The effect of patient positioning aids on PET quantification in PET/MR imaging.

Frederic Mantlik1, Matthias Hofmann, Matthias K Werner, Alexander Sauter, Jürgen Kupferschläger, Bernhard Schölkopf, Bernd J Pichler, Thomas Beyer.   

Abstract

OBJECTIVES: Clinical PET/MR requires the use of patient positioning aids to immobilize and support patients for the duration of the combined examination. Ancillary immobilization devices contribute to overall attenuation of the PET signal, but are not detected with conventional MR sequences and, hence, are ignored in standard MR-based attenuation correction (MR-AC). We report on the quantitative effect of not accounting for the attenuation of patient positioning aids in combined PET/MR imaging.
METHODS: We used phantom and patient data acquired with positioning aids on a PET/CT scanner (Biograph 16, HI-REZ) to mimic PET/MR imaging conditions. Reference CT-based attenuation maps were generated from measured (original) CT transmission images (origCT-AC). We also created MR-like attenuation maps by following the same conversion procedure of the attenuation values except for the prior delineation and subtraction of the positioning aids from the CT images (modCT-AC). First, a uniform (68)Ge cylinder was positioned centrally in the PET/CT scanner and fixed with a vacuum mattress (10 cm thick) and, in a repeat examination, with MR positioning foam pads. Second, 16 patient datasets were selected for subsequent processing. All patients were regionally immobilized with positioning aids: a vacuum mattress for head/neck imaging (nine patients) and a foam mattress for imaging of the lower extremities (seven patients). PET images were reconstructed following CT-based attenuation and scatter correction using the original and modified (MR-like) CT images: PET(origCT-AC) and PET(modCT-AC), respectively. PET images following origCT-AC and modCT-AC were compared visually and in terms of mean differences of voxels with a standardized uptake value of at least 1.0. In addition, we report maximum activity concentration in lesions for selected patients.
RESULTS: In the phantom study employing the vacuum mattress the average voxel activity in PET(modCT-AC) was underestimated by 6.4% compared to PET(origCT-AC), with 3.4% of the PET voxels being underestimated by 10% or more. When the MR foam pads were not accounted for during AC, PET(modCT-AC) was underestimated by 1.1% on average, with none of the PET voxels being underestimated by 10% or more. Evaluation of the head/neck patient data showed a decrease of 8.4% ([(68)Ga]DOTATOC) and 7.4% ([(18)F]FDG) when patient positioning aids were not accounted for during AC, while the corresponding decrease was insignificant for the lower extremities.
CONCLUSION: Depending on the size and density of the positioning aids used, a regionally variable underestimation of PET activity following AC is observed when positioning aids are not accounted for. This underestimation may become relevant in combined PET/MR imaging of patients with neuropsychiatric indications, but appears to be of no clinical relevance in imaging the extremities.

Entities:  

Mesh:

Year:  2011        PMID: 21308373     DOI: 10.1007/s00259-010-1721-9

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  13 in total

1.  Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner.

Authors:  G Delso; A Martinez-Möller; R A Bundschuh; R Ladebeck; Y Candidus; D Faul; S I Ziegler
Journal:  Phys Med Biol       Date:  2010-07-20       Impact factor: 3.609

Review 2.  Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques.

Authors:  Matthias Hofmann; Bernd Pichler; Bernhard Schölkopf; Thomas Beyer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

Review 3.  Combined PET/MRI: a new dimension in whole-body oncology imaging?

Authors:  Gerald Antoch; Andreas Bockisch
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-03       Impact factor: 9.236

4.  PET/MRI: paving the way for the next generation of clinical multimodality imaging applications.

Authors:  Bernd J Pichler; Armin Kolb; Thomas Nägele; Heinz-Peter Schlemmer
Journal:  J Nucl Med       Date:  2010-02-11       Impact factor: 10.057

5.  MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences.

Authors:  Vincent Keereman; Yves Fierens; Tom Broux; Yves De Deene; Max Lonneux; Stefaan Vandenberghe
Journal:  J Nucl Med       Date:  2010-05       Impact factor: 10.057

6.  Fixation devices for whole-body 18F-FDG PET/CT: patient perspectives and technical aspects.

Authors:  Klaus Brechtel; Henriette Heners; Marc Mueller; Philip Aschoff; Susanne M Eschmann; Roland Bares; Claus D Claussen; Anna C Pfannenberg
Journal:  Nucl Med Commun       Date:  2007-02       Impact factor: 1.690

7.  Toward implementing an MRI-based PET attenuation-correction method for neurologic studies on the MR-PET brain prototype.

Authors:  Ciprian Catana; Andre van der Kouwe; Thomas Benner; Christian J Michel; Michael Hamm; Matthias Fenchel; Bruce Fischl; Bruce Rosen; Matthias Schmand; A Gregory Sorensen
Journal:  J Nucl Med       Date:  2010-09       Impact factor: 10.057

8.  Hybrid PET/MRI of intracranial masses: initial experiences and comparison to PET/CT.

Authors:  Andreas Boss; Sotirios Bisdas; Armin Kolb; Matthias Hofmann; Ulrike Ernemann; Claus D Claussen; Christina Pfannenberg; Bernd J Pichler; Matthias Reimold; Lars Stegger
Journal:  J Nucl Med       Date:  2010-07-21       Impact factor: 10.057

9.  Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

Authors:  Axel Martinez-Möller; Michael Souvatzoglou; Gaspar Delso; Ralph A Bundschuh; Christophe Chefd'hotel; Sibylle I Ziegler; Nassir Navab; Markus Schwaiger; Stephan G Nekolla
Journal:  J Nucl Med       Date:  2009-03-16       Impact factor: 10.057

10.  Simultaneous MR/PET imaging of the human brain: feasibility study.

Authors:  Heinz-Peter W Schlemmer; Bernd J Pichler; Matthias Schmand; Ziad Burbar; Christian Michel; Ralf Ladebeck; Kirstin Jattke; David Townsend; Claude Nahmias; Pradeep K Jacob; Wolf-Dieter Heiss; Claus D Claussen
Journal:  Radiology       Date:  2008-09       Impact factor: 11.105

View more
  10 in total

1.  Evaluation of an attenuation correction method for PET/MR imaging of the head based on substitute CT images.

Authors:  Anne Larsson; Adam Johansson; Jan Axelsson; Tufve Nyholm; Thomas Asklund; Katrine Riklund; Mikael Karlsson
Journal:  MAGMA       Date:  2012-09-07       Impact factor: 2.310

2.  Comparison of lesion detection and quantitation of tracer uptake between PET from a simultaneously acquiring whole-body PET/MR hybrid scanner and PET from PET/CT.

Authors:  Marco Wiesmüller; Harald H Quick; Bharath Navalpakkam; Michael M Lell; Michael Uder; Philipp Ritt; Daniela Schmidt; Michael Beck; Torsten Kuwert; Carl C von Gall
Journal:  Eur J Nucl Med Mol Imaging       Date:  2012-10-06       Impact factor: 9.236

Review 3.  Attenuation Correction for Magnetic Resonance Coils in Combined PET/MR Imaging: A Review.

Authors:  Mootaz Eldib; Jason Bini; David D Faul; Niels Oesingmann; Charalampos Tsoumpas; Zahi A Fayad
Journal:  PET Clin       Date:  2015-11-27

Review 4.  MR Imaging-Guided Attenuation Correction of PET Data in PET/MR Imaging.

Authors:  David Izquierdo-Garcia; Ciprian Catana
Journal:  PET Clin       Date:  2016-01-26

5.  Assessment of a carbon fibre MRI flatbed insert for radiotherapy treatment planning.

Authors:  Maysam M Jafar; Jonathan Reeves; Matthieu A Ruthven; Christopher J Dean; Niall D MacDougall; Arthur T Tucker; Marc E Miquel
Journal:  Br J Radiol       Date:  2016-04-01       Impact factor: 3.039

6.  MLAA-based attenuation correction of flexible hardware components in hybrid PET/MR imaging.

Authors:  Thorsten Heußer; Christopher M Rank; Yannick Berker; Martin T Freitag; Marc Kachelrieß
Journal:  EJNMMI Phys       Date:  2017-03-01

7.  Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer.

Authors:  Martin Meyerspeer; Ewald Unger; Ivo Rausch; Alejandra Valladares; Lalith Kumar Shiyam Sundar; Thomas Beyer; Marcus Hacker
Journal:  EJNMMI Phys       Date:  2021-02-18

8.  3D Printing of Abdominal Immobilization Masks for Therapeutics: Dosimetric, Mechanical and Financial Analysis.

Authors:  Jessica Duarte; Maria Amélia Ramos Loja; Ricardo Portal; Lina Vieira
Journal:  Bioengineering (Basel)       Date:  2022-01-29

Review 9.  MRI for attenuation correction in PET: methods and challenges.

Authors:  Gudrun Wagenknecht; Hans-Jürgen Kaiser; Felix M Mottaghy; Hans Herzog
Journal:  MAGMA       Date:  2012-11-21       Impact factor: 2.310

10.  PET/CT versus body coil PET/MRI: how low can you go?

Authors:  P Appenzeller; C Mader; M W Huellner; D Schmidt; D Schmid; A Boss; G von Schulthess; P Veit-Haibach
Journal:  Insights Imaging       Date:  2013-05-15
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.