrene ROI and the background material (water‐equivalent) of the CTP 486 module as the mAs is varied using the small phantom (SP) and large phantom (LP), for (a) OBI and (b) XVI.
Figure 8Image uniformity index (UI) measured on the CTP 486 module as the mAs is varied, using the small phantom (SP) and large phantom (LP), for (a) OBI and (b) XVI. Positive value indicates a ‘cupping’ artifact, whereas a negative value indicates a ‘capping’ artifact.