Karl R VanDerMeid1, Stephanie P Su, Kathleen L Krenzer, Keith W Ward, Jin-Zhong Zhang.
Abstract
PURPOSE: The Schirmer's test is commonly used in the clinic for the diagnosis of dry eye disease by measuring tear volume. This report describes a procedure which can be used to recover tears from the Schirmer strip for the measurement of multiple tear cytokines as well as matrix metalloproteinases (MMPs) by Luminex technology.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21552500 PMCID: PMC3086627
Source DB: PubMed Journal: Mol Vis ISSN: 1090-0535 Impact factor: 2.367
Clinical subject data.
| 1 | 54 | F | 37.5 | 1 | 10 | 15 |
| 2 | 41 | F | 25 | 1 | 6 | 7 |
| 3 | 26 | M | 0 | 0 | 28 | 32 |
| 4 | 53 | F | 20.8 | 1 | 6 | 14 |
| 5 | 25 | F | 0 | 0 | 35 | 35 |
*OSDI scale [34]: Normal (0) 0–20.7; Mild (1) 20.8–41.6; Moderate (2) 41.7–72.9; Severe (3) 73–100.
Percent recovery of cytokines diluted in buffer or extracted from spiked Schirmer strips.
| IL-1α | 77.8±4.7 | 66.9±4.0 | 84.6±5.6 | 79.8±4.2 |
| IL-1β | 81.5±6.2 | 59.0*±3.8 | 84.5±4.3 | 82.9±4.6 |
| IL-1rα | 86.0±4.5 | 67.7±3.5 | 74.5±5.3 | 68.5±3.5 |
| IL-4 | 96.9±8.7 | 37.2±3.5 | 62.9±5.1 | 6.7±1.4 |
| IL-6 | 82.5±3.5 | 68.2±2.7 | 78.2±3.3 | 67.1±3.3 |
| IL-8 | 91.1±5.3 | 68.3*±4.3 | 89.3±4.9 | 76.8±4.2 |
| IL-10 | 83.0±3.2 | 30.5*±2.8 | BLD | BLD |
| IL-13 | 83.8±5.0 | 49.7*±2.6 | 85.2±8.2 | 29.4*±3.5 |
| MCP-1 | 90.1±4.4 | 77.7±4.1 | 87.6±4.4 | 84.2±2.5 |
| TNF-α | 81.9±5.8 | 62.8±2.8 | 91.5±7.0 | 65.1*±4.3 |
*p<0.05 versus cytokine diluted in same assay buffer and calculated from same buffer standard curve. BLD: Below the limit of detection. n=10 for all data.
Percent recovery of MMPs diluted in buffer or extracted from spiked Schirmer strips.
| MMP-1 | 92.1 ± 7.8 | 88.9 ± 4.0 | 91.1 ± 9.7 | 78.5 ± 7.5 |
| MMP-2 | 94.3 ± 3.1 | 89.4 ± 5.1 | 86.8 ± 3.9 | 81.4 ± 5.0 |
| MMP-7 | 93.6 ± 5.0 | 68.2* ± 5.1 | 86.7 ± 2.5 | 76.3 ± 17.3 |
| MMP-9 | 96.7 ± 9.2 | 86.1 ± 8.5 | 100.2 ± 8.3 | 84.5 ± 6.8 |
| MMP-10 | 87.0 ± 6.3 | 79.3 ± 2.7 | 94.4 ± 7.5 | 77.2 ± 3.4 |
*p<0.05 versus MMP from diluted sample with extracted and concentration calculated from equivalent buffers. n=10 for all data.
Figure 1Graph of volume versus Schirmer strip measurement. Each data point represents the mean and standard deviation from 3 individual experiments. The line was determined by linear regression and the correlation coefficient was 0.999.
Figure 2Extraction buffer containing high salt and detergent has no effect on Luminex standard curves. Luminex cytokine standard curve for preparations in kit assay buffer: closed circles with solid line or extraction buffer; open circles with dashed line. *p<0.05 versus equivalent concentration in assay buffer.
Figure 3Concentrations of 10 cytokines extracted from clinical Schirmer strips. All data are expressed as mean±SEM.
Figure 4Concentrations of 5 MMPs extracted from clinical Schirmer strips. All data are expressed as mean±SEM.