| Literature DB >> 29667294 |
Hideharu Miura1,2, Shuichi Ozawa1,2, Toshiya Okazue1, Atsushi Kawakubo1, Kiyoshi Yamada1, Yasushi Nagata1,2.
Abstract
PURPOSE: Dual-source cone-beam computed tomography (DCBCT) is currently available in the Vero4DRT image-guided radiotherapy system. We evaluated the image quality and absorbed dose for DCBCT and compared the values with those for single-source CBCT (SCBCT).Entities:
Keywords: absorbed dose; dual-source cone-beam computed tomography; image quality; image-guided radiation therapy
Mesh:
Year: 2018 PMID: 29667294 PMCID: PMC5978565 DOI: 10.1002/acm2.12328
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Vero4DRT system with two kV imaging systems, each consisting of a kV x‐ray tube and flat panel detector (FPD) aligned at ±45° relative to the MV x‐ray beam axis. The diameter of the O‐ring structure is about 350 cm.
Figure 2Image acquisition using SCBCT (a) tube 1 rotated 215° between 5° to 220° (red) and (b) tube 2 rotated 215° between 275° to 130° (blue). (c) Image acquisition using DCBCT requires both kV x‐ray tubes be rotated 115°(tube1: 315°–70°, tube2: 45°–160°). The purple area indicates the overlap angle (45°–70°).
Figure 3DCBCT image of the uniformity phantom (CTP 486). Five ROIs (center, top, left side, bottom, and right side) on the image were used to evaluate image uniformity.
Figure 4DCBCT image of the HU reproducibility module (CTP 404) with seven inserts of materials with different densities. The blue circles are the ROIs inside the inserts and the yellow circles are the ROIs surrounding the inserts and are used to measure background. Numbers are the HU values from diagnostic CT.
Performance of SCBCT and DCBCT in image uniformity using a uniformity module
| SCBCT (tube 1) | SCBCT (tube 2) | DCBCT | |
|---|---|---|---|
| Center | −53.5 ± 6.3 | −43.7 ± 6.0 | −174.3 ± 5.0 |
| Top | −61.2 ± 8.1 | −59.6 ± 10.1 | −143.2 ± 6.8 |
| Left | −60.9 ± 7.4 | −54.5 ± 8.2 | −144.2 ± 6.5 |
| Bottom | −60.9 ± 7.6 | −60.0 ± 8.8 | −144.6 ± 6.8 |
| Right | −60.2 ± 9.6 | −62.3 ± 9.6 | −153.3 ± 5.7 |
| Image uniformity | 7.4 | 15.4 | 28.0 |
All values are average values ± standard deviation (SD) and are expressed in HU.
Figure 5Reference HU values as a function of measured HU values using SCBCT and DCBCT. Dashed lines are linear regression fits of the data and R 2 is the coefficient of determination. The solid gray line indicates where the reconstructed HU is equal to the reference HU.
Figure 6Comparison between image contrasts obtained using SCBCT and DCBCT of the seven insert materials of different densities. DCBCT yielded poorer image contrast than did SCBCT for all materials (low image contrast value indicates poor contrast). Error bars correspond to 1 SD for the mean value of five measurements.
Performance of SCBCT and DCBCT in MTF using a high‐contrast‐resolution module
| 50% MTF (mm−1) | 10% MTF (mm−1) | |
|---|---|---|
| SCBCT (tube 1) | 0.39 ± 0.01 | 0.77 ± 0.01 |
| SCBCT (tube 2) | 0.37 ± 0.02 | 0.75 ± 0.02 |
| DCBCT | 0.37 ± 0.01 | 0.76 ± 0.01 |
MTF, modulation transfer function.
All values are average values ± standard deviation (SD).
Absorbed doses and CBCTDIw values obtained using 16‐cm‐diameter cylindrical phantom
| SCBCT (tube 1) | SCBCT (tube 2) | DCBCT | |
|---|---|---|---|
| Center | 47.1 ± 0.2 | 50.9 ± 0.1 | 53.3 ± 0.1 |
| Top | 50.1 ± 0.1 | 96.5 ± 0.1 | 89.3 ± 0.1 |
| Left | 25.8 ± 0.1 | 51.4 ± 0.1 | 34.2 ± 0.1 |
| Bottom | 66.4 ± 0.1 | 29.6 ± 0.1 | 42.0 ± 0.1 |
| Right | 89.3 ± 0.1 | 77.0 ± 0.1 | 98.0 ± 0.1 |
| CBCTDIw | 54.3 ± 0.1 | 59.4 ± 0.1 | 61.7 ± 0.1 |
All values are expressed in mGy.
All values are average values ± standard deviation (SD).
Figure 7Axial images of the pelvis region obtained using (a) SCBCT and (b) DCBCT.