[Does osteoporosis lead to reduction the primary stability of cementless hip cups?].

C von Schulze-Pellengahr1, A Bürkner, T Lichtinger, W Teske, A Fottner, B Wegener, T Vogel.   

Abstract

BACKGROUND: Cementless hip cups require adequate primary stability in order to achieve osseointegration. Depending on implant design, a different biomechanical behavior in osteoporotic bone and in bone with normal bone mineral density might be expected.
MATERIAL AND METHODS: The micromovement between bone and implant was determined in macerated human hip acetabula with reduced and with normal bone density for two different cementless screwed cups and two different cementless press-fit cups. The bone mineral density was assessed by means of Q-CT and 20 implantations were performed for each acetabulum with reduced and with normal bone density.
RESULTS: Screwed cups showed significantly less micromovement than press-fit cups. The average micromovement of all cups, with the exception of the Monoblock cup, remained below the value of 100 µ, the upper limit for successful osseointegration. Values of over 200 µ were measured only for the Monoblock cup and only in some of the tests. There was no significant difference between the micro-movement in acetabula with reduced bone mineral density and with normal bone.
CONCLUSION: Impaired osseointegration in osteoporotic bone because of increased micromovement is not to be expected in the investigated cementless hip cups as no relevant differences in micromovement were found between normal and osteoporotic bone stock. On the basis of the good clinical results with the Monoblock cup, the upper limit of 100 µ of micromotion for osseointegation in macerated bone is probably set too low.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21347859     DOI: 10.1007/s00132-010-1733-y

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  10 in total

1.  [Primary stability of cement-free press-fit acetabulum cups. In vitro displacement studies].

Authors:  A Kuhn; G Scheller; M Schwarz
Journal:  Biomed Tech (Berl)       Date:  1999-12       Impact factor: 1.411

2.  Mechanical regulation of localized and appositional bone formation around bone-interfacing implants.

Authors:  C A Simmons; S A Meguid; R M Pilliar
Journal:  J Biomed Mater Res       Date:  2001-04

3.  Differences in osseointegration rate due to implant surface geometry can be explained by local tissue strains.

Authors:  C A Simmons; S A Meguid; R M Pilliar
Journal:  J Orthop Res       Date:  2001-03       Impact factor: 3.494

4.  Acetabular micromotion as a measure of initial implant stability in primary hip arthroplasty. An in vitro comparison of different methods of initial acetabular component fixation.

Authors:  P G Perona; J Lawrence; W G Paprosky; A G Patwardhan; M Sartori
Journal:  J Arthroplasty       Date:  1992-12       Impact factor: 4.757

5.  Mechanical stability of porous-coated acetabular components in total hip arthroplasty.

Authors:  J B Stiehl; E MacMillan; D A Skrade
Journal:  J Arthroplasty       Date:  1991-12       Impact factor: 4.757

6.  [Factors affecting the primary stability of acetabular components. An in vitro study].

Authors:  R P Pitto; J Böhner; V Hofmeister
Journal:  Biomed Tech (Berl)       Date:  1997-12       Impact factor: 1.411

7.  Acetabular strains produced by oversized press fit cups.

Authors:  M D Ries; M Harbaugh
Journal:  Clin Orthop Relat Res       Date:  1997-01       Impact factor: 4.176

8.  Micromotion of cementless hemispherical acetabular components. Does press-fit need adjunctive screw fixation?

Authors:  C H Won; T C Hearn; M Tile
Journal:  J Bone Joint Surg Br       Date:  1995-05

9.  Influence of a functional dynamic loading on bone ingrowth into surface pores of orthopedic implants.

Authors:  P Ducheyne; P De Meester; E Aernoudt
Journal:  J Biomed Mater Res       Date:  1977-11

Review 10.  Hip joint loading during walking and running, measured in two patients.

Authors:  G Bergmann; F Graichen; A Rohlmann
Journal:  J Biomech       Date:  1993-08       Impact factor: 2.712

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.