Literature DB >> 21493795

Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study.

Sebastian T Schindera1, Lars Diedrichsen, Hubert C Müller, Oliver Rusch, Daniele Marin, Bernhard Schmidt, Rainer Raupach, Peter Vock, Zsolt Szucs-Farkas.   

Abstract

PURPOSE: To assess the diagnostic accuracy, image quality, and radiation dose of an iterative reconstruction algorithm compared with a filtered back projection (FBP) algorithm for abdominal computed tomography (CT) at different tube voltages.
MATERIALS AND METHODS: A custom liver phantom with 45 simulated hypovascular liver tumors (diameters of 5, 10, and 15 mm; tumor-to-liver contrast of 10, 25, and 50 HU) was placed in a cylindrical water container that mimicked an intermediate-sized patient. The phantom was scanned at 120, 100, and 80 kVp. The CT data sets were reconstructed with FBP and iterative reconstruction. The image noise was measured, and the contrast-to-noise ratio (CNR) of the tumors was calculated. The radiation dose was assessed with the volume CT dose index. Tumor detection was independently performed by three radiologists. Statistical analysis included analysis of variance.
RESULTS: Compared with the FBP data set at 120 kVp, the iterative reconstruction data set collected at 100 kVp demonstrated significantly lower mean image noise (20.9 and 16.7 HU, respectively; P < .001) and greater mean CNRs for the simulated tumors (P < .001). The iterative reconstruction data set collected at 120 kVp yielded the highest sensitivity for tumor detection, while the FBP data set at 80 kVp yielded the lowest. The sensitivity for the iterative reconstruction data set at 100 kVp was comparable with that for the FBP data set at 120 kVp (79.3% and 74.9%, respectively; P > .99). The volume CT dose index decreased by 39.8% between the 120-kVp protocol and the 100-kVp protocol and by 70.3% between the 120-kVp protocol and the 80-kVp protocol.
CONCLUSION: Results of this phantom study suggest that a 100-kVp abdominal CT protocol with an iterative reconstruction algorithm for simulated intermediate-sized patients increases the image quality and maintains the diagnostic accuracy at a reduced radiation dose when compared with a 120-kVp protocol with an FBP algorithm. © RSNA, 2011.

Entities:  

Mesh:

Year:  2011        PMID: 21493795     DOI: 10.1148/radiol.11102217

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  34 in total

1.  Iterative reconstruction of dual-source coronary CT angiography: assessment of image quality and radiation dose.

Authors:  Eun-Ah Park; Whal Lee; Kwang Woo Kim; Kwang Gi Kim; Allmendinger Thomas; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2011-12-21       Impact factor: 2.357

2.  Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults.

Authors:  Gaurav S Desai; Raul N Uppot; Elaine W Yu; Avinash R Kambadakone; Dushyant V Sahani
Journal:  Eur Radiol       Date:  2012-04-19       Impact factor: 5.315

3.  Aortic CT angiography dose reduction: investigation of optimal noise index and iterative algorithm strength in combination with low kV.

Authors:  Osvaldo Rampado; Stefania Busso; Domenica Garabello; Ezio Marengo; Marco Valerio; Simona Capello; Simona Veglia; Ottavio Davini; Roberto Ropolo
Journal:  Radiol Med       Date:  2015-12-16       Impact factor: 3.469

4.  Consistency of Renal Stone Volume Measurements Across CT Scanner Model and Reconstruction Algorithm Configurations.

Authors:  Alice E Huang; Juan C Montoya; Maria Shiung; Shuai Leng; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2017-04-12       Impact factor: 3.959

5.  Combining automated attenuation-based tube voltage selection and iterative reconstruction: a liver phantom study.

Authors:  Daniela B Husarik; Sebastian T Schindera; Fabian Morsbach; Natalie Chuck; Burkhardt Seifert; Zsolt Szucs-Farkas; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2013-10-24       Impact factor: 5.315

6.  Can virtual monochromatic images from dual-energy CT replace low-kVp images for abdominal contrast-enhanced CT in small- and medium-sized patients?

Authors:  Peijie Lv; Zhigang Zhou; Jie Liu; Yaru Chai; Huiping Zhao; Hua Guo; Daniele Marin; Jianbo Gao
Journal:  Eur Radiol       Date:  2018-11-30       Impact factor: 5.315

7.  Emergency assessment of patients with acute abdominal pain using low-dose CT with iterative reconstruction: a comparative study.

Authors:  Pierre-Alexandre Poletti; Minerva Becker; Christoph D Becker; Alice Halfon Poletti; Olivier T Rutschmann; Habib Zaidi; Thomas Perneger; Alexandra Platon
Journal:  Eur Radiol       Date:  2017-01-12       Impact factor: 5.315

8.  Image quality of adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection.

Authors:  Roh-Eul Yoo; Eun-Ah Park; Whal Lee; Hackjoon Shim; Yeo Koon Kim; Jin Wook Chung; Jae Hyung Park
Journal:  Int J Cardiovasc Imaging       Date:  2012-08-26       Impact factor: 2.357

9.  C-arm flat-panel CT arthrography of the wrist and elbow: first experiences in human cadavers.

Authors:  Roman Guggenberger; Fabian Morsbach; Hatem Alkadhi; Magdalena Vich; Thomas Pfammatter; Juerg Hodler; Gustav Andreisek
Journal:  Skeletal Radiol       Date:  2012-08-29       Impact factor: 2.199

10.  Optimizing CT technique to reduce radiation dose: effect of changes in kVp, iterative reconstruction, and noise index on dose and noise in a human cadaver.

Authors:  Kevin J Chang; Scott Collins; Baojun Li; William W Mayo-Smith
Journal:  Radiol Phys Technol       Date:  2016-10-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.